长轴视野与短轴视野免疫PET/CT的对比:89Zr-曲妥珠单抗的半定量评估
本研究比较了长轴视野(LAFOV)与短轴视野(SAFOV)PET/CT系统在89Zr-曲妥珠单抗免疫PET中的SUV差异,并评估LAFOV扫描时长对图像噪声和SUV指标的影响。共5例转移性乳腺癌患者先后接受两种系统扫描,另4例仅接受LAFOV扫描,注射37 MBq 89Zr-曲妥珠单抗后4天采集约30分钟。结果显示LAFOV因图像噪声较低而SUVmax和SUVpeak低于SAFOV,缩短扫描时长会增加噪声和SUV偏差,但SUVpeak在30分钟缩短至10分钟时偏差极小。
要点
- 30分钟图像中,SAFOV的病灶SUVmax较LAFOV高16.2%±13.4%,SUVpeak高10.1%±7.2%。
- 健康组织SUVmean在两种系统间差异极小(0.7%±5.8%)。
- LAFOV扫描时长从30分钟缩短至10分钟时,SUVpeak仅显示极小偏差。
核心结论
- LAFOV 89Zr免疫PET的SUVmax和SUVpeak低于SAFOV,原因是图像噪声较低。
- 缩短LAFOV扫描时长可能以增加图像噪声和SUV指标偏差为代价。
- 将扫描时长从30分钟缩短至10分钟时,SUVpeak仅显示极小偏差。
研究设计与样本
- 研究纳入5例转移性乳腺癌患者,先后在SAFOV(Biograph Vision)和LAFOV(Biograph Vision Quadra)PET/CT上扫描。
- 另4例患者仅接受LAFOV PET/CT扫描。
- 两种系统扫描时长均约30分钟,于注射37 MBq 89Zr-曲妥珠单抗后4天采集。
- LAFOV列表模式数据被重处理以获得15、10、7.5、5和3分钟的较短采集图像。
主要发现
- 30分钟图像中,SAFOV的病灶SUVmax较LAFOV高16.2%±13.4%,SUVpeak高10.1%±7.2%。
- 健康组织的SUVmean在两种系统间差异极小(0.7%±5.8%)。
- 30分钟SAFOV的肝脏变异系数介于3分钟和5分钟LAFOV之间。
- SAFOV与LAFOV之间SUVmax和SUVpeak差异最小出现在3分钟LAFOV PET。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
The purpose of this study was to quantify any differences between the SUVs of 89Zr immuno-PET scans obtained using a PET/CT system with a long axial field of view (LAFOV; Biograph Vision Quadra) compared to a PET/CT system with a short axial field of view (SAFOV; Biograph Vision) and to evaluate how LAFOV PET scan duration affects image noise and SUV metrics. Methods: Five metastatic breast cancer patients were scanned consecutively on SAFOV and LAFOV PET/CT scanners. Four additional patients were scanned using only LAFOV PET/CT. Scans on both systems lasted approximately 30 min and were acquired 4 d after injection of 37 MBq of 89Zr-trastuzumab. LAFOV list-mode data were reprocessed to obtain images acquired using shorter scan durations (15, 10, 7.5, 5, and 3 min). Volumes of interest were placed in healthy tissues, and tumors were segmented semiautomatically to compare coefficients of variation and to perform Bland-Altman analysis on SUV metrics (SUVmax, SUVpeak, and SUVmean). Results: Using 30-min images, 2 commonly used lesion SUV metrics were higher for SAFOV than for LAFOV PET (SUVmax, 16.2% ± 13.4%, and SUVpeak, 10.1% ± 7.2%), whereas the SUVmean of healthy tissues showed minimal differences (0.7% ± 5.8%). Coefficients of variation in the liver derived from 30-min SAFOV PET were between those of 3- and 5-min LAFOV PET. The smallest SUVmax and SUVpeak differences between SAFOV and LAFOV were found for 3-min LAFOV PET. Conclusion: LAFOV 89Zr immuno-PET showed a lower SUVmax and SUVpeak than SAFOV because of lower image noise. LAFOV PET scan duration may be reduced at the expense of increasing image noise and bias in SUV metrics. Nevertheless, SUVpeak showed only minimal bias when reducing scan duration from 30 to 10 min.