CZT伽马相机与传统伽马相机心肌灌注单光子发射计算机断层扫描诊断心肌梗死及评估心功能的诊断准确性比较
该研究纳入73例已知或疑似慢性冠脉综合征患者,同时用CZT伽马相机和传统伽马相机行门控心肌灌注SPECT,以心脏磁共振为参考,比较心肌梗死检测及左心室容积和射血分数评估的诊断性能。结果显示两种相机检测MI的总体灵敏度、特异度、阳性预测值和阴性预测值相同(67%、100%、100%、69%),CZT在梗死范围>3%时灵敏度82%对传统73%;MPS较CMR显著低估左心室容积,CZT低估程度略轻;两者LVEF准确性均高。结论认为两种相机差异小,似无临床意义。
要点
- CZT与传统伽马相机检测MI的总体灵敏度、特异度、阳性预测值和阴性预测值相同(67%、100%、100%、69%)。
- 梗死范围>3%时灵敏度CZT 82%、传统伽马相机73%。
- MPS较CMR显著低估左心室容积(P≤.002),CZT低估程度较传统相机略轻(2-10 mL,P≤.03)。
核心结论
- CZT与传统伽马相机在检测MI及评估左心室容积和LVEF方面差异小,似无临床意义。
研究设计与样本
- 73例已知或疑似慢性冠脉综合征患者(26%女性)接受CZT伽马相机和传统伽马相机门控MPS及CMR检查。
- 以CMR为参考方法,评估MPS检测MI的存在与范围以及左心室容积、LVEF和左心室质量。
主要发现
- CMR发现42例患者存在MI。
- CZT与传统伽马相机检测MI的总体灵敏度、特异度、阳性预测值和阴性预测值相同(67%、100%、100%、69%)。
- 梗死范围>3%时,灵敏度CZT为82%、传统伽马相机为73%。
- MPS较CMR显著低估左心室容积(所有指标P≤.002),CZT低估程度较传统伽马相机略轻(2-10 mL,所有指标P≤.03)。
- 两种伽马相机LVEF准确性均高。
关键数值与推荐
- MI检测总体灵敏度67%、特异度100%、阳性预测值100%、阴性预测值69%(CZT与传统伽马相机相同)。
- 梗死范围>3%时灵敏度:CZT 82%,传统伽马相机73%。
- 左心室容积低估程度:CZT较传统伽马相机轻2-10 mL(P≤.03)。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The solid-state cadmium-zinc-telluride (CZT) gamma camera for myocardial perfusion single-photon emission computed tomography (MPS) has theoretical advantages compared to the conventional gamma camera technique. This includes more sensitive detectors and better energy resolution. We aimed to explore the diagnostic performance of gated MPS with a CZT gamma camera compared to a conventional gamma camera for detection of myocardial infarct (MI) and assessment of left ventricular (LV) volumes and ejection fraction (LVEF), using cardiac magnetic resonance (CMR) as the reference method. METHODS: Seventy-three patients (26% female) with known or suspected chronic coronary syndrome were examined with gated MPS using both a CZT gamma camera and a conventional gamma camera as well as with CMR. Presence and extent of MI on MPS and late gadolinium enhancement (LGE) CMR was evaluated. For LV volumes, LVEF and LV mass, gated MPS images and cine CMR images were evaluated. RESULTS: MI was found in 42 patients on CMR. The overall sensitivity, specificity, positive and negative predictive values for the CZT and the conventional gamma camera were the same (67%, 100%, 100% and 69%). For infarct size > 3% on CMR, the sensitivity was 82% for the CZT and 73% for the conventional gamma camera, respectively. LV volumes were significantly underestimated by MPS compared to CMR (P ≤ .002 for all measures). The underestimation was slightly less pronounced for the CZT compared to the conventional gamma camera (2-10 mL, P ≤ .03 for all measures). For LVEF, however, accuracy was high for both gamma cameras. CONCLUSION: Differences between a CZT and a conventional gamma camera for detection of MI and assessment of LV volumes and LVEF are small and do not appear to be clinically significant.
引用
- Hedeer F, Akil S, Oddstig J, et al. Diagnostic accuracy for CZT gamma camera compared to conventional gamma camera technique with myocardial perfusion single-photon emission computed tomography: Assessment of myocardial infarction and function. J Nucl Cardiol. 2023. PMID: 36913172. doi:10.1007/s12350-022-03185-0