基于体模与临床数据评估Biograph Vision Quadra PET/CT的EARL合规性及成像优化
本研究在长轴视野(LAFOV)Siemens Biograph Vision Quadra PET/CT上,用NEMA图像质量体模在LAFOV中心及不同位置行EARL合规性体模测量,并采集10例肿瘤患者PET数据,将10分钟列表模式采集重处理为更短扫描时长。体模在所有位置和所有扫描时长均符合EARL标准;患者数据中,符合EARL的重建图像在不同扫描时长间病灶SUVmax、SUVpeak及背景SUVmean、SUVpeak无临床有意义的显著差异,COV仅轻微变化;结论为图像符合EARL规范,扫描时长和/或给药活度最多可降低十倍而不受噪声增加干扰。
要点
- Vision Quadra PET/CT体模测量在LAFOV所有位置及所有扫描时长均符合EARL标准。
- 符合EARL重建的患者图像在不同扫描时长间病灶SUVmax、SUVpeak及背景SUVmean、SUVpeak无临床有意义的显著差异。
- 扫描时长和/或给药活度最多可降低十倍而不受噪声增加干扰。
核心结论
- 使用Vision Quadra PET/CT获得的图像符合EARL规范。
- 扫描时长和/或给药活度最多可降低十倍而不受噪声增加干扰。
研究设计与样本
- 使用NEMA图像质量体模在LAFOV中心及不同位置行EARL合规性体模测量,PET数据最大环差(MRD)=85,采用多种重建参数重建并重处理为更短扫描时长。
- 采集10例肿瘤患者PET数据(MRD=85),采用多种重建设置重建,并将10分钟列表模式采集重处理为更短扫描时长。
- 对病灶SUVmax和SUVpeak行重复测量方差分析,对背景SUVpeak和SUVmean不同扫描时长间差异行Wilcoxon符号秩检验,并计算变异系数(COV)表征噪声。
主要发现
- 体模测量显示LAFOV所有位置、所有扫描时长均符合EARL标准。
- 患者数据中,符合EARL的图像在不同扫描时长间病灶SUVmax和SUVpeak、背景SUVmean和SUVpeak无临床有意义的显著差异。
- 不同扫描时长间COV仅轻微变化。
关键数值与推荐
- 体模测量获得每个球体的最大、平均和峰值活度浓度恢复系数(RC),并与EARL标准规范比较。
- 患者数据中符合EARL的重建图像在不同扫描时长间病灶SUVmax、SUVpeak及背景SUVmean、SUVpeak无临床有意义的显著差异。
- 扫描时长和/或给药活度最多可降低十倍而不受噪声增加干扰。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
PURPOSE: Current European Association of Nuclear Medicine (EANM) Research Ltd. (EARL) guidelines for the standardisation of PET imaging developed for conventional systems have not yet been adjusted for long axial field-of-view (LAFOV) systems. In order to use the LAFOV Siemens Biograph Vision Quadra PET/CT (Siemens Healthineers, Knoxville, TN, USA) in multicentre research and harmonised clinical use, compliance to EARL specifications for 18F-FDG tumour imaging was explored in the current study. Additional tests at various locations throughout the LAFOV and the use of shorter scan durations were included. Furthermore, clinical data were collected to further explore and validate the effects of reducing scan duration on semi-quantitative PET image biomarker accuracy and precision when using EARL-compliant reconstruction settings. METHODS: EARL compliance phantom measurements were performed using the NEMA image quality phantom both in the centre and at various locations throughout the LAFOV. PET data (maximum ring difference (MRD) = 85) were reconstructed using various reconstruction parameters and reprocessed to obtain images at shorter scan durations. Maximum, mean and peak activity concentration recovery coefficients (RC) were obtained for each sphere and compared to EARL standards specifications. Additionally, PET data (MRD = 85) of 10 oncological patients were acquired and reconstructed using various reconstruction settings and reprocessed from 10 min listmode acquisition into shorter scan durations. Per dataset, SUVs were derived from tumour lesions and healthy tissues. ANOVA repeated measures were performed to explore differences in lesion SUVmax and SUVpeak. Wilcoxon signed-rank tests were performed to evaluate differences in background SUVpeak and SUVmean between scan durations. The coefficient of variation (COV) was calculated to characterise noise. RESULTS: Phantom measurements showed EARL compliance for all positions throughout the LAFOV for all scan durations. Regarding patient data, EARL-compliant images showed no clinically meaningful significant differences in lesion SUVmax and SUVpeak or background SUVmean and SUVpeak between scan durations. Here, COV only varied slightly. CONCLUSION: Images obtained using the Vision Quadra PET/CT comply with EARL specifications. Scan duration and/or activity administration can be reduced up to a factor tenfold without the interference of increased noise.
引用
- van Sluis J, van Snick JH, Brouwers AH, et al. EARL compliance and imaging optimisation on the Biograph Vision Quadra PET/CT using phantom and clinical data. Eur J Nucl Med Mol Imaging. 2022. PMID: 35876867. doi:10.1007/s00259-022-05919-1
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。