多针孔准直器提高多巴胺转运体SPECT视觉判读的组内与组间一致性及确定性
研究纳入71例患者,单次注射182±9 MBq 123I-FP-CIT后随机顺序行多针孔(MPH)与LEHRHS准直器SPECT采集,比较视觉判读的组内与组间一致性。MPH图像组内kappa(6分制)为0.84±0.12,高于LEHRHS-OSEM的0.73±0.06和LEHRHS-FBP的0.73±0.08;正常DAT-SPECT的负向Likert评分从LEHRHS-OSEM到MPH设置降低0.87±0.18分。结论认为MPH准直器提高了DAT-SPECT视觉判读的组内与组间一致性及确定性。
要点
- MPH图像6分制组内kappa为0.84±0.12,高于LEHRHS-OSEM的0.73±0.06和LEHRHS-FBP的0.73±0.08。
- MPH图像6分制组间kappa为0.70±0.06,高于LEHRHS-OSEM的0.63±0.08和LEHRHS-FBP的0.48±0.05。
- 正常DAT-SPECT的负向Likert评分从LEHRHS-OSEM到MPH设置降低0.87±0.18分,降低DAT-SPECT的正向评分平均无变化。
核心结论
- MPH准直器提高了DAT-SPECT视觉判读的组内与组间一致性以及确定性。
研究设计与样本
- 研究纳入71例患者(年龄62.1±12.7岁),单次注射182±9 MBq 123I-FP-CIT后按随机顺序行MPH与LEHRHS准直器SPECT采集。
- MPH投影采用迭代3D蒙特卡洛算法重建,LEHRHS投影采用滤波反投影(FBP)或有序子集期望最大化联合分辨率恢复(OSEM)重建。
- 由三名独立判读者对图像进行两次视觉评估,判断纹状体123I-FP-CIT摄取是否存在帕金森型降低,采用Likert 6分制(-3=明确正常,…,3=明确降低)。
- 组内不一致时经组内共识确定,采用Cohen kappa表征6分制及二分化评分的组内与组间一致性。
主要发现
- MPH/LEHRHS-OSEM/LEHRHS-FBP图像的组内kappa(6分制,三名判读者均值)分别为0.84±0.12/0.73±0.06/0.73±0.08。
- MPH/LEHRHS-OSEM/LEHRHS-FBP图像的组内kappa(二分化评分)分别为1.00±0.00/0.96±0.04/0.97±0.03。
- MPH/LEHRHS-OSEM/LEHRHS-FBP图像的组间kappa(6分制,三对判读者均值)分别为0.70±0.06/0.63±0.08/0.48±0.05。
- MPH/LEHRHS-OSEM/LEHRHS-FBP图像的组间kappa(二分化评分)分别为1.00±0.00/0.92±0.04/0.90±0.06。
- 正常DAT-SPECT的负向Likert评分从LEHRHS-OSEM到MPH设置降低0.87±0.18分,降低DAT-SPECT的正向Likert评分平均无变化。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Multiple-pinhole (MPH) collimators improve the resolution-sensitivity trade-off compared to parallel-hole collimators. This study evaluated the impact of MPH collimators on intra- and between-rater agreement, and on the certainty of visual interpretation in dopamine transporter (DAT)-SPECT. METHODS: The study included 71 patients (62.1 ± 12.7 y). Two SPECT acquisitions were performed in randomized order after a single injection of 182 ± 9 MBq 123I-FP-CIT, one with MPH and one with low-energy-high-resolution-high-sensitivity (LEHRHS) collimators. MPH projections were reconstructed with an iterative 3d Monte Carlo algorithm. LEHRHS projections were reconstructed with filtered backprojection (FBP) or with ordered-subsets expectation-maximization and resolution recovery (OSEM). Images were visually evaluated twice by three independent raters with respect to presence/absence of Parkinson-typical reduction of striatal 123I-FP-CIT uptake using a Likert 6-score (- 3 = clearly normal, …, 3 = clearly reduced). In case of intra-rater discrepancy, an intra-rater consensus was obtained. Intra- and between-rater agreement with respect to the Likert score (6-score and dichotomized score) was characterized by Cohen's kappa. RESULTS: Intra-rater kappa of visual scoring of MPH/LEHRHS-OSEM/LEHRHS-FBP images was 0.84 ± 0.12/0.73 ± 0.06/0.73 ± 0.08 (6-score, mean of three raters) and 1.00 ± 0.00/0.96 ± 0.04/0.97 ± 0.03 (dichotomized score). Between-rater kappa of visual scoring (intra-rater consensus) of MPH/LEHRHS-OSEM/LEHRHS-FBP images was 0.70 ± 0.06/0.63 ± 0.08/0.48 ± 0.05 (6-score, mean of three pairs of raters) and 1.00 ± 0.00/0.92 ± 0.04/0.90 ± 0.06 (dichotomized score). There was a decrease of (negative) Likert scores in normal DAT-SPECT by 0.87 ± 0.18 points from the LEHRHS-OSEM to the MPH setting. The (positive) Likert scores of reduced DAT-SPECT did not change on average. CONCLUSIONS: MPH collimators improve intra- and between-rater agreement as well as the certainty of the visual interpretation of DAT-SPECT.