门控小动物18F-FDG PET/CT显像中左心室容积与功能的评估:三种市售软件工具的对比研究
本研究在健康小鼠异氟烷麻醉下进行心电门控18F-FDG小动物nanoPET/CT显像,用PMOD、MIM和QGS三种软件分析左心室舒张末期容积(EDV)、收缩末期容积(ESV)、每搏输出量(SV)和射血分数(EF)。三种软件测量的心脏容积相关性良好,但Bland-Altman一致性分析显示左心室参数和心功能存在差异,EF相关性差;作者结论为三种软件均可用于小动物心脏容积评估,但因软件相关一致性差异,在小鼠研究中不能互换使用。
要点
- PMOD、MIM和QGS三种软件测量小鼠左心室EDV、ESV、SV相关性良好,但EF相关性差。
- Bland-Altman分析显示三种软件左心室参数和心功能一致性存在差异。
- 三种软件重复性佳(组内和组间相关系数均>0.9),但小鼠研究中不能互换使用。
核心结论
- 三种市售软件工具均可用于门控18F-FDG PET/CT小动物心脏容积评估,但因软件相关一致性差异,PMOD、MIM和QGS在小鼠研究中不能互换使用。
研究设计与样本
- 健康小鼠在异氟烷麻醉下使用小动物nanoPET/CT(Mediso)进行心电门控18F-FDG显像,重建门控PET图像后用PMOD、MIM和QGS三种软件分析心脏容积和功能。
主要发现
- EDV相关性:PMOD vs QGS r=0.85(p<0.001),MIM vs QGS r=0.92(p<0.001),MIM vs PMOD r=0.88(p<0.001)。
- ESV相关性:PMOD vs QGS r=0.48(p=0.07),MIM vs QGS r=0.79(p<0.001),MIM vs PMOD r=0.69(p<0.01)。
- SV相关性:PMOD vs QGS r=0.73(p<0.01),MIM vs QGS r=0.86(p<0.001),MIM vs PMOD r=0.92(p<0.001)。
- EF相关性差:PMOD vs QGS r=-0.31(p=0.26),MIM vs QGS r=0.48(p=0.07),MIM vs PMOD r=0.23(p=0.41)。
- 组内和组间相关系数均>0.9,提示PMOD、MIM和QGS用于心脏容积和功能评估的重复性良好。
临床怎么用
- 作者结论认为三种软件均可用于小动物心脏容积评估,但因软件相关一致性差异,在小鼠研究中不能互换使用。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Several software tools have been developed for gated PET imaging that use distinct algorithms to analyze tracer uptake, myocardial perfusion, and left ventricle volumes and function. Studies suggest that different software tools cannot be used interchangeably in humans. In this study, we sought to compare the left ventricular parameters in gated 18F-FDG PET/CT imaging in mice by three commercially available software tools: PMOD, MIM, and QGS. METHODS AND RESULTS: Healthy mice underwent ECG-gated 18F-FDG imaging using a small-animal nanoPET/CT (Mediso) under isoflurane narcosis. Reconstructed gates PET images were subsequently analyzed in three different software tools, and cardiac volume and function (end-diastolic (EDV), end-systolic volumes (ESV), stroke volume (SV), and ejection fraction (EF)) were evaluated. While cardiac volumes correlated well between PMOD, MIM, and QGS, the left ventricular parameters and cardiac function differed in agreement using Bland-Altman analysis. EDV in PMOD vs. QGS: r = 0.85; p < 0.001, MIM vs. QGS: r = 0.92; p < 0.001, and MIM vs. PMOD: r = 0.88; p < 0.001, showed good correlations. Correlation was also found in ESV: PMOD vs. QGS: r = 0.48; p = 0.07, MIM vs QGS: r = 0.79; p < 0.001, and MIM vs. PMOD: r = 0.69; p < 0.01. SV showed good correlations in: PMOD vs. QGS: r = 0.73; p < 0.01, MIM vs. QGS: r = 0.86; p < 0.001, and MIM vs. PMOD: r = 0.92; p < 0.001. However, EF among correlated poorly: PMOD vs. QGS: r = -0.31; p = 0.26, MIM vs. QGS: r = 0.48; p = 0.07, and MIM vs. PMOD: r = 0.23; p = 0.41. Inter-class and intra-class correlation coefficient were > 0.9 underlining repeatability in using PMOD, MIM, and QGS for cardiac volume and function assessment. CONCLUSIONS: All three commercially available software tools are feasible in small animal cardiac volume assessment in gated 18F-FDG PET/CT imaging. However, due to software-related differences in agreement analysis for cardiac volumes and function, PMOD, MIM, and QGS cannot be used interchangeably in murine research.