文献速递肿瘤 PET/CT放射性药物与辐射防护

步行式平板全身PET:采用具备DOI能力的高分辨率单片探测器,实现以患者为中心的低成本高通量成像

Walk-through flat panel total-body PET: a patient-centered design for high throughput imaging at lower cost using DOI-capable high-resolution monolithic detectors.
PubMed 文献 · Eur J Nucl Med Mol Imaging · 2023-07-19

为克服长轴视野全身PET因患者摆位耗时和系统成本高而难以普及的问题,作者提出一种双平板步行式全身PET(WT-TB-PET)设计,患者直立位成像,采用BGO或LYSO单片探测器并具备作用深度(DOI)能力。基于40例随机PET-CT扫描确定设计尺寸,并与短轴视野(26 cm)和长轴视野(106 cm)LSO PET系统比较灵敏度、组件成本和患者通量,结果显示WT-TB-PET组件成本约为106 cm长轴系统的1/3.3倍,8小时工作日最大扫描患者数可达87例。

要点

  • WT-TB-PET采用两块70 cm宽、106 cm高的平板探测器,间距50 cm,患者直立位通过,避免床旁摆位耗时。
  • 与106 cm长轴LSO PET相比,探测器面积减少1.9倍、闪烁体体积减少2.2倍,但灵敏度相当且视野内均匀空间分辨率<2 mm。
  • WT-TB-PET组件成本比106 cm长轴系统低3.3倍,仅比短轴视野PET组件成本高20%;8小时工作日最大扫描患者数从短轴视野的28例增至WT-TB-PET的87例。

核心结论

  • 作者提出一种基于单片BGO或LYSO探测器并具备DOI测量的WT-TB-PET系统,患者直立位于两块平板之间,摆脱传统床旁摆位。
  • WT-TB-PET以接近短轴视野系统的成本实现全身PET,并通过快速摆位和高灵敏度探测器提高患者通量。
  • 采用具备DOI能力的单片闪烁体可在整个视野内获得均匀的高空间分辨率(<2 mm)。

研究设计与样本

  • 本研究为系统设计与性能比较研究,基于40例随机选择的PET-CT扫描获取患者宽度、高度和深度以确定WT-TB-PET设计尺寸。
  • 将WT-TB-PET与商用Siemens Biograph Vision 600(短轴视野,26 cm)和Siemens Quadra(长轴视野,106 cm)PET-CT在组件成本、系统灵敏度和患者通量方面进行比较。
  • 系统成本比较基于估计PET系统中两种主要组件(硅光电倍增管SiPM和闪烁体)的成本;灵敏度值通过Gate蒙特卡罗模拟确定。
  • 患者通量时间(包括CT和定位扫描、床上摆位和转移)在Siemens Vision 600 PET上记录1天,用于估算WT-TB-PET、短轴视野和长轴视野PET的预期患者通量。

主要发现

  • WT-TB-PET由两块平板组成,每块70 cm宽、106 cm高,两块板之间间距50 cm;每块板由14×20个50×50×16 mm的单片BGO探测器块组成,耦合到6×6 mm SiPM阵列读出。
  • 与长轴视野PET系统相比,WT-TB-PET探测器表面积减少1.9倍、闪烁体体积减少2.2倍,同时表现出相当的灵敏度和更好的均匀空间分辨率(视野内所有方向<2 mm)。
  • WT-TB-PET估计组件成本比106 cm长轴视野系统低3.3倍,仅比短轴视野PET组件成本高20%。
  • 标准8小时工作日估计最大扫描患者数从短轴视野的28例增至长轴视野的53-60例(有限/完全验收),WT-TB-PET可达87例。
  • WT-TB-PET每例患者所需订购活度比短轴视野系统少66%。

关键数值与推荐

  • WT-TB-PET设计尺寸:每块平板70 cm宽、106 cm高,两板间距50 cm,由40例随机PET-CT扫描的患者尺寸确定。
  • 探测器单元:每块板14×20个单片BGO探测器块,尺寸50×50×16 mm,耦合6×6 mm SiPM阵列读出。
  • 与长轴视野PET相比,WT-TB-PET探测器表面积减少1.9倍、闪烁体体积减少2.2倍,灵敏度相当,视野内均匀空间分辨率<2 mm。
  • 组件成本:WT-TB-PET比106 cm长轴视野系统低3.3倍,比短轴视野PET组件成本高20%。
  • 患者通量:8小时工作日最大扫描患者数短轴视野28例、长轴视野53-60例(有限/完全验收)、WT-TB-PET 87例;WT-TB-PET每例订购活度比短轴视野少66%。

