IQ-SPECT系统测量小体积心脏患者左心室容积与射血分数的心血管磁共振验证
研究纳入49例(20例正常心脏大小、29例小体积心脏)接受99mTc标记药物门控心肌灌注显像的患者,比较IQ-SPECT、常规SPECT与心血管磁共振(CMR)测量左心室射血分数(LVEF)的差异。IQ-SPECT测得的LVEF低于常规SPECT,且与CMR无显著差异;小体积心脏患者最佳重建参数为5个子集和12次迭代。
要点
- 常规SPECT在小体积心脏患者中高估LVEF,IQ-SPECT测值更接近CMR。
- 小体积心脏患者IQ-SPECT与CMR的LVEF无显著差异(62.1±7.8% vs 64.6±8.8%,P>0.05)。
- 小体积心脏最佳重建参数为5个子集和12次迭代。
核心结论
- 以CMR为标准,IQ-SPECT在小体积心脏人群中测得的LVEF比常规SPECT更可靠。
- 小体积心脏患者IQ-SPECT的最佳重建参数为5个子集和12次迭代。
研究设计与样本
- 研究纳入49例连续患者(20例正常心脏大小、29例小体积心脏),在负荷或静息状态下接受99mTc标记药物门控心肌灌注显像以评估冠状动脉疾病风险。
- 常规SPECT数据采用滤波反投影(FBP)重建,IQ-SPECT数据采用有序子集共轭梯度最小化(OSCGM)重建,并用定量门控SPECT(QGS)计算ESV、EDV和LVEF。
- 通过比较SPECT与CMR对应的LVEF测量值来确定最佳有序子集重建参数。
主要发现
- IQ-SPECT与常规SPECT测得的EDV、ESV和LVEF显著相关,但IQ-SPECT测得的EDV和ESV高于常规SPECT。
- 正常心脏大小组IQ-SPECT的LVEF低于常规SPECT(50.6±4.3% vs 73.4±8.4%,P=0.002)。
- 小体积心脏组IQ-SPECT的LVEF低于常规SPECT(62.1±7.8% vs 75.0±11.4%,P<0.001)。
- 正常心脏大小组IQ-SPECT与CMR的LVEF无显著差异(50.6±4.3% vs 53.2±5.8%,P>0.05)。
- 小体积心脏组IQ-SPECT与CMR的LVEF无显著差异(62.1±7.8% vs 64.6±8.8%,P>0.05)。
关键数值与推荐
- 小体积心脏定义:男性EDV≤60 ml且ESV≤25 ml;女性EDV≤45 ml且ESV≤20 ml。
- 小体积心脏患者5个子集和12次迭代时CMR与IQ-SPECT的LVEF无显著差异(64.6±8.8% vs 62.1±7.8%,P=0.120)。
- 正常心脏大小患者3个子集和10次迭代为最佳参数(50.6±4.3% vs 53.1±5.8%,P=0.117)。
临床怎么用
- 原文未据此给出临床推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The IQ-SPECT system is equipped with multifocal collimators and uses ordered-subset conjugate gradient minimization (OSCGM) as its reconstruction algorithm, achieving a shorter acquisition time than conventional SPECT. Left ventricular ejection fraction (LVEF) is overestimated by conventional SPECT in patients with small heart size. In this study, we compared IQ-SPECT with conventional SPECT and cardiovascular magnetic resonance (CMR) for the estimation of LVEF in patients with small hearts (males: EDV ≤ 60 ml, ESV ≤ 25 ml; females: EDV ≤ 45 ml, ESV ≤ 20 ml). METHODS: The study consisted of 49 consecutive patients (20 normal and 29 with small heart size) undergoing gated myocardial perfusion imaging (GMPI) with a 99mTc-labelled agent during stress or rest to assess the risk of coronary artery disease (CAD). The data were reconstructed using filtered back-projection (FBP) for conventional SPECT and OSCGM for IQ-SPECT. ESV, EDV, and LVEF were calculated using quantitative gated SPECT (QGS). To determine the optimal ordered-subset reconstruction parameters, we compared the LVEF from SPECT to the corresponding measurement from CMR. RESULTS: EDV, ESV, and LVEF values obtained from IQ-SPECT and conventional SPECT showed that the results of these two forms of SPECT were significantly correlated, although the EDV and ESV obtained by IQ-SPECT were higher than those obtained by conventional SPECT. IQ-SPECT yielded lower LVEF measurements than conventional SPECT (normal heart size: 50.6 ± 4.3% vs. 73.4 ± 8.4%, P = 0.002; small heart size: 62.1 ± 7.8% vs. 75.0 ± 11.4%, P < 0.001). There were no significant differences in LVEF measurements made by IQ-SPECT and CMR (normal heart size: 50.6 ± 4.3% vs. 53.2 ± 5.8%, P > 0.05; small heart size: 62.1 ± 7.8% vs. 64.6 ± 8.8%, P > 0.05). Five subsets (S) and 12 iterations (I) did not differ significantly in LVEF between CMR and IQ-SPECT for patients with small hearts (64.6 ± 8.8% vs. 62.1 ± 7.8%, P = 0.120), while 3 S and 10 I were the best parameters for patients with normal heart size (50.6 ± 4.3% vs. 53.1 ± 5.8%, P = 0.117). CONCLUSION: With CMR as the standard, IQ-SPECT yields more reliable LVEF values than conventional SPECT for populations with small heart size. The best reconstruction parameters from IQ-SPECT were 5 S and 12 I for patients with small hearts.