胸主动脉微钙化活动在正电子发射断层显像与磁共振成像联合检查中的评估

Thoracic aortic microcalcification activity in combined positron emission tomography and magnetic resonance imaging.
PubMed 文献 · Eur J Nucl Med Mol Imaging · 2024-03-08

研究纳入30例无症状二叶式主动脉瓣疾病监测患者,同次行钠[18F]氟化物PET-CT与PET-MRI升主动脉显像,PET-MRI采用三种衰减校正重建。PET-MRI各重建法与PET-CT的主动脉总标准化摄取值强相关(Dixon R=0.70;RadialVIBE-2 R=0.63;RadialVIBE-4 R=0.64;均p<0.001),RadialVIBE-2与PET-CT定量一致性最佳且受伪影影响较小。

要点

  • 30例无症状二叶式主动脉瓣疾病患者同次行钠[18F]氟化物PET-CT与PET-MRI升主动脉显像。
  • PET-MRI各衰减校正重建的主动脉总标准化摄取值与PET-CT强相关(Dixon R=0.70;RadialVIBE-2 R=0.63;RadialVIBE-4 R=0.64;均p<0.001)。
  • RadialVIBE-2衰减校正与PET-CT参照标准定量一致性最佳且受伪影影响较小,但PET-MRI组织分割仍存挑战。

核心结论

  • PET-MRI升主动脉微钙化分析与PET-CT结果可比,技术可行。
  • RadialVIBE-2组织衰减校正与PET-CT参照标准相关性最佳且较少受伪影影响。
  • PET-MRI重建中组织类型分割仍存挑战,需要改进衰减校正方法。

研究设计与样本

  • 30例无症状二叶式主动脉瓣疾病监测患者在同一次就诊中接受升胸主动脉钠[18F]氟化物PET-CT与PET-MRI检查。
  • PET-MRI数据采用三种迭代衰减校正重建:Dixon、RadialVIBE-2与RadialVIBE-4组织分割。
  • 对PET-CT与PET-MRI图像进行主动脉钠[18F]氟化物摄取的定性与定量分析。

主要发现

  • PET-MRI各重建的主动脉钠[18F]氟化物摄取视觉上与PET-CT可比。
  • PET-CT主动脉总标准化摄取值与各PET-MRI衰减校正法强相关:Dixon R=0.70;RadialVIBE-2 R=0.63;RadialVIBE-4 R=0.64;均p<0.001。
  • Dixon与RadialVIBE-4重建可检出软组织与肺之间的呼吸相关伪影,RadialVIBE-2未检出;该伪影邻近心房时导致血池活动估计变化。
  • PET-CT与PET-MRI放射性示踪剂活动的定量一致性以RadialVIBE-2最为一致。

局限性

  • 原文指出PET-MRI重建中组织类型分割仍存挑战,需要改进衰减校正方法。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

INTRODUCTION: Non-invasive detection of pathological changes in thoracic aortic disease remains an unmet clinical need particularly for patients with congenital heart disease. Positron emission tomography combined with magnetic resonance imaging (PET-MRI) could provide a valuable low-radiation method of aortic surveillance in high-risk groups. Quantification of aortic microcalcification activity using sodium [18F]fluoride holds promise in the assessment of thoracic aortopathies. We sought to evaluate aortic sodium [18F]fluoride uptake in PET-MRI using three methods of attenuation correction compared to positron emission tomography computed tomography (PET-CT) in patients with bicuspid aortic valve, METHODS: Thirty asymptomatic patients under surveillance for bicuspid aortic valve disease underwent sodium [18F]fluoride PET-CT and PET-MRI of the ascending thoracic aorta during a single visit. PET-MRI data were reconstructed using three iterations of attenuation correction (Dixon, radial gradient recalled echo with two [RadialVIBE-2] or four [RadialVIBE-4] tissue segmentation). Images were qualitatively and quantitatively analysed for aortic sodium [18F]fluoride uptake on PET-CT and PET-MRI. RESULTS: Aortic sodium [18F]fluoride uptake on PET-MRI was visually comparable with PET-CT using each reconstruction and total aortic standardised uptake values on PET-CT strongly correlated with each PET-MRI attenuation correction method (Dixon R = 0.70; RadialVIBE-2 R = 0.63; RadialVIBE-4 R = 0.64; p < 0.001 for all). Breathing related artefact between soft tissue and lung were detected using Dixon and RadialVIBE-4 but not RadialVIBE-2 reconstructions, with the presence of this artefact adjacent to the atria leading to variations in blood pool activity estimates. Consequently, quantitative agreements between radiotracer activity on PET-CT and PET-MRI were most consistent with RadialVIBE-2. CONCLUSION: Ascending aortic microcalcification analysis in PET-MRI is feasible with comparable findings to PET-CT. RadialVIBE-2 tissue attenuation correction correlates best with the reference standard of PET-CT and is less susceptible to artefact. There remain challenges in segmenting tissue types in PET-MRI reconstructions, and improved attenuation correction methods are required.

来源:Eur J Nucl Med Mol Imaging,查看原文。版权归原作者所有。

引用

  1. Nash J, Debono S, Whittington B, et al. Thoracic aortic microcalcification activity in combined positron emission tomography and magnetic resonance imaging. Eur J Nucl Med Mol Imaging. 2024. PMID: 38456972. doi:10.1007/s00259-024-06670-5