数据驱动运动校正算法对正电子发射断层摄影灌注成像质量的影响
本研究采用原型数据驱动运动校正(DDMC)算法对40例连续存在运动伪影的患者生成校正图像,并与未校正(NMC)图像对比。结果显示所有校正图像质量均改善且可判读,机器运动量化与医师判读间呈显著相关(Spearman相关系数0.626,P<.001)。
要点
- DDMC算法可改善存在运动的心脏PET图像质量,所有校正图像均可判读。
- 机器运动量化与医师判读间显著相关(Spearman相关系数0.626,P<.001)。
- 40例患者中轻度运动占40%、中度30%、重度15%、极轻度15%。
核心结论
- 新型DDMC算法改善了存在运动的心脏PET图像质量。
- 机器运动量化与医师判读之间的相关性显著。
研究设计与样本
- 研究纳入40例连续存在运动伪影的患者,采集静息和负荷NMC与DDMC图像。
- 由2名盲法研究者采用4级视觉有序量表(0:极轻度运动;1:轻度运动;2:中度运动;3:重度运动/不可判读)进行评分。
- 运动同时用Dwell Fraction量化,即运动向量显示心脏位于校正位置6 mm以内的时间比例,由NMC图像的listmode数据导出。
主要发现
- 极轻度运动见于15%的患者,轻度、中度和重度运动分别见于40%、30%和15%的患者。
- 所有校正图像均显示质量改善,处理后均可判读。
- 机器运动量化与医师判读之间呈显著相关(Spearman相关系数0.626,P<.001)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: Cardiac motion frequently reduces the interpretability of PET images. This study utilized a prototype data-driven motion correction (DDMC) algorithm to generate corrected images and compare DDMC images with non-corrected images (NMC) to evaluate image quality and change of perfusion defect size and severity. METHODS: Rest and stress images with NMC and DDMC from 40 consecutive patients with motion were rated by 2 blinded investigators on a 4-point visual ordinal scale (0: minimal motion; 1: mild motion; 2: moderate motion; 3: severe motion/uninterpretable). Motion was also quantified using Dwell Fraction, which is the fraction of time the motion vector shows the heart to be within 6 mm of the corrected position and was derived from listmode data of NMC images. RESULTS: Minimal motion was seen in 15% of patients, while 40%, 30%, and 15% of patients had mild moderate and severe motion, respectively. All corrected images showed an improvement in quality and were interpretable after processing. This was confirmed by a significant correlation (Spearman's correlation coefficient 0.626, P < .001) between machine measurement of motion quantification and physician interpretation. CONCLUSION: The novel DDMC algorithm improved quality of cardiac PET images with motion. Correlation between machine measurement of motion quantification and physician interpretation was significant.
引用
- Han Y, Ahmed AI, Hayden C, et al. Change in positron emission tomography perfusion imaging quality with a data-driven motion correction algorithm. J Nucl Cardiol. 2022. PMID: 35275348. doi:10.1007/s12350-021-02902-5
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。