接受肺移植评估的患者中吸入99mTc-锝气体与133Xe定量相对肺通气可比性的研究

Comparability of Quantifying Relative Lung Ventilation with Inhaled 99mTc-Technegas and 133Xe in Patients Undergoing Evaluation for Lung Transplantation.
PubMed 文献 · J Nucl Med · 2025-01-03

该研究对接受肺移植评估的3期试验参与者进行二次分析,对比吸入133Xe与99mTc-锝气体显像定量相对肺通气的相关性与一致性。74例受试者中,两种显像剂定量结果呈强正相关,Bland-Altman图显示高度一致,支持99mTc-锝气体用于相对肺通气定量。

要点

  • 74例受试者中,133Xe与99mTc-锝气体定量相对肺通气呈强正相关。
  • 74例中72例(97.3%)两种方法的相对通气百分比差异在±15%以内。
  • 133Xe与后位99mTc-锝气体图像的相关性最强(r=0.79)。

核心结论

  • 133Xe与99mTc-锝气体获得的相对肺通气百分比之间呈强正相关且高度一致,为99mTc-锝气体用于相对肺通气定量提供了重要临床证据。

研究设计与样本

  • 对前瞻性招募的3期试验参与者数据进行二次分析,受试者均接受133Xe与99mTc-锝气体通气显像,共74例。
  • 采用半自动分割异步分析图像,提取相对肺通气百分比;分析后位、前位99mTc-锝气体图像得出三套相对通气百分比(后位、前位和几何均值数据),并与后位133Xe图像数值比较。

主要发现

  • 74例受试者中,133Xe与99mTc-锝气体定量的相对肺通气呈强正相关,Bland-Altman图显示高度一致。
  • 74例中72例(97.3%)两种显像剂的相对通气百分比测量值差异在±15%以内。
  • 74例中70例(94.6%)相对通气测量差异在均值的95%CI范围内,74例中68例(91.9%)在更窄的±10%阈值内。
  • 133Xe与后位99mTc-锝气体图像的相关系数最强(r=0.79);几何均值法相关系数略低但仍可比(r=0.77);与前位99mTc-锝气体图像的相关系数最差(r=0.70)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

99mTc-Technegas was recently approved by the U.S. Food and Drug Administration as a radiopharmaceutical for ventilation scintigraphy. However, there are currently no data comparing the quantification of relative lung ventilation with 99mTc-Technegas with that performed using the standard approach with inhaled 133Xe. Methods: We performed a secondary analysis of data from prospectively recruited participants in a phase 3 trial undergoing evaluation for lung transplantation who received both 133Xe and 99mTc-Technegas ventilation imaging. The 133Xe and 99mTc-Technegas images were analyzed asynchronously using semiautomatic segmentation to extract the relative lung ventilation percentages. The anterior and posterior 99mTc-Technegas images were analyzed to derive 3 sets of relative ventilation percentages (posterior, anterior, and geometric mean data) and compared with the values from posterior 133Xe images. We evaluated for correlation and agreement between the relative lung ventilation percentages obtained using these 2 radiopharmaceuticals. Results: In a cohort of 74 participants, we found a strong positive correlation in the relative lung ventilation quantified using 133Xe with that using 99mTc-Technegas. A high level of agreement was demonstrated on the Bland-Altman plot comparing the 2 imaging modalities. Seventy-two of 74 participants (97.3%) had their relative ventilation percentage measurements within ±15% for 133Xe and 99mTc-Technegas. The differences in relative ventilation measurements were within the 95% CI limits of the mean for 70 of 74 participants (94.6%) and within a narrower ±10% threshold for 68 of 74 participants (91.9%), again reflecting the comparability of the 2 techniques. The strongest correlation coefficient (r = 0.79) was observed between the relative ventilation percentages obtained from 133Xe and posterior 99mTc-Technegas images. The geometric mean method had a slightly lower but still comparable correlation (r = 0.77), and as expected, the correlation with the anterior 99mTc-Technegas images was worst (r = 0.70). Conclusion: We showed a strong positive correlation and high agreement between the relative lung ventilation percentages obtained using 133Xe and 99mTc-Technegas. These data provide important clinical evidence supporting the use of 99mTc-Technegas for quantification of relative lung ventilation.

来源:J Nucl Med,查看原文。版权归原作者所有。