术中影像时代脑深部电刺激靶点获取或可不再依赖电生理定位
该回顾性研究分析50例患者61根需电生理定位(EM)引导重新定位的DBS电极,发现74%的重新定位是朝初始靶点方向移动,而非纠正靶点选择错误。术前计划与靶点接触坐标的平均径向误差从2.2 mm降至1.5 mm(P < 0.001)。结论认为EM主要纠正电极放置误差,术中影像和立体定向技术改进可能减少甚至消除对EM的依赖。
要点
- EM引导的重新定位多数是纠正电极放置误差而非靶点选择错误。
- 微电极记录比测试刺激更常引导电极朝初始靶点移动(88% vs. 63%)。
- 术中影像和立体定向技术改进可能减少甚至消除对EM的依赖。
核心结论
- 电生理定位主要纠正电极放置误差,而非提供靶点选择错误的新信息。
- 术中影像和立体定向技术改进可能减少甚至消除对电生理定位的依赖。
研究设计与样本
- 回顾性研究,纳入50例患者共61根需电生理定位引导重新定位的DBS电极。
- 通过术中CT确定每根电极的立体定向坐标,评估重新定位方向是远离还是朝向初始靶点。
主要发现
- 61根电极中45根(74%)的电生理引导重新定位使电极朝初始靶点移动。
- 术前计划与靶点接触坐标的平均径向误差在重新定位前为2.2 mm,重新定位后为1.5 mm(P < 0.001)。
- 微电极记录比测试刺激更常引导电极朝初始靶点移动(88% vs. 63%;P = 0.03)。
- 所针对的核团与朝初始靶点移动的可能性相关。
临床怎么用
- 在具备术中影像验证电极位置的条件时,可重新评估电生理定位的必要性。
- 电生理定位引导的重新定位多提示需纠正立体定向放置误差,而非靶点选择错误。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Electrophysiologic mapping (EM) has been instrumental in advancing neuroscience and ensuring accurate lead placement for deep brain stimulation. However, EM is associated with increased operative time, expense, and potential risk. Intraoperative imaging to verify lead placement provides an opportunity to reassess the clinical role of EM. We investigated whether EM 1) provides new information that corrects suboptimal preoperative target selection by the physician or 2) simply corrects intraoperative stereotactic error, which can instead be quickly corrected with intraoperative imaging. METHODS: Deep brain stimulation lead location errors were evaluated by measuring whether repositioning leads based on EM directed the final lead placement 1) away from or 2) toward the original target. We retrospectively identified 50 patients with 61 leads that required repositioning directed by EM. The stereotactic coordinates of each lead were determined with intraoperative computed tomography. RESULTS: In 45 of 61 leads (74%), the electrophysiologically directed repositioning moved the lead toward the initial target. The mean radial errors between the preoperative plan and targeted contact coordinates before and after repositioning were 2.2 and 1.5 mm, respectively (P < 0.001). Microelectrode recording was more likely than test stimulation to direct leads toward the initial target (88% vs. 63%; P = 0.03). The nucleus targeted was associated with the likelihood of moving toward the initial target. CONCLUSIONS: Electrophysiologic mapping corrected primarily for errors in lead placement rather than providing new information regarding errors in target selection. Thus, intraoperative imaging and improvements in stereotactic techniques may reduce or even eliminate dependence on EM.
引用
- Zavala B, Mirzadeh Z, Chen T, et al. Electrophysiologic Mapping for Target Acquisition in Deep Brain Stimulation May Become Unnecessary in the Era of Intraoperative Imaging. World Neurosurg. 2021. PMID: 33905908. doi:10.1016/j.wneu.2021.04.069
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。