文献速递心脏核医学

IQ·SPECT高速扫描在肥胖患者中的诊断效能:一项初步质量控制研究

Diagnostic performance of IQ·SPECT with high-speed scanning: A preliminary quality control study in obese patients.
PubMed 文献 · J Nucl Cardiol · 2022-04-06

这项前瞻性研究对20例BMI>30 kg/m2的肥胖患者先后进行常规准直器和IQ·SPECT的静息-负荷心肌灌注显像,IQ·SPECT图像质量评分中位数≥3.25,低于常规准直器的≥4.0(P≤0.01),但射血分数、SSS和SRS与常规准直器显著相关,SDS未见显著相关。作者认为IQ·SPECT在肥胖患者中与常规SPECT相当,采集时间缩短可能提高检查通量而不损失诊断准确性。

要点

  • 20例BMI>30 kg/m2肥胖患者中,IQ·SPECT图像质量评分中位数≥3.25,低于常规准直器≥4.0(P≤0.01)。
  • IQ·SPECT与常规准直器的EF(r=0.86)、SSS(r=0.75)、SRS(r=0.60)显著相关,SDS(r=0.15)未见显著相关。
  • 作者认为IQ·SPECT在肥胖患者中与常规SPECT相当,采集时间缩短可能提高通量而不损失诊断准确性。

核心结论

  • 作者认为IQ·SPECT在肥胖患者中与常规SPECT相当,其缩短的采集时间可能提高检查通量而不损失诊断准确性。

研究设计与样本

  • 前瞻性研究,纳入20例BMI>30 kg/m2的患者,先后采集常规准直器和IQ·SPECT的静息-负荷心肌灌注显像,采集时间分别为20分钟和7分钟。
  • 由两名盲法独立阅片者评估图像质量(5分制)及心肌灌注指标(SSS、SRS、SDS),并比较软件计算的左心室射血分数。

主要发现

  • 平均BMI为39.6 ± 7.6 kg/m2,12例(BMI 44.1 ± 6.8 kg/m2)为2级或3级肥胖。
  • IQ·SPECT门控/非门控图像质量评分为中位数≥3.25,低于常规准直器的中位数≥4.0(P≤0.01)。
  • IQ·SPECT与常规准直器比较,EF(r=0.86,P<0.01)、SSS(r=0.75,P<0.0001)和SRS(r=0.60,P<0.005)显著相关,SDS(r=0.15)未见显著相关。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: IQ·SPECT is a recently introduced collimator design for myocardial perfusion imaging (MPI). Little data exist on use of this collimator type in obese patients, particularly Class 2 or 3 [body mass index (BMI) > 35 kg/m2]. METHODS: Two consecutive rest-stress MPI scans were prospectively acquired using a conventional collimator and IQ·SPECT (acquisition times of 20 and 7 minutes, respectively) in 20 patients with a BMI of >30 kg/m2. Assigned by two blinded, independent readers, image quality (on a 5-point scale) and metrics of myocardial perfusion [summed stress score (SSS), summed rest score (SRS) and summed difference score (SDS)] were compared. Software-based left ventricular ejection fraction (EF) was also correlated. RESULTS: Mean BMI was 39.6 ± 7.6 kg/m2. Class 2 or 3 obesity was present in 12 patients (BMI, 44.1 ± 6.8 kg/m2). Gated/non-gated images from IQ·SPECT revealed fair to good quality scores (median ≥ 3.25), which were inferior to the conventional collimator (median ≥ 4.0; P ≤ 0.01). Significant correlative indices were achieved when comparing IQ·SPECT and conventional collimators for EF values (r = 0.86, P < 0.01), SSS (r = 0.75, P < 0.0001) and SRS (r = 0.60, P < 0.005), but not for SDS (r = 0.15). CONCLUSION: IQ·SPECT was comparable to conventional SPECT in obese patients. The reduced acquisition time of IQ·SPECT may allow for improved throughput with no loss in diagnostic accuracy.

来源:J Nucl Cardiol,查看原文。版权归原作者所有。

引用

  1. Werner RA, Martinez RB, Marcus C, et al. Diagnostic performance of IQ·SPECT with high-speed scanning: A preliminary quality control study in obese patients. J Nucl Cardiol. 2022. PMID: 35386095. doi:10.1007/s12350-022-02928-3

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。