文献速递肿瘤 PET/CT放射性药物与辐射防护

利用全身PET成像减轻SUV的不确定性

Mitigating SUV uncertainties using total body PET imaging.
PubMed 文献 · Eur J Nucl Med Mol Imaging · 2023-11-13

研究基于长轴视野(LAFOV)PET/CT系统,用25例患者扫描建立回归拟合,从全身(TB)和半身(HB)图像推导18F-FDG活度和体重,并在56例独立扫描中验证。结果显示推导值与报告值高度一致,图像推导值可降低肝脏和病灶SUV变异性并纠正病灶SUV误差,作者认为应将其纳入质控以获得更可靠、可重复的定量摄取测量。

要点

  • 从全身和半身18F-FDG PET图像可准确推导注射活度和体重
  • 图像推导值可降低肝脏SUVmean和病灶SUVpeak的变异性
  • 作者认为图像推导值应作为质控纳入,以提高定量摄取测量的可靠性

核心结论

  • 18F-FDG活度和体重可从全身和半身扫描中准确推导,图像推导值改善了SUV精密度并纠正了病灶SUV误差。
  • 作者认为图像推导值应作为质控纳入,以生成更可靠、可重复的定量摄取测量。

研究设计与样本

  • 研究使用LAFOV PET/CT系统,以81例患者扫描中的25例确定推导图像活度和体重的回归拟合。
  • 将回归拟合应用于同一扫描仪的56例独立18F-FDG PET扫描,以确定能否从全身和半身扫描准确获得注射活度和体重。
  • 研究图像推导值对肝脏SUVmean和病灶SUVpeak精密度的影响。
  • 从短轴视野(SAFOV)PET/CT系统获取20例扫描,以确定回归拟合是否也适用于SAFOV系统的半身扫描。

主要发现

  • 全身和半身18F-FDG活度及体重均显著预测报告的注射活度(r=0.999;r=0.984)和体重(r=0.999;r=0.987)。
  • 应用回归拟合后,从全身扫描准确推导的18F-FDG活度和体重误差分别为4.8%和3.2%,从半身扫描分别为4.9%和3.1%。
  • 与报告值相比,图像推导值降低了肝脏和病灶SUV的变异性。
  • 从SAFOV扫描仪获得的18F-FDG活度和体重推导误差分别为6.7%和4.5%。

临床怎么用

  • 作者认为图像推导值应作为质控纳入,以生成更可靠、可重复的定量摄取测量。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: Standardised uptake values (SUV) are commonly used to quantify 18F-FDG lesion uptake. However, SUVs may suffer from several uncertainties and errors. Long-axial field-of-view (LAFOV) PET/CT systems might enable image-based quality control (QC) by deriving 18F-FDG activity and weight from total body (TB) 18F-FDG PET images. In this study, we aimed to develop these image-based QC to reduce errors and mitigate SUV uncertainties. METHODS: Twenty-five out of 81 patient scans from a LAFOV PET/CT system were used to determine regression fits for deriving of image-derived activity and weight. Thereafter, the regression fits were applied to 56 independent 18F-FDG PET scans from the same scanner to determine if injected activity and weight could be obtained accurately from TB and half-body (HB) scans. Additionally, we studied the impact of image-based values on the precision of liver SUVmean and lesion SUVpeak. Finally, 20 scans were acquired from a short-axial field-of-view (SAFOV) PET/CT system to determine if the regression fits also applied to HB scans from a SAFOV system. RESULTS: Both TB and HB 18F-FDG activity and weight significantly predicted reported injected activity (r = 0.999; r = 0.984) and weight (r = 0.999; r = 0.987), respectively. After applying the regression fits, 18F-FDG activity and weight were accurately derived within 4.8% and 3.2% from TB scans and within 4.9% and 3.1% from HB, respectively. Image-derived values also mitigated liver and lesion SUV variability compared with reported values. Moreover, 18F-FDG activity and weight obtained from a SAFOV scanner were derived within 6.7% and 4.5%, respectively. CONCLUSION: 18F-FDG activity and weight can be derived accurately from TB and HB scans, and image-derived values improved SUV precision and corrected for lesion SUV errors. Therefore, image-derived values should be included as QC to generate a more reliable and reproducible quantitative uptake measurement.

来源:Eur J Nucl Med Mol Imaging,查看原文。版权归原作者所有。

引用

  1. Smith CLC, Zwezerijnen GJC, den Hollander ME, et al. Mitigating SUV uncertainties using total body PET imaging. Eur J Nucl Med Mol Imaging. 2023. PMID: 37953391. doi:10.1007/s00259-023-06503-x

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。