18F-FDG全身PET/CT短采集时间识别儿童癫痫灶的有效性

The efficacy of short acquisition time using 18F-FDG total-body PET/CT for the identification of pediatric epileptic foci.
PubMed 文献 · EJNMMI Res · 2024-02-26

回顾性纳入25例接受全身PET/CT的儿童癫痫患者,18F-FDG给药剂量3.7 MBq/kg、采集时间600秒,通过截断列表模式数据模拟60秒、150秒和300秒短采集。结果显示300秒与150秒主观图像质量无显著差异(P>0.05),60秒时病灶检出率仍为100%,提示全身PET/CT采集时间可缩短至60秒且病灶检出率可接受,150秒足以达到诊断性能和图像质量。

要点

  • 300秒与150秒采集的主观PET图像质量无显著差异(P>0.05)。
  • 60秒采集的总体图像质量评分3.21±0.46、病灶可见度评分4.08±0.74,均大于3。
  • 25例患者26个癫痫灶在采集时间缩短至60秒时病灶检出率仍为100%。

核心结论

  • 本研究提出全身PET/CT采集时间可缩短至60秒且病灶检出率可接受。
  • 研究提示150秒采集时间足以达到儿童癫痫患者的诊断性能和图像质量。

研究设计与样本

  • 回顾性纳入25例接受全身PET/CT的儿童癫痫患者,使用uEXPLORER扫描仪,18F-FDG给药剂量3.7 MBq/kg,采集时间600秒。
  • 通过截断列表模式PET数据模拟60秒、150秒和300秒短采集时间以降低计数密度。
  • 主观图像质量采用5分制评分,对疑似癫痫灶、对侧对应部位及健康小脑皮质行感兴趣区分析,比较短时图像与600秒图像的半定量摄取指标及癫痫灶检出率。

主要发现

  • 300秒与150秒采集的主观PET图像质量无显著差异(P>0.05)。
  • 60秒采集的总体图像质量评分为3.21±0.46、病灶可见度评分为4.08±0.74,均大于3。
  • 随着采集时间缩短,小脑皮质SUVmax和SD的变化逐渐增大(P<0.01)。
  • 各组间不对称指数(AI)差异无显著差异,所有组中癫痫灶的AI均>15%。
  • 25例患者26个癫痫灶在采集时间缩短至60秒时病灶检出率仍为100%。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: 18F-FDG positron emission tomography (PET) plays a crucial part in the evaluation for pediatric epileptic patients prior to therapy. Short-term scanning holds significant importance, especially for pediatrics epileptic individuals who exhibited involuntary movements. The aim was to evaluate the effects of short acquisition time on image quality and lesion detectability in pediatric epileptic patients using total-body (TB) PET/CT. A total of 25 pediatric patients who underwent TB PET/CT using uEXPLORER scanner with an 18F-FDG administered dose of 3.7 MBq/kg and an acquisition time of 600 s were retrospectively enrolled. Short acquisition times (60 s, 150 and 300 s) were simulated by truncating PET data in list mode to reduce count density. Subjective image quality was scored on a 5-point scale. Regions of interest analysis of suspected epileptogenic zones (EZs), corresponding locations contralateral to EZs, and healthy cerebellar cortex were used to compare the semi-quantitative uptake indices of short-time images and then were compared with 600 s images. The comparison of EZs detectability based on time-dependent PET images was performed. RESULTS: Our study demonstrated that a short acquisition time of 150 s is sufficient to maintain subjective image quality and lesion significance. Statistical analysis revealed no significant difference in subjective PET image quality between imaging at 300 s and 150 s (P > 0.05). The overall impression scores of image quality and lesion conspicuity in G60s were both greater than 3 (overall quality, 3.21 ± 0.46; lesion conspicuity, 4.08 ± 0.74). As acquisition time decreased, the changes of SUVmax and SD in the cerebellar cortex gradually increased (P < 0.01). There was no significant difference in asymmetry index (AI) difference between the groups and the AIs of EZs were > 15% in all groups. In 26 EZs of 25 patients, the lesion detection rate was still 100% when the time was reduced to 60 s. CONCLUSIONS: This study proposed that TB PET/CT acquisition time could be reduced to 60 s with acceptable lesion detectability. Furthermore, it was suggested that a 150 s acquisition time would be sufficient to achieve diagnostic performance and image quality for children with epilepsy.