长轴视野PET/CT:心血管疾病中的技术要点
本文综述长轴视野(LAFOV)PET/CT在心血管疾病中的技术要点。LAFOV系统提供更大解剖覆盖和显著提高的PET灵敏度,可实现单床位从头到骨盆成像,约2米轴视野系统甚至可单次扫描完成全身成像。文章讨论其在心血管疾病应用中的机遇与挑战,包括流程考虑、数据处理、动脉输入函数提取和伪影管理。
要点
- LAFOV PET/CT提供更大解剖覆盖和显著提高的PET灵敏度。
- 约2米轴视野系统可单次扫描完成全身成像。
- 在心血管疾病中的应用面临数据处理、动脉输入函数提取和伪影管理挑战。
核心结论
- LAFOV PET/CT在心血管疾病中具有改善全身心血管评估的潜力,并可解决传统短轴视野(SAFOV)设备的固有局限。
- LAFOV系统在心血管疾病中的流程考虑、数据处理复杂性、动脉输入函数提取和伪影管理等问题尚未被充分探索。
研究设计与样本
- 本文为综述,讨论LAFOV PET/CT与心血管疾病相关的技术方面,重点介绍关键机遇与挑战,并分析这些因素对最常见心血管疾病评估的影响。
主要发现
- LAFOV PET/CT系统提供更大的解剖覆盖和显著增强的PET灵敏度。
- LAFOV系统可实现单床位从头到骨盆成像。
- 在轴视野约2米的系统中,甚至可单次扫描完成全身成像。
局限性
- LAFOV系统在心血管疾病中的流程考虑、数据处理复杂性、动脉输入函数提取和伪影管理尚未被充分探索。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Positron emission tomography / computed tomography (PET/CT) plays a pivotal role in the assessment of cardiovascular diseases (CVD), particularly in the context of ischemic heart disease. Nevertheless, its application in other forms of CVD, such as infiltrative, infectious, or inflammatory conditions, remains limited. Recently, PET/CT systems with an extended axial field of view (LAFOV) have been developed, offering greater anatomical coverage and significantly enhanced PET sensitivity. These advancements enable head-to-pelvis imaging with a single bed position, and in systems with an axial field of view (FOV) of approximately 2 meters, even total body (TB) imaging is feasible in a single scan session. The application of LAFOV PET/CT in CVD presents a promising opportunity to improve systemic cardiovascular assessments and address the limitations inherent to conventional short axial field of view (SAFOV) devices. However, several technical challenges, including procedural considerations for LAFOV systems in CVD, complexities in data processing, arterial input function extraction, and artefact management, have not been fully explored. This review aims to discuss the technical aspects of LAFOV PET/CT in relation to CVD by highlighting key opportunities and challenges and examining the impact of these factors on the evaluation of most relevant CVD.
引用
- Martinez-Lucio TS, Mendoza-Ibañez OI, Liu W, et al. Long Axial Field of View PET/CT: Technical Aspects in Cardiovascular Diseases. Semin Nucl Med. 2024. PMID: 39537432. doi:10.1053/j.semnuclmed.2024.10.009
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。