99mTc-PYP SPECT与SPECT/CT定量诊断心脏转甲状腺素蛋白淀粉样变性的对比研究
回顾318例疑似心脏转甲状腺素蛋白淀粉样变性(ATTR)患者的99mTc-PYP SPECT/CT数据,评估心肌/血池计数比(MBP)的可重复性与准确性,并比较SPECT/CT与单纯SPECT的视觉和定量结果。SPECT/CT的MBP比值可重复,且较单纯SPECT更准确地对应视觉评分≥2,图像质量更优、模棱两可结果更少。
要点
- 25%的患者心肌摄取阳性,即心肌/肋骨摄取评分≥2分。
- SPECT/CT的MBP比值可重复(1.35±0.68 vs 1.33±0.74,p=.09)。
- SPECT/CT对应视觉评分≥2的ROC AUC为99±1%,优于单纯SPECT的93±3%(p=.02)。
核心结论
- SPECT/CT产生的MBP比值可重复,并能准确识别阳性扫描,图像质量优于单纯SPECT且模棱两可病例更少。
研究设计与样本
- 回顾318例疑似心脏转甲状腺素蛋白淀粉样变性(ATTR)并接受99mTc-PYP SPECT/CT检查的患者数据。
- MBP比值计算后间隔超过1个月独立重复测量。
- 医师独立对SPECT/CT的左心室心肌/肋骨摄取进行评分(0分阴性、1分小于肋骨为模棱两可、2分等于肋骨为阳性、3分大于肋骨为阳性),并对图像质量评分(1分差、2分合格、3分好)。
- 对前191例连续患者子组单独评估单纯SPECT的MBP比值和视觉评分。
主要发现
- 25%的患者心肌摄取阳性(心肌/肋骨摄取评分≥2分)。
- SPECT/CT的MBP比值可重复(1.35±0.68 vs 1.33±0.74,p=.09)。
- SPECT/CT的MBP比值对应视觉评分≥2的ROC AUC为99±1%,单纯SPECT为93±3%(p=.02)。
- SPECT/CT图像质量优于单纯SPECT(2.7±0.5 vs 2.1±0.5,p<.0001),模棱两可结果更少(2.6% vs 22.5%,p<.0001)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: We investigated quantitative 99mTc-pyrophosphate (PYP) SPECT/CT reproducibility and accuracy for diagnosing cardiac transthyretin amyloidosis (ATTR), and whether SPECT/CT improved visual and quantitative results compared to SPECT-only. METHODS: Data were reviewed for 318 patients with suspected ATTR who underwent PYP SPECT/CT. Myocardial-to-blood pool count (MBP) ratios were computed and repeated independently > 1 month later. A physician independently scored LV myocardial-to-rib uptake on SPECT/CT as: 0 (negative), 1 < rib (equivocal), 2 = rib (positive) or 3 > rib (positive), and the image quality as: 1 (poor), 2 (adequate), and 3 (good). SPECT-only MBP ratios and visual scores were assessed separately for a subgroup of the first sequential 191 patients. RESULTS: 25% of patients had positive myocardial uptake (myocardial-to-rib uptake score of ≥ 2). SPECT/CT MBP ratios were reproducible (1.35 ± .68 vs 1.33 ± .74, p = .09) and corresponded with visual scores ≥ 2 (ROC AUC = 99 ± 1%) more accurately than SPECT-only MBPs (93 ± 3%, p = .02). SPECT/CT image quality was better than that of SPECT-only (2.7 ± .5 vs 2.1 ± .5, p < .0001) with fewer equivocal results (2.6% vs 22.5%, p < .0001). CONCLUSION: SPECT/CT produces MBP ratios that are reproducible and accurately identify a positive scan, with better image quality and fewer equivocal cases than SPECT-only.