文献速递心脏核医学

剂量一致的动态SPECT

Dose-consistent dynamic SPECT.
PubMed 文献 · J Nucl Cardiol · 2022-12-08

该研究提出剂量一致的动态SPECT方法,并在20例临床研究中验证。结果显示,所有感兴趣区位置的平均动脉输入函数比值与注射放射性示踪剂剂量比一致,总灌注缺损与冠状动脉血流储备在检出阻塞性冠状动脉疾病患者时一致率>80%,但个体患者中感兴趣区位置可使冠状动脉血流储备改变约30%。

要点

  • 所有感兴趣区位置的平均动脉输入函数比值与注射剂量比一致。
  • 总灌注缺损与冠状动脉血流储备检出阻塞性冠状动脉疾病的一致率>80%。
  • 个体患者中感兴趣区位置改变可使冠状动脉血流储备典型改变约30%。

核心结论

  • 剂量一致的动态SPECT可能有助于评估冠状动脉血流储备测量的不确定性,并可能通过改善个体患者的感兴趣区定位来降低不确定性。

研究设计与样本

  • 研究引入剂量一致的动态SPECT方法,并在20例患者临床研究中展示其应用,比较不同感兴趣区位置的影响并与心肌灌注显像进行对比。

主要发现

  • 所有检查的感兴趣区位置的平均动脉输入函数比值与注射剂量比一致。
  • 在检出阻塞性冠状动脉疾病患者时,所有考虑的感兴趣区位置的总灌注缺损与冠状动脉血流储备一致率>80%。
  • 个体患者中观察到冠状动脉血流储备对感兴趣区位置有显著依赖,典型改变约30%。
  • 所提出方法的不确定性经评估约为7%,小于因感兴趣区位置选择所致的变异性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Coronary flow reserve (CFR) values measured by dynamic SPECT systems are typically consistent with other modalities (e.g., PET). However, large discrepancies are often observed for individual patients. Positioning of the region-of-interest (ROI), representing the arterial input function (AIF) could explain some of these discrepancies. We explored the possibility of positioning the ROI in a manner that evaluates its consistency with patient-based injected radiotracer doses. METHODS: Dose-consistent dynamic SPECT methodology was introduced, and its application was demonstrated in a twenty-patient clinical study. The effect of various ROI positions was investigated and comparison to myocardial perfusion imaging was performed. RESULTS: Mean AIF ratios were consistent with the injected dose ratios for all examined ROI positions. Good agreement (> 80%) between total perfusion deficit and CFR was found in the detection of obstructive CAD patients for all ROIs considered. However, for individual patients, significant dependence on ROI position was observed (altering CFR by typically 30%). The proposed methodology's uncertainty was evaluated (~ 7%) and found to be smaller than the variability due to choice of ROI position. CONCLUSION: Dose-consistent dynamic SPECT may contribute to evaluating uncertainty of CFR measurements and may potentially decrease uncertainty by allowing improved ROI positioning for individual patients.

来源:J Nucl Cardiol,查看原文。版权归原作者所有。

引用

  1. Krakovich A, Gelbart E, Moalem I, et al. Dose-consistent dynamic SPECT. J Nucl Cardiol. 2022. PMID: 36477896. doi:10.1007/s12350-022-03160-9

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。