回顾性门控心肌灌注成像及衰减、散射、呼吸、运动和心律失常的综合校正
本研究比较了原型xSPECT心脏(xSC)与常规SPECT(Flash3D)在56例疑似冠心病患者静息态下的图像,xSC采用列表模式采集、连续旋转和回顾性心电门控迭代重建,并综合校正衰减、散射、运动等。结果显示xSC图像质量均可接受,极化图分布更均匀,心肌梗死缺损对比度更高(p=0.0004),回顾性门控平均使用98%的心搏,排除2%的心搏。结论认为基于列表模式的xSC原型进行定量心肌灌注成像可行。
要点
- xSC原型采用列表模式采集、连续旋转和回顾性心电门控迭代重建,并综合校正衰减、散射及运动。
- xSC图像质量均可接受,极化图分布较Flash3D更均匀,心肌梗死缺损对比度更高(p=0.0004)。
- 回顾性门控平均使用98%的心搏,排除2%的心搏;xSC计数数据较前瞻性门控等效数据集高18%。
核心结论
- 基于列表模式的xSPECT心脏原型进行定量心肌灌注成像可行,图像质量至少可接受。
研究设计与样本
- 研究比较了原型xSC与常规SPECT(Flash3D)在56例年龄71±12岁疑似冠心病患者静息态下的图像。
主要发现
- 所有xSC图像质量均可满足临床用途。
- xSC极化图分布较Flash3D更均匀,心尖计数更低,心肌梗死缺损对比度更高(p=0.0004)。
- 壁运动、16门控容积曲线和射血分数至少可接受,并有改善迹象。
- 临床前瞻性门控方法在6例患者中剔除≥20%的心搏,而回顾性门控平均使用98%的心搏,排除2%的心搏。
- 利用列表模式数据创建的前瞻性门控等效数据集显示,xSC方法生成的计数数据高18%,图像噪声更少,容积和LVEF等功能变量相当(p=ns)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Absolute quantitative myocardial perfusion SPECT requires addressing of aleatory and epistemic uncertainties in conjunction with providing image quality sufficient for lesion detection and characterization. Iterative reconstruction methods enable the mitigation of the root causes of image degradation. This study aimed to determine the feasibility of a new SPECT/CT method with integrated corrections attempting to enable absolute quantitative cardiac imaging (xSPECT Cardiac; xSC). METHODS: We compared images of prototype xSC and conventional SPECT (Flash3DTM) acquired at rest from 56 patients aged 71 ± 12 y with suspected coronary heart disease. The xSC prototype comprised list-mode acquisitions with continuous rotation and subsequent iterative reconstructions with retrospective electrocardiography (ECG) gating. Besides accurate image formation modeling, patient-specific CT-based attenuation and energy window-based scatter correction, additionally we applied mitigation for patient and organ motion between views (inter-view), and within views (intra-view) for both the gated and ungated reconstruction. We then assessed image quality, semiquantitative regional values, and left ventricular function in the images. RESULTS: The quality of all xSC images was acceptable for clinical purposes. A polar map showed more uniform distribution for xSC compared with Flash3D, while lower apical count and higher defect contrast of myocardial infarction (p = 0.0004) were observed on xSC images. Wall motion, 16-gate volume curve, and ejection fraction were at least acceptable, with indication of improvements. The clinical prospectively gated method rejected beats ≥20% in 6 patients, whereas retrospective gating used an average of 98% beats, excluding 2% of beats. We used the list-mode data to create a product equivalent prospectively gated dataset. The dataset showed that the xSC method generated 18% higher count data and images with less noise, with comparable functional variables of volume and LVEF (p = ns). CONCLUSIONS: Quantitative myocardial perfusion imaging with the list-mode-based prototype xSPECT Cardiac is feasible, resulting in images of at least acceptable image quality.
引用
- Nakajima K, Shibutani T, Massanes F, et al. Myocardial perfusion imaging with retrospective gating and integrated correction of attenuation, scatter, respiration, motion, and arrhythmia. J Nucl Cardiol. 2023. PMID: 37758961. doi:10.1007/s12350-023-03374-5
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。