新一代30厘米轴向视野Discovery MI PET/CT的NEMA NU 2-2018性能评估

NEMA NU 2-2018 performance evaluation of a new generation 30-cm axial field-of-view Discovery MI PET/CT.
PubMed 文献 · Eur J Nucl Med Mol Imaging · 2022-03-14

该研究依据NEMA NU 2-2018标准,评估6环升级的第二代30厘米轴向视野Discovery MI PET/CT,并与升级前同一系统的5环配置比较;6环系统平均灵敏度为32.76 cps/kBq、噪声等效计数率峰值434.3 kcps,较5环分别提高约47%和39%。

要点

  • 6环DMI平均系统灵敏度32.76 cps/kBq,较5环22.29 cps/kBq提高约47%。
  • 6环噪声等效计数率峰值434.3 kcps(23.6 kBq/mL),较第二代5环提高39%。
  • 6环背景变异度1.9–6.8%,低于5环的2.3–8.5%;肺误差3.16%低于5环的4.0%。

核心结论

  • 作者认为,6环DMI较5环配置灵敏度更高,可能改善缩短扫描时间下的临床图像质量。
  • 作者认为,6环配置同样可用于改善动态采集的时间采样和/或缩短总扫描时间。
  • 作者认为,时间采样和扫描时间可按应用权衡,以进一步降低注射剂量水平。

研究设计与样本

  • 采用NEMA NU 2-2018标准评估PET性能,包括空间分辨率、灵敏度、图像质量、计数率性能、时间分辨率和图像配准精度。
  • 评估对象为6环升级的第二代(Gen2)系统,同时报告同一系统升级前5环配置的数值。
  • 使用患者图像评估图像质量。

主要发现

  • 平均系统灵敏度为32.76 cps/kBq,较5环的22.29 cps/kBq提高约47%。
  • 噪声等效计数率峰值为434.3 kcps,对应23.6 kBq/mL,较第一代(Gen1)提高约60%,较第二代5环提高39%。
  • 对比度恢复范围为54.5%至85.8%,与5环相似。
  • 背景变异度5环为2.3%至8.5%,6环为1.9%至6.8%;肺误差5环为4.0%,6环为3.16%。
  • 1 cm处横向/轴向半高全宽(FWHM)为3.79/4.26 mm,10 cm处为4.29/4.55 mm;散射分数40.2%,能量分辨率9.63%,飞行时间(TOF)分辨率在0 kBq/mL时为389.6 ps。

关键数值与推荐

  • 6环配置平均系统灵敏度32.76 cps/kBq,5环配置22.29 cps/kBq。
  • 噪声等效计数率峰值434.3 kcps对应23.6 kBq/mL。
  • 背景变异度6环1.9%至6.8%,5环2.3%至8.5%。
  • 肺误差6环3.16%,5环4.0%。
  • TOF分辨率在0 kBq/mL时为389.6 ps。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

PURPOSE: The DMI PET/CT is a modular silicon photomultiplier-based scanner with an axial field-of-view (FOV) between 15 and 25 cm depending on ring configuration (3, 4, or 5 rings). A new generation of the system includes a reengineered detector module, featuring improved electronics and an additional 6th ring, extending the axial FOV to 30 cm. We report on the performance evaluation of the 6-ring upgraded Generation 2 (Gen2) system while values are also reported for the 5-ring configuration of the very same system prior to the upgrade. METHODS: PET performance was evaluated using the NEMA NU 2-2018 standard for spatial resolution, sensitivity, image quality, count rate performance, timing resolution, and image co-registration accuracy. Patient images were used to assess image quality. RESULTS: The average system sensitivity was measured at 32.76 cps/kBq (~ 47% increase to 5 rings at 22.29 cps/kBq) while noise equivalent count rate peaked at 434.3 kcps corresponding to 23.6 kBq/mL (~ 60% increase to Generation 1 (Gen1) and 39% to Gen2 5 rings). Contrast recovery ranged between 54.5 and 85.8% similar to 5 rings, while the 6 rings provided lower background variability (2.3-8.5% for 5 rings vs 1.9-6.8% for 6 rings) and lower lung error (4.0% for the 5 rings and 3.16% for the 6 rings). Transverse/axial full width at half-maximum (FWHM) at 1 cm (3.79/4.26 mm) and 10 cm (4.29/4.55 mm), scatter fraction (40.2%), energy resolution (9.63%), and time-of-flight (TOF) resolution (389.6 ps at 0 kBq/mL) were in line to previously reported values measured across different system configurations. Improved patient image quality is obtained with the 6 rings compared to the 5 rings, while image quality is retained even at reduced scan times, enabling WB dynamic acquisitions. CONCLUSIONS: The higher sensitivity of the 6-ring DMI compared to the 5-ring configuration may lead to improved image quality of clinical images at reduced scan time. Additionally, it could equally be used to allow improved temporal sampling and/or reduced overall scan time in dynamic acquisitions. Conversely, temporal sampling and scan time could be traded per application to further drive injected dose at lower levels.