文献速递肿瘤 PET/CT放射性药物与辐射防护

新型贝叶斯惩罚似然重建算法(HYPER Iterative)对68Ga-DOTA-NOC PET/CT图像质量影响的模体与临床评估

Phantom and clinical evaluation of the effect of a new Bayesian penalized likelihood reconstruction algorithm (HYPER Iterative) on 68Ga-DOTA-NOC PET/CT image quality.
PubMed 文献 · EJNMMI Res · 2022-12-12

研究纳入模体和25例神经内分泌肿瘤患者,比较HYPER Iterative算法不同惩罚因子(0.03–0.5)与OSEM在2和3分钟/床位的68Ga-DOTA-NOC PET/CT图像质量。结果显示惩罚因子0.2时病灶对比度更高、图像噪声更低,可在减少注射放射性和缩短采集时间下达到与OSEM相当的图像质量。

要点

  • HYPER Iterative惩罚因子0.2时病灶SUVmax较OSEM 3 m/b组分别提高19–22%(2 m/b组)和25%(3 m/b组),均p<0.001。
  • HYPER Iterative 2 m/b组惩罚因子0.07–0.2的对比度恢复与OSEM 3 m/b组相当,背景变异性更优。
  • 惩罚因子0.2时HYPER Iterative 2 m/b组主观评分3.20±0.52与OSEM 3 m/b组3.09±0.36相当(p=0.388),3 m/b组3.70±0.36更高(p<0.001)。

核心结论

  • HYPER Iterative算法在惩罚因子0.2时较OSEM获得更高的病灶对比度和更低的图像噪声,可在减少注射放射性和缩短采集时间下达到相同图像质量。

研究设计与样本

  • 研究纳入模体和25例接受68Ga-DOTA-NOC PET/CT的神经内分泌肿瘤患者。
  • PET数据以列表模式采集于数字PET/CT扫描仪,采用OSEM和HYPER Iterative算法重建,惩罚因子为0.03至0.5共七个,采集时间为每床位2分钟和3分钟,均包含飞行时间和点扩散函数恢复。

主要发现

  • 四个热球的对比度恢复、背景变异性和放射性浓度比随惩罚因子增大而降低。
  • HYPER Iterative 2 m/b组惩罚因子0.07至0.2的对比度恢复与OSEM 3 m/b组相当,背景变异性更优。
  • 所有重建组的肝脏SUVmean近似相等(范围5.95–5.97)。
  • HYPER Iterative 2 m/b和3 m/b组惩罚因子0.1至0.2的肝脏变异系数与OSEM 3 m/b组相当(分别为p=0.113–0.711和p=0.079–0.287),而病灶SUVmax分别显著增加19–22%和25%(均p<0.001)。
  • HYPER Iterative 2 m/b组惩罚因子0.2的主观评分最高(3.20±0.52),3 m/b组为3.70±0.36,与OSEM 3 m/b组(3.09±0.36)相比分别p=0.388和p<0.001。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Bayesian penalized likelihood (BPL) algorithm is an effective way to suppress noise in the process of positron emission tomography (PET) image reconstruction by incorporating a smooth penalty. The strength of the smooth penalty is controlled by the penalization factor. The aim was to investigate the impact of different penalization factors and acquisition times in a new BPL algorithm, HYPER Iterative, on the quality of 68Ga-DOTA-NOC PET/CT images. A phantom and 25 patients with neuroendocrine neoplasms who underwent 68Ga-DOTA-NOC PET/CT were included. The PET data were acquired in a list-mode with a digital PET/CT scanner and reconstructed by ordered subset expectation maximization (OSEM) and the HYPER Iterative algorithm with seven penalization factors between 0.03 and 0.5 for acquisitions of 2 and 3 min per bed position (m/b), both including time-of-flight and point of spread function recovery. The contrast recovery (CR), background variability (BV) and radioactivity concentration ratio (RCR) of the phantom; The SUVmean and coefficient of variation (CV) of the liver; and the SUVmax of the lesions were measured. Image quality was rated by two radiologists using a five-point Likert scale. RESULTS: The CR, BV, and RCR decreased with increasing penalization factors for four "hot" spheres, and the HYPER Iterative 2 m/b groups with penalization factors of 0.07 to 0.2 had equivalent CR and superior BV performance compared to the OSEM 3 m/b group. The liver SUVmean values were approximately equal in all reconstruction groups (range 5.95-5.97), and the liver CVs of the HYPER Iterative 2 m/b and 3 m/b groups with the penalization factors of 0.1 to 0.2 were equivalent to those of the OSEM 3 m/b group (p = 0.113-0.711 and p = 0.079-0.287, respectively), while the lesion SUVmax significantly increased by 19-22% and 25%, respectively (all p < 0.001). The highest qualitative score was attained at a penalization factor of 0.2 for the HYPER Iterative 2 m/b group (3.20 ± 0.52) and 3 m/b group (3.70 ± 0.36); those scores were comparable to or greater than that of the OSEM 3 m/b group (3.09 ± 0.36, p = 0.388 and p < 0.001, respectively). CONCLUSIONS: The HYPER Iterative algorithm with a penalization factor of 0.2 resulted in higher lesion contrast and lower image noise than OSEM for 68Ga-DOTA-NOC PET/CT, allowing the same image quality to be achieved with less injected radioactivity and a shorter acquisition time.

来源:EJNMMI Res,查看原文。版权归原作者所有。

引用

  1. Xu L, Cui C, Li R, et al. Phantom and clinical evaluation of the effect of a new Bayesian penalized likelihood reconstruction algorithm (HYPER Iterative) on 68Ga-DOTA-NOC PET/CT image quality. EJNMMI Res. 2022. PMID: 36504014. doi:10.1186/s13550-022-00945-4

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。