长轴视野PET扫描仪上基于深度学习的散射校正评估
该研究在长轴视野(LAFOV)全身PET上评估基于卷积神经网络U-Net的深度学习散射估计(DLSE)方法,用XCAT体模蒙特卡洛模拟训练网络,并在7例[18F]-FDG和[18F]-PSMA临床数据中与单散射模拟(SSS)比较。DLSE在体模数据上精度更优、对患者体型和剂量变化更稳健、病灶对比度恢复更好,临床结果有前景。
要点
- DLSE在体模数据上较SSS精度更优,对患者体型和剂量变化更稳健。
- DLSE在[18F]-FDG数据集上改善病灶对比度,在[18F]-PSMA数据集上与SSS表现一致。
- LAFOV PET散射可从原始数据用DLSE方法准确估计。
核心结论
- LAFOV PET散射可从原始数据用所提出的DLSE方法准确估计。
研究设计与样本
- 方法学研究,为既往在常规PET系统上开展的DLSE评估的扩展,在LAFOV全身PET上评估DLSE性能。
- DLSE基于卷积神经网络U-Net架构,使用发射和衰减正弦图估计散射正弦图。
- 网络由XCAT体模[18F]-FDG PET采集的蒙特卡洛模拟训练,采用Siemens Biograph Vision Quadra扫描仪模型,含多种形态和剂量分布。
- 先在模拟数据上于正弦图域和图像域与蒙特卡洛基准及SSS散射正弦图比较,再在7例[18F]-FDG和[18F]-PSMA临床数据集上与SSS估计比较。
主要发现
- DLSE在体模数据上精度更优,较SSS对患者体型和剂量变化更稳健,病灶对比度恢复更好。
- 临床结果有前景:DLSE在[18F]-FDG数据集上改善病灶对比度。
- DLSE在[18F]-PSMA数据集上与SSS表现一致,尽管未用[18F]-PSMA训练。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Long-axial field-of-view (LAFOV) positron emission tomography (PET) systems allow higher sensitivity, with an increased number of detected lines of response induced by a larger angle of acceptance. However this extended angle increases the number of multiple scatters and the scatter contribution within oblique planes. As scattering affects both quality and quantification of the reconstructed image, it is crucial to correct this effect with more accurate methods than the state-of-the-art single scatter simulation (SSS) that can reach its limits with such an extended field-of-view (FOV). In this work, which is an extension of our previous assessment of deep learning-based scatter estimation (DLSE) carried out on a conventional PET system, we aim to evaluate the DLSE method performance on LAFOV total-body PET. APPROACH: The proposed DLSE method based on an convolutional neural network (CNN) U-Net architecture uses emission and attenuation sinograms to estimate scatter sinogram. The network was trained from Monte-Carlo (MC) simulations of XCAT phantoms [ 18 F]-FDG PET acquisitions using a Siemens Biograph Vision Quadra scanner model, with multiple morphologies and dose distributions. We firstly evaluated the method performance on simulated data in both sinogram and image domain by comparing it to the MC ground truth and SSS scatter sinograms. We then tested the method on seven [ 18 F]-FDG and [ 18 F]-PSMA clinical datasets, and compare it to SSS estimations. RESULTS: DLSE showed superior accuracy on phantom data, greater robustness to patient size and dose variations compared to SSS, and better lesion contrast recovery. It also yielded promising clinical results, improving lesion contrasts in [ 18 F]-FDG datasets and performing consistently with [ 18 F]-PSMA datasets despite no training with [ 18 F]-PSMA. SIGNIFICANCE: LAFOV PET scatter can be accurately estimated from raw data using the proposed DLSE method.
引用
- Laurent B, Bousse A, Merlin T, et al. Evaluation of deep learning-based scatter correction on a long-axial field-of-view PET scanner. Eur J Nucl Med Mol Imaging. 2025. PMID: 39918764. doi:10.1007/s00259-025-07120-6
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。