文献速递肿瘤 PET/CT甲状腺核医学

利用小体素与最优采集时间优化儿童神经母细胞瘤[18F]mFBG全身PET/CT显像的初步研究

Leveraging small voxel with optimal acquisition time for [18F]mFBG total-body PET/CT imaging in pediatric patients with neuroblastoma: a preliminary study.
PubMed 文献 · Eur J Nucl Med Mol Imaging · 2025-01-31

该研究回顾性纳入33例神经母细胞瘤患儿共37次[18F]mFBG全身uEXPLORER PET/CT扫描,比较不同采集时间(0.5–10分钟)与矩阵大小(192×192、512×512、1024×1024)对图像质量和诊断信心的影响。结果显示512×512矩阵至少采集4分钟即可获得足够主观图像质量,且对不确定病灶的诊断信心显著提高。作者推荐[18F]mFBG全身PET/CT采用512×512矩阵、最短4分钟采集。

要点

  • 512×512矩阵至少采集4分钟可获得足够主观图像质量(所有评分≥3分)。
  • 192×192矩阵至少采集2分钟即可获得足够主观图像质量。
  • 512×512矩阵至少采集2分钟及1024×1024矩阵10分钟采集对不确定病灶诊断信心显著更高。

核心结论

  • 作者推荐[18F]mFBG全身PET/CT采用512×512矩阵、最短4分钟采集,以获得足够图像质量并提高诊断信心,尤其适用于不确定病灶的检测。

研究设计与样本

  • 回顾性纳入33例神经母细胞瘤患儿,共37次[18F]mFBG全身uEXPLORER PET/CT扫描,采集时长10分钟;按不同采集时间(0.5–10分钟)和三种矩阵大小(192×192、512×512、1024×1024)重建图像,由核医学医师分析主观评分(5分制)与客观信噪比(SNR)。

主要发现

  • 33例患儿37次扫描中,17例患儿20次扫描[18F]mFBG PET/CT结果为阳性。
  • 192×192矩阵至少采集2分钟、512×512矩阵至少采集4分钟即可获得足够主观图像质量(所有评分≥3分)。
  • 同一体素大小下SNR随采集时间延长而升高,体素缩小时SNR降低。
  • G192、G512与G1024-10三组的Curie与SIOPEN评分保持一致。
  • 512×512矩阵至少采集2分钟组及1024×1024矩阵10分钟组对G192-10图像中不确定病灶的诊断信心评分显著更高,几乎所有不确定病灶被评为非常有信心。

关键数值与推荐

  • 主观图像质量足够的标准为5分制评分全部≥3分:192×192矩阵需至少2分钟采集,512×512矩阵需至少4分钟采集。
  • G512组至少2分钟采集与G1024-10组对不确定病灶的诊断信心评分显著高于G192-10组。

临床怎么用

  • 作者推荐[18F]mFBG全身PET/CT采用512×512矩阵、最短4分钟采集,以获得足够图像质量并改善诊断信心,尤其适用于不确定病灶的检测。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: The advent of total-body PET/CT presents an opportunity for significant advancements in imaging of neuroblastoma with [18F]meta-fluorobenzylguanidine ([18F]mFBG). Small voxel imaging has proven to have better lesion detectability but need enough radioactivity counts. This study aims to balance shortened acquisition times and small voxel reconstruction to keep sufficient image quality and diagnostic confidence on [18F]mFBG total-body PET for neuroblastoma. METHODS: We retrospectively enrolled 33 pediatric patients with neuroblastoma who underwent 37 [18F]mFBG total-body uEXPLORER PET/CT scans of 10-min duration. PET images were reconstructed with varying acquisition times (0.5-10 min) and three matrix sizes (192 × 192, 512 × 512 and 1024 × 1024). The subjective (scored on a 5-point scale) and objective image quality (signal-to-noise ratio, SNR) of all the sets of reconstructed images were analyzed by nuclear medicine physicians. For indeterminate lesions identified in the group of 192 × 192 matrix with the 10-min scan (G192-10), diagnostic confidence was further evaluated in images reconstructed with the 512 × 512 and 1024 × 1024 matrices (G512 and G1024). RESULTS: Of the 33 patients with 37 [18F]mFBG PET/CT scans, 17 patients with 20 scans had positive [18F]mFBG PET/CT findings. Sufficient subjective image quality was achieved with at least 2-min acquisition of 192 × 192 matrix and 4-min acquisition of 512 × 512 matrix (with all scores ≥ 3). SNR increased with longer acquisition times for the same voxel size, while decreased as voxel size shrunk. Although the Curie and SIOPEN scores remained consistent across G192, G512, and G1024-10 groups, the G512 groups with at least 2-min acquisition and G1024-10 showed significantly higher confidence scores for characterizing indeterminate lesions on the G192-10 images, with almost all indeterminate lesions being rated as very confident. CONCLUSIONS: A matrix of 512 × 512 with a minimum of 4-min acquisition on [18F]mFBG total-body PET/CT is recommended for sufficient image quality and improved diagnostic confidence, particularly in detecting indeterminate lesions.