颅骨密度比直方图偏度预测MRgFUS热升温更准确
本研究回顾性分析55例接受单侧VIM丘脑毁损术的特发性震颤患者,量化1024个换能器单元的颅骨密度比(SDR)直方图偏度、峰度、熵和均匀性,发现SDR偏度是热升温的显著预测因子,其预测峰值温度>55°C的AUC为0.847,优于SDR平均值的0.784,建议截断值取0.68。
要点
- SDR偏度是MRgFUS热升温的独立预测因子,优于传统SDR平均值。
- SDR偏度预测峰值温度>55°C的AUC为0.847,SDR平均值为0.784。
- 建议SDR偏度截断值设为0.68,可能纳入被SDR平均值标准排除的可治疗患者。
核心结论
- SDR偏度比传统SDR平均值更准确地预测MRgFUS热升温,建议截断值0.68。
研究设计与样本
- 回顾性分析55例接受单侧VIM丘脑毁损术的特发性震颤患者,量化1024个换能器单元的SDR直方图偏度、峰度、熵和均匀性。
主要发现
- 单因素和多因素logistic回归分析显示SDR偏度是热升温的显著预测因子(p<0.001,p=0.013)。
- ROC曲线分析显示SDR偏度预测峰值温度>55°C的AUC为0.847,SDR平均值为0.784。
关键数值与推荐
- SDR偏度预测热升温的截断值建议设为0.68。
临床怎么用
- SDR偏度可用于筛选适合MRgFUS治疗的特发性震颤患者,可能纳入被SDR平均值标准排除的可治疗患者。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Sufficient thermal increase capable of generating thermocoagulation is indispensable for an effective clinical outcome in patients undergoing magnetic resonance-guided focused ultrasound (MRgFUS). The skull density ratio (SDR) is one of the most dominant predictors of thermal increase prior to treatment. However, users currently rely only on the average SDR value (SDRmean) as a screening criterion, although some patients with low SDRmean values can achieve sufficient thermal increase. The present study aimed to examine the numerical distribution of SDR values across 1024 elements to identify more precise predictors of thermal increase during MRgFUS. METHODS: The authors retrospectively analyzed the correlations between the skull parameters and the maximum temperature achieved during unilateral ventral intermediate nucleus thalamotomy with MRgFUS in a cohort of 55 patients. In addition, the numerical distribution of SDR values was quantified across 1024 elements by using the skewness, kurtosis, entropy, and uniformity of the SDR histogram. Next, the authors evaluated the correlation between the aforementioned indices and a peak temperature > 55°C by using univariate and multivariate logistic regression analyses. Receiver operating characteristic curve analysis was performed to compare the predictive ability of the indices. The diagnostic performance of significant factors was also assessed. RESULTS: The SDR skewness (SDRskewness) was identified as a significant predictor of thermal increase in the univariate and multivariate logistic regression analyses (p < 0.001, p = 0.013). Moreover, the receiver operating characteristic curve analysis indicated that the SDRskewness exhibited a better predictive ability than the SDRmean, with area under the curve values of 0.847 and 0.784, respectively. CONCLUSIONS: The SDRskewness is a more accurate predictor of thermal increase than the conventional SDRmean. The authors suggest setting the SDRskewness cutoff value to 0.68. SDRskewness may allow for the inclusion of treatable patients with essential tremor who would have been screened out based on the SDRmean exclusion criterion.
引用
- Iijima K, Yokota H, Yamaguchi T, et al. Predictors of thermal increase in magnetic resonance-guided focused ultrasound treatment for essential tremor: histogram analysis of skull density ratio values for 1024 elements. J Neurosurg. 2021. PMID: 34653973. doi:10.3171/2021.5.JNS21669
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。