现代PET技术与方法对EANM儿童剂量卡的影响
本文基于已发表的PET/MR图像质量数据,参照EANM成人注射活度指南及床重叠与MBq·min·床⁻¹·kg⁻¹概念,更新了PET/MR及轴向视野≥25 cm的PET/CT系统的EANM儿童剂量卡。新参数为B类、基线值10.7 MBq、最低推荐活度10 MBq;3 kg儿童推荐11 MBq [18F]FDG,68 kg儿童推荐149 MBq,假设每床位扫描3 min;1岁及以上各年龄平均有效剂量为2.85 mSv。
要点
- 新剂量卡针对PET/MR及轴向视野≥25 cm的PET/CT系统,参数为B类、基线值10.7 MBq、最低推荐活度10 MBq。
- 3 min/床位的假设下,3 kg儿童推荐11 MBq [18F]FDG,68 kg儿童推荐149 MBq。
- 1岁及以上各年龄平均有效剂量为2.85 mSv。
核心结论
- 本文为先前未被考虑的PET/MR及轴向视野≥25 cm的PET/CT系统给出了儿童给药推荐。
- 新参数参照EANM成人注射活度指南及MBq·min·床⁻¹·kg⁻¹概念制定。
研究设计与样本
- 本文为简短通讯,参照EANM成人注射活度指南及床重叠与MBq·min·床⁻¹·kg⁻¹概念,利用已发表的PET/MR系统图像质量数据更新EANM儿童剂量卡。
- 更新范围限于PET/MR及轴向视野≥25 cm的PET/CT系统,不涵盖近期上市的半身及全身扫描仪的儿童剂量扩展。
主要发现
- 新参数为B类、基线值10.7 MBq、最低推荐活度10 MBq。
- 在每床位扫描3 min的假设下,3 kg儿童推荐11 MBq [18F]FDG,68 kg儿童推荐149 MBq [18F]FDG。
- 1岁及以上各年龄的平均有效剂量为2.85 mSv。
关键数值与推荐
- 新剂量卡参数:B类,基线值10.7 MBq,最低推荐活度10 MBq。
- 3 kg儿童推荐11 MBq [18F]FDG,假设每床位扫描3 min。
- 68 kg儿童推荐149 MBq [18F]FDG,假设每床位扫描3 min。
- 1岁及以上各年龄平均有效剂量为2.85 mSv。
临床怎么用
- 本文为PET/MR及轴向视野≥25 cm的PET/CT系统给出儿童[18F]FDG给药推荐,参数为B类、基线值10.7 MBq、最低推荐活度10 MBq。
- 推荐活度基于每床位扫描3 min的假设:3 kg儿童11 MBq,68 kg儿童149 MBq。
- 本文的推荐不涵盖半身及全身扫描仪的儿童剂量扩展。
局限性
- 本文为简短通讯,未涵盖近期上市的半身及全身扫描仪的儿童剂量扩展。
- 剂量推荐基于每床位扫描3 min的假设。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIM: Recent advancements in PET technology have brought with it significant improvements in PET performance and image quality. In particular, the extension of the axial field of view of PET systems, and the introduction of semiconductor technology into the PET detector, initially for PET/MR, and more recently available long-field-of-view PET/CT systems (≥ 25 cm) have brought a step change improvement in the sensitivity of PET scanners. Given the requirement to limit paediatric doses, this increase in sensitivity is extremely welcome for the imaging of children and young people. This is even more relevant with PET/MR, where the lack of CT exposures brings further dose reduction benefits to this population. In this short article, we give some details around the benefits around new PET technology including PET/MR and its implications on the EANM paediatric dosage card. MATERIAL AND METHODS : Reflecting on EANM adult guidance on injected activities, and making reference to bed overlap and the concept of MBq.min bed-1 kg-1, we use published data on image quality from PET/MR systems to update the paediatric dosage card for PET/MR and extended axial field of view (≥ 25 cm) PET/CT systems. However, this communication does not cover the expansion of paediatric dosing for the half-body and total-body scanners that have recently come to market. RESULTS: In analogy to the existing EANM dosage card, new parameters for the EANM paediatric dosage card were developed (class B, baseline value: 10.7 MBq, minimum recommended activity 10 MBq). The recommended administered activities for the systems considered in this communication range from 11 MBq [18F]FDG for a child with a weight of 3 kg to 149 MBq [18F]FDG for a paediatric patient weight of 68 kg, assuming a scan of 3 min per bed position. The mean effective dose over all ages (1 year and older) is 2.85 mSv. CONCLUSION: With this, recommendations for paediatric dosing are given for systems that have not been considered previously.
引用
- Dickson J, Eberlein U, Lassmann M, et al. The effect of modern PET technology and techniques on the EANM paediatric dosage card. Eur J Nucl Med Mol Imaging. 2021. PMID: 34910233. doi:10.1007/s00259-021-05635-2
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。