文献速递肿瘤 PET/CT放射性药物与辐射防护

基于深度学习的框架利用LSO本底辐射在长轴视野PET扫描仪中生成全身衰减图的定量评估

Quantitative evaluation of a deep learning-based framework to generate whole-body attenuation maps using LSO background radiation in long axial FOV PET scanners.
PubMed 文献 · Eur J Nucl Med Mol Imaging · 2022-07-19

该研究利用长轴视野(LAFOV)PET扫描仪的LSO闪烁体本底辐射,开发并评估基于深度学习与联合重建的方法生成衰减图(μ-map)。18例受试者数据用于训练卷积神经网络,深度学习增强的μ-map与MLACF及CT来源图对比;结果显示DL-MLACF-POST的rMAE较MLACF-POST平均降低3.3倍,PET图像rMAE降低2.6倍,SUV平均绝对误差在健康器官<5%、脑灰质<7%、全部肿瘤合计4.3%。

要点

  • 深度学习增强的μ-map(DL-MLACF-PRE与DL-MLACF-POST)在脂肪、水基软组织及骨骼的rME值与CT法无统计学显著差异。
  • DL-MLACF-POST的rMAE较MLACF-POST平均降低3.3倍,相应重建PET图像rMAE降低2.6倍。
  • PETDL-MLACF-POST对比PETCT的SUV平均绝对误差在健康器官<5%、脑灰质<7%、全部肿瘤合计4.3%。

核心结论

  • 作者描述了一种基于深度学习的方法,可从PET发射数据和LSO本底辐射准确生成μ-map,在LAFOV PET扫描仪中实现无CT的衰减和散射校正。

研究设计与样本

  • 研究使用18例受试者数据训练卷积神经网络,以增强基于联合活度与衰减重建算法(MLACF)及LSO本底辐射透射数据生成的初始μ-map。
  • 分别使用18F-FDG给药前(PRE)和给药后(POST)的LSO本底辐射数据生成μ-mapMLACF-PRE和μ-mapMLACF-POST,并训练生成对应的深度学习增强图μ-mapDL-MLACF-PRE和μ-mapDL-MLACF-POST。
  • 将深度学习增强的μ-map与MLACF衍生图及CT来源图(μ-mapCT)进行比较,并通过评估各μ-map重建的PET图像、计算基于感兴趣区的标准摄取值及相对CT法的百分比相对平均误差(rME)和相对平均绝对误差(rMAE)来评估方法性能。

主要发现

  • μ-mapDL-MLACF-PRE与μ-mapDL-MLACF-POST在脂肪基、水基软组织及骨骼中的rME值均无统计学显著差异,提示体内放射性药物活性的存在对所得μ-map影响可忽略。
  • μ-mapDL-MLACF-POST的rMAE值较μ-mapMLACF-POST的rMAE平均降低3.3倍。
  • 使用μ-mapDL-MLACF-POST重建的PET图像(PETDL-MLACF-POST)的平均rMAE值较使用μ-mapMLACF-POST重建的PET图像平均rMAE值小2.6倍。
  • PETDL-MLACF-POST对比PETCT的SUV平均绝对误差在健康器官中小于5%,在脑灰质中小于7%,全部肿瘤合计为4.3%。

关键数值与推荐

  • 深度学习增强μ-map的rME值与CT法无统计学显著差异(脂肪基、水基软组织及骨骼)。
  • μ-mapDL-MLACF-POST的rMAE较μ-mapMLACF-POST平均降低3.3倍。
  • PETDL-MLACF-POST的rMAE较PETMLACF-POST平均降低2.6倍。
  • PETDL-MLACF-POST对比PETCT的SUV平均绝对误差:健康器官<5%,脑灰质<7%,全部肿瘤合计4.3%。

临床怎么用

  • 原文未据此给出临床推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: Attenuation correction is a critically important step in data correction in positron emission tomography (PET) image formation. The current standard method involves conversion of Hounsfield units from a computed tomography (CT) image to construct attenuation maps (µ-maps) at 511 keV. In this work, the increased sensitivity of long axial field-of-view (LAFOV) PET scanners was exploited to develop and evaluate a deep learning (DL) and joint reconstruction-based method to generate µ-maps utilizing background radiation from lutetium-based (LSO) scintillators. METHODS: Data from 18 subjects were used to train convolutional neural networks to enhance initial µ-maps generated using joint activity and attenuation reconstruction algorithm (MLACF) with transmission data from LSO background radiation acquired before and after the administration of 18F-fluorodeoxyglucose (18F-FDG) (µ-mapMLACF-PRE and µ-mapMLACF-POST respectively). The deep learning-enhanced µ-maps (µ-mapDL-MLACF-PRE and µ-mapDL-MLACF-POST) were compared against MLACF-derived and CT-based maps (µ-mapCT). The performance of the method was also evaluated by assessing PET images reconstructed using each µ-map and computing volume-of-interest based standard uptake value measurements and percentage relative mean error (rME) and relative mean absolute error (rMAE) relative to CT-based method. RESULTS: No statistically significant difference was observed in rME values for µ-mapDL-MLACF-PRE and µ-mapDL-MLACF-POST both in fat-based and water-based soft tissue as well as bones, suggesting that presence of the radiopharmaceutical activity in the body had negligible effects on the resulting µ-maps. The rMAE values µ-mapDL-MLACF-POST were reduced by a factor of 3.3 in average compared to the rMAE of µ-mapMLACF-POST. Similarly, the average rMAE values of PET images reconstructed using µ-mapDL-MLACF-POST (PETDL-MLACF-POST) were 2.6 times smaller than the average rMAE values of PET images reconstructed using µ-mapMLACF-POST. The mean absolute errors in SUV values of PETDL-MLACF-POST compared to PETCT were less than 5% in healthy organs, less than 7% in brain grey matter and 4.3% for all tumours combined. CONCLUSION: We describe a deep learning-based method to accurately generate µ-maps from PET emission data and LSO background radiation, enabling CT-free attenuation and scatter correction in LAFOV PET scanners.