临床怎么用

  • 原文未据此给出临床推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: Long axial field-of-view (LAFOV) systems have a much higher sensitivity than standard axial field-of-view (SAFOV) PET systems for imaging the torso or full body, which allows faster and/or lower dose imaging. Despite its very high sensitivity, current total-body PET (TB-PET) throughput is limited by patient handling (positioning on the bed) and often a shortage of available personnel. This factor, combined with high system costs, makes it hard to justify the implementation of these systems for many academic and nearly all routine nuclear medicine departments. We, therefore, propose a novel, cost-effective, dual flat panel TB-PET system for patients in upright standing positions to avoid the time-consuming positioning on a PET-CT table; the walk-through (WT) TB-PET. We describe a patient-centered, flat panel PET design that offers very efficient patient throughput and uses monolithic detectors (with BGO or LYSO) with depth-of-interaction (DOI) capabilities and high intrinsic spatial resolution. We compare system sensitivity, component costs, and patient throughput of the proposed WT-TB-PET to a SAFOV (= 26 cm) and a LAFOV (= 106 cm) LSO PET systems. METHODS: Patient width, height (= top head to start of thighs) and depth (= distance from the bed to front of patient) were derived from 40 randomly selected PET-CT scans to define the design dimensions of the WT-TB-PET. We compare this new PET system to the commercially available Siemens Biograph Vision 600 (SAFOV) and Siemens Quadra (LAFOV) PET-CT in terms of component costs, system sensitivity, and patient throughput. System cost comparison was based on estimating the cost of the two main components in the PET system (Silicon Photomultipliers (SiPMs) and scintillators). Sensitivity values were determined using Gate Monte Carlo simulations. Patient throughput times (including CT and scout scan, patient positioning on bed and transfer) were recorded for 1 day on a Siemens Vision 600 PET. These timing values were then used to estimate the expected patient throughput (assuming an equal patient radiotracer injected activity to patients and considering differences in system sensitivity and time-of-flight information) for WT-TB-PET, SAFOV and LAFOV PET. RESULTS: The WT-TB-PET is composed of two flat panels; each is 70 cm wide and 106 cm high, with a 50-cm gap between both panels. These design dimensions were justified by the patient sizes measured from the 40 random PET-CT scans. Each panel consists of 14 × 20 monolithic BGO detector blocks that are 50 × 50 × 16 mm in size and are coupled to a readout with 6 × 6 mm SiPMs arrays. For the WT-TB-PET, the detector surface is reduced by a factor of 1.9 and the scintillator volume by a factor of 2.2 compared to LAFOV PET systems, while demonstrating comparable sensitivity and much better uniform spatial resolution (< 2 mm in all directions over the FOV). The estimated component cost for the WT-TB-PET is 3.3 × lower than that of a 106 cm LAFOV system and only 20% higher than the PET component costs of a SAFOV. The estimated maximum number of patients scanned on a standard 8-h working day increases from 28 (for SAFOV) to 53-60 (for LAFOV in limited/full acceptance) to 87 (for the WT-TB-PET). By scanning faster (more patients), the amount of ordered activity per patient can be reduced drastically: the WT-TB-PET requires 66% less ordered activity per patient than a SAFOV. CONCLUSIONS: We propose a monolithic BGO or LYSO-based WT-TB-PET system with DOI measurements that departs from the classical patient positioning on a table and allows patients to stand upright between two flat panels. The WT-TB-PET system provides a solution to achieve a much lower cost TB-PET approaching the cost of a SAFOV system. High patient throughput is increased by fast patient positioning between two vertical flat panel detectors of high sensitivity. High spatial resolution (< 2 mm) uniform over the FOV is obtained by using DOI-capable monolithic scintillators.

来源:Eur J Nucl Med Mol Imaging,查看原文。版权归原作者所有。

引用

  1. Vandenberghe S, Muller FM, Withofs N, et al. Walk-through flat panel total-body PET: a patient-centered design for high throughput imaging at lower cost using DOI-capable high-resolution monolithic detectors. Eur J Nucl Med Mol Imaging. 2023. PMID: 37466650. doi:10.1007/s00259-023-06341-x

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。