用于大鼠同步PET扫描的3D打印双支架系统:设计及其对定量分析的影响

3D-printed dual holder system for simultaneous rat PET scanning: design and influence on quantification.
PubMed 文献 · EJNMMI Res · 2023-09-07

本研究设计了一种低成本3D打印双支架系统,用于在INVEON小动物PET中同步动态扫描两只大鼠,并评估其对脑图像定量的影响。体模实验显示单个体模与两个体模同时测量的浓度准确度分别为5.5%和5.7%,体内分布无差异;全脑ROI定量变异性单扫为4.4%、双扫为3.5%,重测偏差单扫8%、双扫5%,组内相关系数单扫0.969 vs 双扫0.966。结论为双鼠同步扫描不影响定量,可降低协议成本与时长。

要点

  • 3D打印双支架支持两只大鼠同时进行动态PET扫描
  • 体模浓度测量准确度单个体模5.5%、两个体模同时5.7%
  • 全脑ROI定量变异性单扫4.4% vs 双扫3.5%,重测偏差单扫8% vs 双扫5%

核心结论

  • 在INVEON中两只大鼠同步扫描不影响定量分析,双支架系统可降低协议成本与时长。

研究设计与样本

  • 设计低成本3D打印双支架系统,用于两只仰卧大鼠的同步动态PET扫描,并研究其对脑图像定量的影响。

主要发现

  • 体模浓度测量准确度:视野内单个体模为5.5%,两个体模同时测量为5.7%。
  • 体模间浓度比值误差:单独上位6.7%±5.1%,单独下位6.7%±3.7%,双位上位5.9%±4.3%,双位下位7.4%±6%;分离测量为6.7%,同时测量为6.6%。
  • 体内分布曲线显示单独与双鼠扫描的摄取无差异。
  • 全脑感兴趣区定量变异性:单独扫描4.4%,双鼠扫描3.5%。
  • 重测偏差:单独扫描8%,双鼠扫描5%;组内相关系数单独扫描0.969 vs 双鼠扫描0.966,两者相当。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The low throughout of small animal positron emission tomography (PET) images acquisitions represents a substantial limitation. The aims of this study were (i) to design a low-cost support for simultaneous dynamic PET scanning of two lying rats and (ii) to study its impact on brain image quantification. RESULTS: Accuracy of concentration measurement was 5.5% for one phantom in the field of view, and 5.7% for two phantoms measured simultaneously. Ratio concentration between phantoms showed an error of 6.7% ± 5.1% for Solo upper position, 6.7% ± 3.7% for Solo lower position, 5.9% ± 4.3% for Duo upper position, and 7.4% ± 6% for Duo lower position 6.7% for separated measures, and 6.6% for simultaneous measures. In vivo distribution profiles showed no difference between solo and duo uptakes. Region of Interest quantification in the whole brain showed 4.4% variability solo and 3.5% duo. The quantified test-retest bias was 8% in solo and 5% in duo, and the Intraclass Correlation Coefficient was comparable in solo and duo (0.969 vs. 0.966). CONCLUSIONS: Our results showed that simultaneous scans of two rats in INVEON do not affect quantification. The dual support system will allow us to reduce protocol costs and duration.