长轴视野PET/CT提高放射组学特征可靠性
研究分析34例接受长轴视野(LAFOV)PET/CT的患者,涵盖2-[18F]-FDG、[18F]PSMA-1007和[68Ga]Ga-DOTATOC三类显像剂。通过不同采集时长与接受角重建评估放射组学特征可靠性,发现采集时间延长与最大环差提高可改善特征一致性,全计数全环差数据更能捕捉特征异质性。
要点
- 采集至240秒时几乎所有特征与全计数数据至少呈中等一致,至360秒时几乎均呈优良一致。
- 与模拟标准轴视野采集相比,长轴视野数据中呈中等或差一致的特征更少,[68Ga]Ga-DOTATOC最为明显。
- 环差85与环差322在全部大于2秒的采集时长下均显示高度保真。
核心结论
- 较长采集时间与较高环差下特征可靠性改善,显示高灵敏度长轴视野系统可获得可重复、低噪声数据。
研究设计与样本
- 对34例患者进行分析:12例接受肿瘤2-[18F]-FDG PET/CT,14例接受[18F]PSMA-1007,8例接受[68Ga]Ga-DOTATOC。
- 使用106 cm长轴视野系统采集10分钟,由列表模式数据生成2、5、10、20、30、60、120、240、360和600秒的投影数据,分别采用标准接受角(最小环差MRD 85晶体)与最大接受角(MRD 322)。
- 对目标病灶进行分割并计算放射组学特征,以皮尔逊积矩相关系数(PPMCC)评估特征与全时长采集的相关性。
- 按优良(r > 0.9)、中等(r > 0.7且 < 0.9)和差(r ≤ 0.7)统计特征数作为稳定性指标,并用四分位离散系数评估组内异质性。
主要发现
- 随采集时间延长,所有特征的PPMCC提高。
- 至240秒时几乎所有特征与全计数(C100%)数据至少呈中等一致,至360秒时几乎均呈优良一致。
- 与标准轴视野(SAFOV)等效扫描相比,长轴视野中呈中等或差一致的特征更少,[68Ga]Ga-DOTATOC最为明显。
- MRD 322下C100%数据更能捕捉患者间异质性。
关键数值与推荐
- 特征相关性分级:优良为r > 0.9,中等为r > 0.7且 < 0.9,差为r ≤ 0.7。
- 采集时长设置为2、5、10、20、30、60、120、240、360和600秒,全时长为10分钟。
- 标准接受角为MRD 85晶体,最大接受角为MRD 322。
- 长轴视野系统轴向长度为106 cm。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: To assess the influence of long-axial field-of-view (LAFOV) PET/CT systems on radiomics feature reliability, to assess the suitability for short-duration or low-activity acquisitions for textural feature analysis and to investigate the influence of acceptance angle. METHODS: 34 patients were analysed: twelve patients underwent oncological 2-[18F]-FDG PET/CT, fourteen [18F]PSMA-1007 and eight [68Ga]Ga-DOTATOC. Data were obtained using a 106 cm LAFOV system for 10 min. Sinograms were generated from list-mode data corresponding to scan durations of 2, 5, 10, 20, 30, 60, 120, 240, 360 and 600s using both standard (minimum ring difference MRD 85 crystals) and maximum acceptance angles (MRD 322). Target lesions were segmented and radiomics features were calculated. To assess feature correlation, Pearson's product-moment correlation coefficient (PPMCC) was calculated with respect to the full duration acquisition for MRD 85 and 322 respectively. The number of features with excellent (r > 0.9), moderate (r > 0.7 and < 0.9) and poor (r ≤ 0.7) correlation was compared as a measure of feature stability. Intra-class heterogeneity was assessed by means of the quartile coefficient of dispersion. RESULTS: As expected, PPMCC improved with acquisition time for all features. By 240s almost all features showed at least moderate agreement with the full count (C100%) data, and by 360s almost all showed excellent agreement. Compared to standard-axial field of view (SAFOV) equivalent scans, fewer features showed moderate or poor agreement, and this was most pronounced for [68Ga]Ga-DOTATOC. Data obtained at C100% at MRD 322 were better able to capture between-patient heterogeneities. CONCLUSION: The improved feature reliability at longer acquisition times and higher MRD demonstrate the advantages of high sensitivity LAFOV systems for reproducible and low-noise data. High fidelity between MRD 85 and MRD 322 was seen at all scan durations > 2s. When contrasted with data comparable to a simulated SAFOV acquisition, full-count and full-MRD data were better able to capture underlying feature heterogeneities.
引用
- Alberts IL, Xue S, Sari H, et al. Long-axial field-of-view PET/CT improves radiomics feature reliability. Eur J Nucl Med Mol Imaging. 2024. PMID: 39477863. doi:10.1007/s00259-024-06921-5
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。