脑室腹腔分流引流随重力与脑脊液压力搏动增加:台式模型研究
该研究利用刚性假脑室台式阀门测试平台,在模拟生理条件下测量差压分流阀的流量。结果显示搏动压力可从脑室传递至阀门,较低顺应性和较高搏动发生器力度导致假脑室内搏动压力升高,重力与较高搏动压力均导致流量增加,阀门可减轻但不能消除这些高流量。结论提示分流阀易发生重力依赖性过度引流,搏动压力在水平和垂直位置均增加流出且依赖顺应性。
要点
- 搏动压力可从脑室传递至分流阀,较低顺应性和较高搏动力度使搏动压力升高。
- 重力与较高搏动压力均导致流量增加,阀门可减轻但不能消除高流量。
- 过度引流不限于垂直位置,水平位置下生理性搏动压力同样增加流出。
核心结论
- 分流阀易发生重力依赖性过度引流,搏动压力在水平和垂直位置均增加流出且依赖顺应性。
研究设计与样本
- 体外台式研究,使用由刚性假脑室、顺应性腔、搏动发生器和压力传感器组成的假脑室台式阀门测试平台。
- 在模拟生理条件下测量差压分流阀流量,条件包括方向(水平/垂直)、顺应性(低/中/高)和搏动发生器力度(低/中/高)。
主要发现
- 搏动压力从脑室忠实传递至阀门。
- 较低顺应性和较高搏动发生器力度导致假脑室内搏动压力升高。
- 重力和较高搏动压力均导致流量增加。
- 阀门的存在减轻但不能消除这些较高流量。
临床怎么用
- 提示临床关注分流阀在水平和垂直位置下的搏动相关过度引流风险。
- 为开发重力阀和抗虹吸装置之外的新型阀门提供生理参数理解基础。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: There have been few improvements in cerebrospinal fluid (CSF) shunt technology since John Holter introduced the silicon valve, with overdrainage remaining a major source of complications. OBJECTIVE: To better understand why valves are afflicted by supra-normal CSF flow rates. We present in Vitro benchtop analyses of flow through a differential pressure valve under simulated physiological conditions. METHODS: The pseudo-ventricle benchtop valve testing platform that comprises a rigid pseudo-ventricle, compliance chamber, pulsation generator, and pressure sensors was used to measure flow rates through a differential pressure shunt valve under the following simulated physiological conditions: orientation (horizontal/vertical), compliance (low/medium/high), and pulsation generator force (low/medium/high). RESULTS: Our data show that pulse pressures are faithfully transmitted from the ventricle to the valve, that lower compliance and higher pulse generator forces lead to higher pulse pressures in the pseudo-ventricle, and that both gravity and higher pulse pressure lead to higher flow rates. The presence of a valve mitigates but does not eliminate these higher flow rates. CONCLUSION: Shunt valves are prone to gravity-dependent overdrainage, which has motivated the development of gravitational valves and antisiphon devices. This study shows that overdrainage is not limited to the vertical position but that pulse pressures that simulate rhythmic (eg, cardiac) and provoked (eg, Valsalva) physiological CSF pulsations increase outflow in both the horizontal and vertical positions and are dependent on compliance. A deeper understanding of the physiological parameters that affect intracranial pressure and flow through shunt systems is prerequisite to the development of novel valves.
引用
- Koueik J, Iskandar BJ, Yang Z, et al. Ventriculoperitoneal Shunt Drainage Increases With Gravity and Cerebrospinal Fluid Pressure Pulsations: Benchtop Model. Neurosurgery. 2021. PMID: 34528096. doi:10.1093/neuros/nyab336
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。