评估已知透壁瘢痕区域静息心肌血流量以确认三维心脏正电子发射断层扫描的准确性和精密度

Assessment of resting myocardial blood flow in regions of known transmural scar to confirm accuracy and precision of 3D cardiac positron emission tomography.
PubMed 文献 · EJNMMI Res · 2023-09-27

本研究在当代三维PET-CT系统上使用82Rb,评估6种软件包在已知透壁心肌瘢痕(TMS)区域测量静息心肌血流量(rMBF)的准确性和精密度。结果显示各软件包低端准确性差异很大,HeartSee软件准确(0.26 mL/min/g,与既定值一致)且精密度高(变异系数COV=0.07),Emory-Ottawa次之(0.37 mL/min/g),而4DM、4DM-FVD、Cedars和Emory-Votaw持续给出高于既定金标准的值且扫描间变异性更高。

要点

  • 在已知TMS区域,HeartSee软件测得的rMBF中位数为0.26 mL/min/g,与既定金标准值~0.30 mL/min/g或更低一致,COV=0.07。
  • Emory-Ottawa软件准确性次之,rMBF中位数为0.37 mL/min/g,但在约5%的静息扫描中rMBF高于既定金标准值。
  • 4DM、4DM-FVD、Cedars和Emory-Votaw软件测得的rMBF中位数分别为0.71、0.41、0.66、0.51 mL/min/g,均高于既定金标准,且扫描间变异性更高(COV分别为0.16、0.11、0.11、0.09)。

核心结论

  • 基于已知TMS区域rMBF测量,各软件包低端准确性差异很大;使用82Rb的三维PET和HeartSee软件能准确(0.26 mL/min/g,与既定值一致)且精密(COV=0.07)地定量TMS区域的rMBF。

研究设计与样本

  • 在当代三维PET-CT系统上使用82Rb,评估目前可用的软件包能否准确且精密地测量已知TMS区域的rMBF。
  • 临床试验注册号NCT05286593,注册日期2021年12月28日。

主要发现

  • TMS区域rMBF中位数(mL/min/g)和COV:4DM为0.71(IQR 0.52-1.02)和0.16;4DM-FVD为0.41(IQR 0.34-0.54)和0.10;Cedars为0.66(IQR 0.51-0.85)和0.11;Emory-Votaw为0.51(IQR 0.43-0.61)和0.08;Emory-Ottawa为0.37(IQR 0.30-0.42)和0.07;HeartSee为0.26(IQR 0.23-0.32)和COV 0.07。
  • Emory-Ottawa软件准确性次之(0.37 mL/min/g),但在约5%的静息扫描中rMBF高于既定金标准值。
  • 4DM、4DM-FVD、Cedars和Emory-Votaw软件持续给出高于既定金标准的值(分别为0.71、0.41、0.66、0.51 mL/min/g),且扫描间变异性更高(分别为0.16、0.11、0.11、0.09)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Composite invasive and non-invasive data consistently demonstrate that resting myocardial blood flow (rMBF) in regions of known transmural myocardial scar (TMS) converge on a value of ~ 0.30 mL/min/g or lower. This value has been confirmed using the 3 most common myocardial perfusion agents (13N, 15O-H2O and 82Rb) incorporating various kinetic models on older 2D positron emission tomography (PET) systems. Thus, rMBF in regions of TMS can serve as a reference "truth" to evaluate low-end accuracy of various PET systems and software packages (SWPs). Using 82Rb on a contemporary 3D-PET-CT system, we sought to determine whether currently available SWP can accurately and precisely measure rMBF in regions of known TMS. RESULTS: Median rMBF (in mL/min/g) and COV in regions of TMS were 0.71 [IQR 0.52-1.02] and 0.16 with 4DM; 0.41 [0.34-0.54] and 0.10 with 4DM-FVD; 0.66 [0.51-0.85] and 0.11 with Cedars; 0.51 [0.43-0.61] and 0.08 with Emory-Votaw; 0.37 [0.30-0.42], 0.07 with Emory-Ottawa, and 0.26 [0.23-0.32], COV 0.07 with HeartSee. CONCLUSIONS: SWPs varied widely in low end accuracy based on measurement of rMBF in regions of known TMS. 3D PET using 82Rb and HeartSee software accurately (0.26 mL/min/g, consistent with established values) and precisely (COV = 0.07) quantified rMBF in regions of TMS. The Emory-Ottawa software yielded the next-best accuracy (0.37 mL/min/g), though rMBF was higher than established gold-standard values in ~ 5% of the resting scans. 4DM, 4DM-FDV, Cedars and Emory-Votaw SWP consistently resulted values higher than the established gold standard (0.71, 0.41, 0.66, 0.51 mL/min/g, respectively), with higher interscan variability (0.16, 0.11, 0.11, and 0.09, respectively). TRIAL REGISTRATION: clinicaltrial.gov, NCT05286593, Registered December 28, 2021, https://clinicaltrials.gov/ct2/show/NCT05286593 .

来源:EJNMMI Res,查看原文。版权归原作者所有。