文献速递肿瘤 PET/CT放射性药物与辐射防护

全身PET/CT中的面部匿名化与隐私担忧

Facial Anonymization and Privacy Concerns in Total-Body PET/CT.
PubMed 文献 · J Nucl Med · 2023-06-02

研究开发并验证了一种在三维体数据中遮蔽(去面部)受试者面部的工作流,以应对共享数据时的隐私与可识别性担忧。30名健康受试者接受[18F]FDG PET和CT成像,采用FaceNet计算面部嵌入并通过聚类分析评估可识别性。去面部后CT和PET图像的正确匹配率分别从93%和最高64%(PET)降至6%和7%,去面部CT可用于PET重建的衰减校正,在靠近面部的脑皮层区域引入最大偏差-3.3%。

要点

  • 去面部后CT图像面部正确匹配率从93%降至6%。
  • 去面部后PET图像面部正确匹配率从最高64%降至7%。
  • 去面部CT用于PET衰减校正,近面部脑皮层区域最大偏差-3.3%。

核心结论

  • 作者认为所提出的方法在在线或机构间共享图像数据时提供了匿名性和谨慎性的基线,并将有助于促进合作和未来的监管合规。

研究设计与样本

  • 研究纳入30名健康受试者,在3或6个时间点接受[18F]FDG PET和CT成像,开发并验证在三维体数据中遮蔽面部的工作流。
  • 采用谷歌FaceNet计算面部嵌入,并通过聚类分析估计可识别性。

主要发现

  • CT图像渲染的面部与其他时间点CT扫描的正确匹配率为93%,去面部后降至6%。
  • PET图像渲染的面部与其他时间点PET图像的正确匹配率最高为64%,与CT图像的正确匹配率最高为50%,去面部后均降至7%。
  • 去面部CT图像可用于PET重建时的衰减校正,在靠近面部的脑皮层区域引入最大偏差-3.3%。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Total-body PET/CT images can be rendered to produce images of a subject's face and body. In response to privacy and identifiability concerns when sharing data, we have developed and validated a workflow that obscures (defaces) a subject's face in 3-dimensional volumetric data. Methods: To validate our method, we measured facial identifiability before and after defacing images from 30 healthy subjects who were imaged with both [18F]FDG PET and CT at either 3 or 6 time points. Briefly, facial embeddings were calculated using Google's FaceNet, and an analysis of clustering was used to estimate identifiability. Results: Faces rendered from CT images were correctly matched to CT scans at other time points at a rate of 93%, which decreased to 6% after defacing. Faces rendered from PET images were correctly matched to PET images at other time points at a maximum rate of 64% and to CT images at a maximum rate of 50%, both of which decreased to 7% after defacing. We further demonstrated that defaced CT images can be used for attenuation correction during PET reconstruction, introducing a maximum bias of -3.3% in regions of the cerebral cortex nearest the face. Conclusion: We believe that the proposed method provides a baseline of anonymity and discretion when sharing image data online or between institutions and will help to facilitate collaboration and future regulatory compliance.

来源:J Nucl Med,查看原文。版权归原作者所有。

引用

  1. Selfridge AR, Spencer BA, Abdelhafez YG, et al. Facial Anonymization and Privacy Concerns in Total-Body PET/CT. J Nucl Med. 2023. PMID: 37268426. doi:10.2967/jnumed.122.265280

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。