5-氨基乙酰丙酸诱导的多种脑肿瘤中卟啉含量:对成像设备设计与验证的启示

5-Aminolevulinic Acid-Induced Porphyrin Contents in Various Brain Tumors: Implications Regarding Imaging Device Design and Their Validation.
PubMed 文献 · Neurosurgery · 2021-01-01

该研究利用高光谱成像系统分析162例患者524份肿瘤样本,校准后计算原卟啉IX浓度(cPPIX),并由经验丰富的神经外科医生记录BLUE400滤镜下的肉眼荧光。结果定义0.9 μg/ml为可肉眼检测和区分荧光的最小cPPIX阈值,为当前显微镜技术的灵敏度与特异性提供依据,并提示新设备需达到类似性能方可安全使用。

要点

  • BLUE400滤镜下肉眼可检测荧光的最小cPPIX阈值为0.9 μg/ml。
  • 新荧光成像技术需验证与现有标准相当的检测性能。
  • 若新设备更灵敏,需进一步评估肿瘤选择性。

核心结论

  • 首次报道当前一代显微镜肉眼检测荧光所需的cPPIX阈值,为0.9 μg/ml。

研究设计与样本

  • 利用高光谱成像系统分析来自162例患者的524份不同脑肿瘤组织样本,包含有或无可见荧光者。
  • 以已知cPPIX的荧光体模校准系统后计算绝对cPPIX值。
  • 由经验丰富的神经外科医生记录BLUE400滤镜下的肉眼荧光。

主要发现

  • 共分析524份肿瘤样本,来自162例患者。
  • 定义0.9 μg/ml的cPPIX为肉眼检测和区分荧光所需的最小浓度。

关键数值与推荐

  • 肉眼荧光检测阈值:cPPIX 0.9 μg/ml(BLUE400滤镜下)。

临床怎么用

  • 新荧光成像技术应展示与BLUE400系统相似的检测特性方可安全有效使用。
  • 若新技术更灵敏,需进一步评估其肿瘤选择性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Fluorescence-guided resections using 5-aminolevulinic acid (5-ALA)-induced tumor porphyrins have been established as an adjunct for malignant glioma surgery based on a phase III study using specifically adapted microscopes for visualizing fluorescing protoporphyrin IX (PPIX). New hardware technologies are being introduced, which claim the same performance as the original technology for visualizing fluorescence. This assumes that qualitative fluorescence detection is equivalent to the established standard, an assumption that needs to be critically assessed. OBJECTIVE: To determine PPIX concentrations (cPPIX) in tissue that can be detected visually using the established BLUE400 filter system (Carl Zeiss Meditec, Oberkochen, Germany) as a basis for defining the performance of this system. METHODS: Utilizing a hyperspectral imaging system, tumor samples from patients harboring different tumor tissues, with or without visible fluorescence, were analyzed. Absolute values of cPPIX were calculated after calibrating the system with fluorescence phantoms with known cPPIX. RESULTS: A total of 524 tumor samples from 162 patients were analyzed. Visual fluorescence under the BLUE400 filter was documented by experienced neurosurgeons. A 0.9 μg/ml threshold of cPPIX was defined as the minimal concentration required to detect and discriminate visual fluorescence. CONCLUSION: This is the first report providing data on the threshold of cPPIX, which is visually detected using the current generation of microscopes, thus defining the specificity and sensitivity of this technology as initially tested in a randomized trial. Novel technologies should show similar characteristics in order to be used safely and effectively. If more sensitive, such technologies require further assessments of tumor selectivity.

来源:Neurosurgery,查看原文。版权归原作者所有。