影响MRgFUS丘脑毁损术治疗特发性震颤的技术与手术因素:250例经验
本研究回顾性分析250例单侧MRgFUS丘脑毁损术治疗特发性震颤的患者数据,评估12个月疗效与安全性及相关因素。结果显示73.6%患者12个月时CRST子评分<10,年轻年龄和较高Tmax与震颤改善相关,Tmax预测CRST<10的cutoff值为55.8°C。高温组平衡障碍发生率高于低温组。
要点
- 年轻年龄和较高Tmax与12个月震颤改善独立相关。
- Tmax预测CRST子评分<10的cutoff值为55.8°C。
- 高温组平衡障碍发生率显著高于低温组。
核心结论
- 年轻年龄和较高Tmax与单侧MRgFUS丘脑毁损术后12个月震颤改善相关。
- Tmax预测CRST子评分<10的cutoff值为55.8°C。
- 高温组平衡障碍发生率高于低温组。
研究设计与样本
- 回顾性分析2013年10月至2019年1月三个地理区域的单侧MRgFUS丘脑毁损术数据,共250例特发性震颤患者。
- 疗效基于治疗后1年CRST子评分改善,安全性基于中重度丘脑毁损相关不良事件发生率。
主要发现
- 250例中184例(73.6%)在12个月随访时CRST子评分<10,即无或极轻微震颤相关残疾。
- 多变量分析显示年轻年龄(OR 0.948,p=0.013)和较高Tmax(OR 1.188,p=0.025)与震颤改善相关。
- 多变量分析中未发现与Tmax独立相关的单一显著因素。
- 颅骨密度比(SDR)与加热效率正相关(β=0.005,p<0.001),但与震颤改善无显著关系。
- 低温组(52°C≤Tmax≤54°C)对右侧靶点重复1-3次即可在随访期内产生持续震颤抑制;高温组平衡障碍发生率高于低温组(p=0.04)。
关键数值与推荐
- CRST子评分<10定义为无或极轻微震颤相关残疾。
- Tmax预测CRST子评分<10的cutoff值为55.8°C。
- 低温组条件为52°C≤Tmax≤54°C,对右侧靶点重复1-3次。
临床怎么用
- 术前可参考年龄和Tmax目标值(cutoff 55.8°C)优化患者筛选和手术终点设定。
- 在保证疗效的前提下,可考虑采用低温方案(52-54°C重复1-3次)以降低平衡障碍风险。
- SDR可用于评估加热效率,但不应作为震颤改善的预测指标。
局限性
- 回顾性研究设计。
- 数据来源于原始关键研究和继续获取研究,非前瞻性统一方案。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Magnetic resonance imaging-guided focused ultrasound (MRgFUS) provides real-time monitoring of patients to assess tremor control and document any adverse effects. MRgFUS of the ventral intermediate nucleus (VIM) of the thalamus has become an effective treatment option for medically intractable essential tremor (ET). The aim of this study was to analyze the correlations of clinical and technical parameters with 12-month outcomes after unilateral MRgFUS thalamotomy for ET to help guide future clinical treatments. METHODS: From October 2013 to January 2019, data on unilateral MRgFUS thalamotomy from the original pivotal study and continued-access studies from three different geographic regions were collected. Authors of the present study retrospectively reviewed those data and evaluated the efficacy of the procedure on the basis of improvement in the Clinical Rating Scale for Tremor (CRST) subscore at 1 year posttreatment. Safety was based on the rates of moderate and severe thalamotomy-related adverse events. Treatment outcomes in relation to various patient- and sonication-related parameters were analyzed in a large cohort of patients with ET. RESULTS: In total, 250 patients were included in the present analysis. Improvement was sustained throughout the 12-month follow-up period, and 184 (73.6%) of 250 patients had minimal or no disability due to tremor (CRST subscore < 10) at the 12-month follow-up. Younger age and higher focal temperature (Tmax) correlated with tremor improvement in the multivariate analysis (OR 0.948, p = 0.013; OR 1.188, p = 0.025; respectively). However, no single statistically significant factor correlated with Tmax in the multivariate analysis. The cutoff value of Tmax in predicting a CRST subscore < 10 was 55.8°C. Skull density ratio (SDR) was positively correlated with heating efficiency (β = 0.005, p < 0.001), but no significant relationship with tremor improvement was observed. In the low-temperature group, 1-3 repetitions to the right target with 52°C ≤ Tmax ≤ 54°C was sufficient to generate sustained tremor suppression within the investigated follow-up period. The high-temperature group had a higher rate of balance disturbances than the low-temperature group (p = 0.04). CONCLUSIONS: The authors analyzed the data of 250 patients with the aim of improving practices for patient screening and determining treatment endpoints. These results may improve the safety, efficacy, and efficiency of MRgFUS thalamotomy for ET.