儿童分流过度引流:胸部植入可调重力阀的替代方案
结果显示年并发症率<0.02,随访期间无导管断开,提示胸部植入是儿童过度引流可行且稳定的分流改良方式。
要点
- 胸部植入可调重力阀可作为儿童分流过度引流的可行分流改良方式。
- 该技术操作简单,装置植入耗时不超过15分钟。
核心结论
- 胸部植入可调重力装置是儿童过度引流患者合适的分流改良方式。
主要发现
- 平均植入年龄9.62岁,平均随访38个月。
- 平均压力调整次数2.48次。
- 装置相对身体纵轴的「偏差角」平均5.8°。
- 年分流并发症率<0.02,随访期间无导管断开。
临床怎么用
- 对儿童分流过度引流,尤其顶枕分流者,可考虑胸部植入可调重力阀,操作简单且耗时不超过15分钟。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Shunt overdrainage is a potential complication in pediatric hydrocephalus. The addition of adjustable gravitational units to previous shunt systems has been proposed as effective management for this problem. These devices have been traditionally implanted over the occipital bone. We propose chest implantation as an easier, safer, and more stable alternative in the pediatric population, especially in those cases with parieto-occipital shunts. METHODS: This study comprises a retrospective analysis from a unicentric case series of pediatric patients affected by overdrainage and managed with adjustable gravitational valves implanted in the chest. The device implantation technique is described in detail and takes no more than 15 minutes. RESULTS: Thirty-seven patients met the criteria. The mean age of implantation was 9.62 years. The mean follow-up in the series was 38 months. The mean number of pressure adjustments was 2.48. The mean "deviation angle" of the device to the longitudinal body axis was 5.8°. The complications per year of shunt were <0.02 with no disconnection of the catheters in any case during follow-up. CONCLUSIONS: In our experience, chest implantation for adjustable gravitational devices was a suitable shunt modification in pediatric patients suffering from overdrainage.
引用
- Tirado-Caballero J, Rivero-Garvia M, Moreno-Madueño G, et al. Chest Implantation of Adjustable Gravitational Valves: An Easy, Safe, and Stable Alternative to Control Symptomatic Overdrainage in Shunted Children. World Neurosurg. 2020. PMID: 33171320. doi:10.1016/j.wneu.2020.11.001
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。