使用PLANETDose进行放射性药物治疗剂量学的质量保证考量:一项国际原子能机构研究

Quality Assurance Considerations in Radiopharmaceutical Therapy Dosimetry Using PLANETDose: An International Atomic Energy Agency Study.
PubMed 文献 · J Nucl Med · 2024-01-02

该研究基于南非Tygerberg医院一名接受[177Lu]Lu-DOTATATE治疗患者的SPECT/CT数据集,在9个中心使用统一协议和PLANETDose(3.1.1版)软件对3个治疗周期进行剂量学分析。结果显示,肝脏、肾脏和病灶的初始剂量学结果在操作者间存在相当大的差异,即使经过多轮培训和反馈,仍观察到显著差异。作者认为,强化培训、设立中间检查点、进行合理性核查以及交叉验证结果,可减少或避免人为误差导致的变异。

要点

  • 同一患者数据集用同一剂量学流程和软件分析,操作者间结果仍存在显著差异。
  • 肝脏、肾脏和病灶的初始剂量学结果表现出相当大的操作者间变异。
  • 需通过中间检查点区分用户失误与合理的操作者差异,以推动临床剂量学的质量保证。

核心结论

  • 使用同一患者数据集、同一剂量学流程和软件时,尽管经过多轮培训和反馈,结果仍存在显著差异。
  • 人为误差导致的变异可通过强化培训、设立中间检查点、合理性核查及跨物理师或标准化数据集的交叉验证来减少或避免。

研究设计与样本

  • 对国际原子能机构协调研究项目E23005中南非Tygerberg医院一名接受[177Lu]Lu-DOTATATE治疗患者的SPECT/CT数据集进行结构化分析。
  • 注射后5个时间点采集SPECT/CT图像,患者与校准图像在工作站重建,独立得出校准因子122.6 Bq/count并提供给参与者。
  • 定义标准剂量学协议,在9个中心安装PLANETDose(3.1.1版)软件,对3个治疗周期进行剂量学计算。
  • 协议包括刚性图像配准、分割(器官半手动,肿瘤用活度阈值)及剂量体素核卷积活度,随后积分吸收剂量率以获得吸收剂量。
  • 参与者个人及集体培训中反复执行协议,分析剂量学变异、质量保证与误差,并建立中间检查点以区分用户失误与合理的操作者差异。

主要发现

  • 肝脏、肾脏和病灶的初始剂量学结果表现出相当大的操作者间变异。
  • 除总计数、体积和活度等中间检查点外,还需活度与计数比、活度浓度、吸收剂量率与活度浓度比来确定变异来源。
  • 多次培训和反馈后,使用同一患者数据集、同一剂量学流程和软件分析,结果仍存在显著差异。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Implementation of radiopharmaceutical therapy dosimetry varies depending on the clinical application, dosimetry protocol, software, and ultimately the operator. Assessing clinical dosimetry accuracy and precision is therefore a challenging task. This work emphasizes some pitfalls encountered during a structured analysis, performed on a single-patient dataset consisting of SPECT/CT images by various participants using a standard protocol and clinically approved commercial software. Methods: The clinical dataset consisted of the dosimetric study of a patient administered with [177Lu]Lu-DOTATATE at Tygerberg Hospital, South Africa, as a part of International Atomic Energy Agency-coordinated research project E23005. SPECT/CT images were acquired at 5 time points postinjection. Patient and calibration images were reconstructed on a workstation, and a calibration factor of 122.6 Bq/count was derived independently and provided to the participants. A standard dosimetric protocol was defined, and PLANETDose (version 3.1.1) software was installed at 9 centers to perform the dosimetry of 3 treatment cycles. The protocol included rigid image registration, segmentation (semimanual for organs, activity threshold for tumors), and dose voxel kernel convolution of activity followed by absorbed dose (AD) rate integration to obtain the ADs. Iterations of the protocol were performed by participants individually and within collective training, the results of which were analyzed for dosimetric variability, as well as for quality assurance and error analysis. Intermediary checkpoints were developed to understand possible sources of variation and to differentiate user error from legitimate user variability. Results: Initial dosimetric results for organs (liver and kidneys) and lesions showed considerable interoperator variability. Not only was the generation of intermediate checkpoints such as total counts, volumes, and activity required, but also activity-to-count ratio, activity concentration, and AD rate-to-activity concentration ratio to determine the source of variability. Conclusion: When the same patient dataset was analyzed using the same dosimetry procedure and software, significant disparities were observed in the results despite multiple sessions of training and feedback. Variations due to human error could be minimized or avoided by performing intensive training sessions, establishing intermediate checkpoints, conducting sanity checks, and cross-validating results across physicists or with standardized datasets. This finding promotes the development of quality assurance in clinical dosimetry.