文献速递肿瘤 PET/CT放射性药物与辐射防护

新一代数字锗酸铋PET/CT系统Omni Legend 32按照NEMA NU 2-2018标准的性能特征

Performance Characteristics of a New-Generation Digital Bismuth Germanium Oxide PET/CT System, Omni Legend 32, According to NEMA NU 2-2018 Standards.
PubMed 文献 · J Nucl Med · 2023-12-01

本研究按照美国国家电气制造商协会(NEMA)NU 2-2018标准,评估基于硅光电倍增管(SiPM)和锗酸铋晶体的Omni Legend 32 PET/CT系统的空间分辨率、灵敏度、计数率性能和图像质量,并展示2例患者图像。结果显示该系统在全身PET系统中具有同类领先的灵敏度和计数率,空间分辨率与其他当前SiPM-based PET/CT系统大致相当,图像质量良好。

要点

  • Omni Legend 32轴向视野32 cm,采用锗酸铋晶体与SiPM探测器。
  • NEMA灵敏度为47.30 cps/kBq(轴向中心)和47.05 cps/kBq(径向偏移10 cm)。
  • 峰值噪声等效计数率为501.7 kcps(15.7 kBq/mL)。

核心结论

  • Omni Legend 32在全身PET系统中具有同类领先的灵敏度和计数率,空间分辨率与其他当前SiPM-based PET/CT系统大致相当,图像质量良好。

研究设计与样本

  • 按照NEMA NU 2-2018标准评估Omni Legend 32 PET/CT系统的空间分辨率、灵敏度、计数率性能和图像质量。
  • 系统PET组件包含22个探测器模块,每个模块含24个探测器块,每个探测器块含72个体积为4.1×4.1×30 mm的锗酸铋晶体,耦合18个面积为6×6 mm的SiPM器件,轴向视野32 cm。
  • 另对2例患者的PET图像进行视觉评估,结合Precision DL。

主要发现

  • 距中心轴1、10、20 cm的平均空间分辨率在滤波反投影下分别为4.3、5.3、6.2 mm,在有序子集期望最大化下分别为3.7、4.3、5.1 mm。
  • NEMA灵敏度在视野轴向中心为47.30 cps/kBq,在径向偏移10 cm处为47.05 cps/kBq。
  • 在15.7 kBq/mL时,散射分数为35.4%,计数率准确性为1.7%,峰值噪声等效计数率为501.7 kcps。
  • NEMA体模中从最小到最大球体的对比度恢复范围为61.3%至93.0%,背景变异度为5.4%至11.7%,Q.Clear(β值50)的肺误差为5.1%。
  • 使用Omni Legend 32获得了良好的患者图像质量。

关键数值与推荐

  • 空间分辨率:滤波反投影下距中心轴1、10、20 cm分别为4.3、5.3、6.2 mm;有序子集期望最大化下分别为3.7、4.3、5.1 mm。
  • NEMA灵敏度:视野轴向中心47.30 cps/kBq,径向偏移10 cm处47.05 cps/kBq。
  • 散射分数35.4%、计数率准确性1.7%、峰值噪声等效计数率501.7 kcps(15.7 kBq/mL)。
  • 对比度恢复61.3%至93.0%,背景变异度5.4%至11.7%,Q.Clear(β值50)肺误差5.1%。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

The Omni Legend 32 PET/CT system features silicon photomultiplier (SiPM)-based detectors with bismuth germanium oxide crystals and a 32-cm axial field of view (FOV). The present study aimed to determine the performance characteristics of the Omni Legend 32 PET/CT system according to National Electrical Manufacturers Association (NEMA) NU 2-2018 standards. Methods: The PET component of this system comprises 22 detector modules; each module contains 24 detector blocks with 72 bismuth germanium oxide crystals with a volume of 4.1 × 4.1 × 30 mm coupled to 18 SiPM devices with a 6 × 6 mm area, resulting in an axial FOV of 32 cm. The spatial resolution, sensitivity, count rate performance, and image quality delivered by PET were evaluated using the NEMA NU 2-2018 standard. PET images of 2 patients were evaluated to get a visual first impression of the Omni Legend 32 PET/CT system together with Precision DL. Results: The average spatial resolution at 1, 10, and 20 cm from the central axis was 4.3, 5.3, and 6.2 mm, respectively, for filtered backprojection and 3.7, 4.3, and 5.1 mm, respectively, for ordered-subset expectation maximization. The NEMA sensitivity was 47.30 and 47.05 cps/kBq at the axial center of the FOV and at a 10-cm radial offset, respectively. The scatter fraction, count rate accuracy, and peak noise-equivalent count rates were 35.4%, 1.7%, and 501.7 kcps, respectively, at 15.7 kBq/mL. Contrast recovery for the NEMA body phantom from the smallest to the largest sphere ranged from 61.3% to 93.0%, with a background variability of 5.4%-11.7% and a lung error of 5.1% for Q.Clear (β-value, 50). Good patient image quality was obtained with the Omni Legend 32. Conclusion: The Omni Legend 32 has class-leading sensitivity and count rates within the category of whole-body PET systems while maintaining spatial resolution broadly comparable to that of other current SiPM-based PET/CT systems. This combination of properties results in a very good image quality.

来源:J Nucl Med,查看原文。版权归原作者所有。

引用

  1. Yamagishi S, Miwa K, Kamitaki S, et al. Performance Characteristics of a New-Generation Digital Bismuth Germanium Oxide PET/CT System, Omni Legend 32, According to NEMA NU 2-2018 Standards. J Nucl Med. 2023. PMID: 37857503. doi:10.2967/jnumed.123.266140

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。