人工智能驱动的单次PET图像伪影检测与解缠:迈向常规临床图像质量保证
本研究提出并验证了PET图像质量保证网络(PET-QA-NET),可在无需参考标准或金标准的前提下检测和分离图像伪影。网络用669/100/100例无伪影肿瘤18F-FDG PET/CT图像训练、验证和测试,并在8个PET中心的384例扫描上微调与评估;另对200例含伪影扫描由2名核医学医师盲法评估,显示PET-QA-NET较常规CT衰减散射校正有改善图像质量、提高诊断信心和减少伪影的总体趋势。
要点
- PET-QA-NET可在无参考标准或金标准条件下检测并分离18F-FDG PET图像伪影。
- 100例患者各22个感兴趣区中,SUVmean、SUVmax、SUVpeak的MAE分别为0.13±0.04、0.24±0.1、0.21±0.06,无统计学显著差异。
- 2名核医学医师盲法评估200例含伪影扫描,PET-QA-NET较常规CT衰减散射校正呈改善图像质量、提高诊断信心和减少伪影的总体趋势。
核心结论
- 作者认为已开发出一种高效可靠的质量保证工具,可常规嵌入临床用于检测和校正18F-FDG PET图像伪影,并显著改善PET图像质量和定量能力。
研究设计与样本
- 研究为深度学习网络PET-QA-NET的开发与评估,训练/验证/测试集由669/100/100例无伪影肿瘤18F-FDG PET/CT图像组成。
- 网络随后在来自8个不同PET中心的384例扫描(20%用于微调)上微调并评估。
- 另对200例含伪影的18F-FDG PET/CT扫描(分别用常规CT衰减散射校正和PET-QA-NET生成)由2名核医学医师盲法评估伪影存在、诊断信心、图像质量和不同身体区域检出病灶数。
主要发现
- 100例患者各22个感兴趣区中,SUVmean、SUVmax、SUVpeak的MAE分别为0.13±0.04、0.24±0.1、0.21±0.06,无统计学显著差异。
- 定性评估显示,与常规CT衰减散射校正相比,PET-QA-NET呈改善图像质量和诊断信心、减少图像伪影的总体趋势。
关键数值与推荐
- 100例患者各22个感兴趣区中,SUVmean的MAE为0.13±0.04。
- 100例患者各22个感兴趣区中,SUVmax的MAE为0.24±0.1。
- 100例患者各22个感兴趣区中,SUVpeak的MAE为0.21±0.06。
- 上述SUVmean、SUVmax、SUVpeak的MAE之间无统计学显著差异。
临床怎么用
- 作者认为该质量保证工具可常规嵌入临床,用于检测和校正18F-FDG PET图像伪影,并改善图像质量和定量能力。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: Medical imaging artifacts compromise image quality and quantitative analysis and might confound interpretation and misguide clinical decision-making. The present work envisions and demonstrates a new paradigm PET image Quality Assurance NETwork (PET-QA-NET) in which various image artifacts are detected and disentangled from images without prior knowledge of a standard of reference or ground truth for routine PET image quality assurance. METHODS: The network was trained and evaluated using training/validation/testing data sets consisting of 669/100/100 artifact-free oncological 18 F-FDG PET/CT images and subsequently fine-tuned and evaluated on 384 (20% for fine-tuning) scans from 8 different PET centers. The developed DL model was quantitatively assessed using various image quality metrics calculated for 22 volumes of interest defined on each scan. In addition, 200 additional 18 F-FDG PET/CT scans (this time with artifacts), generated using both CT-based attenuation and scatter correction (routine PET) and PET-QA-NET, were blindly evaluated by 2 nuclear medicine physicians for the presence of artifacts, diagnostic confidence, image quality, and the number of lesions detected in different body regions. RESULTS: Across the volumes of interest of 100 patients, SUV MAE values of 0.13 ± 0.04, 0.24 ± 0.1, and 0.21 ± 0.06 were reached for SUV mean , SUV max , and SUV peak , respectively (no statistically significant difference). Qualitative assessment showed a general trend of improved image quality and diagnostic confidence and reduced image artifacts for PET-QA-NET compared with routine CT-based attenuation and scatter correction. CONCLUSION: We developed a highly effective and reliable quality assurance tool that can be embedded routinely to detect and correct for 18 F-FDG PET image artifacts in clinical setting with notably improved PET image quality and quantitative capabilities.