基于心脏心电图门控15O-水正电子发射断层扫描的左心室容积与射血分数:与同步混合PET/MR心脏磁共振成像的比较
本研究在24例已知或疑似冠心病患者中,使用同步心电图门控心脏PET/MR扫描,比较15O-水PET首过分析法与心脏磁共振成像(CMR)计算左心室容积和射血分数的可行性与一致性。结果显示PET与CMR的容积相关性高(r>0.84),射血分数相关性中等(r=0.52),所有参数一致性高,PET仅对舒张末期容积有轻微高估。结论认为15O-水PET首过分析可高精度计算左心室容积和射血分数,精确度与CMR相当。
要点
- 15O-水PET首过分析计算的左心室容积与CMR相关性高(r>0.84),射血分数相关性中等(r=0.52)。
- PET对舒张末期容积有轻微高估,其他参数无显著平均偏倚,所有参数一致性高。
- 容积的观察者间和观察者内一致性在PET与CMR间均高且相当;射血分数观察者间一致性PET更高,观察者内一致性CMR更高。
核心结论
- 15O-水PET首过分析可高精度计算左心室容积和射血分数,精确度与CMR相当。
研究设计与样本
- 研究纳入24例已知或疑似冠心病患者,进行同步心电图门控心脏PET/MR扫描。
- 15O-水首过图像(0-50秒)使用CarPET软件分析,CMR图像使用Segment软件分析,计算左心室容积和射血分数。
- 采用相关分析和Bland-Altman分析比较PET与CMR的左心室容积和射血分数,并评估两种模态的观察者间和观察者内变异性。
主要发现
- PET与CMR的容积相关性高(r>0.84),射血分数相关性中等(r=0.52),射血分数相关性中等部分因射血分数值范围小。
- 所有参数一致性高,PET对舒张末期容积有轻微高估,其他参数无显著平均偏倚。
- 容积的观察者间和观察者内一致性高,且在PET与CMR间相当。
- 射血分数的观察者间一致性PET更高,观察者内一致性CMR更高。
局限性
- 射血分数相关性中等,部分因射血分数值范围小。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: 15O-water PET is the gold standard for noninvasive quantification of myocardial blood flow. In addition to evaluation of ischemia, the assessment of cardiac function and remodeling is important in all cardiac diseases. However, since 15O-water is freely diffusible and standard uptake images show little contrast between the myocardium and blood pool, the assessment of left-ventricular (LV) volumes and ejection fraction (EF) is challenging. Therefore, the aim of the present study was to investigate the feasibility of calculating LV volumes and EF from first-pass analysis of 15O-water PET, by comparison with cardiac magnetic resonance imaging (CMR) using a hybrid PET/MR scanner. METHODS: Twenty-four patients with known or suspected CAD underwent a simultaneous ECG-gated cardiac PET/MR scan. The 15O-water first-pass images (0-50 seconds) were analyzed using the CarPET software and the CMR images were analyzed using the software Segment, for LV volumes and EF calculations. The LV volumes and EF were compared using correlation and Bland-Altman analysis. In addition, inter- and intra-observer variability of LV volumes and EF were assessed for both modalities. RESULTS: The correlation between PET and CMR was strong for volumes (r > 0.84) and moderate for EF (r = 0.52), where the moderate correlation for EF was partly due to the small range of EF values. Agreement was high for all parameters, with a slight overestimation of PET values for end-diastolic volume but with no significant mean bias for other parameters. Inter- and intra-observer agreement of volumes was high and comparable between PET and CMR. For EF, inter-observer agreement was higher for PET and intra-observer agreement was higher for CMR. CONCLUSION: LV volumes and EF can be calculated by first-pass analysis of a 15O-water PET scan with high accuracy and comparable precision as with CMR.