三维打印颅内动脉瘤缺损模型比较封堵球囊性能

Comparing Occlusive Balloon Performance Using 3-Dimensional Printed Models of Intracranial Aneurysmal Defects.
PubMed 文献 · World Neurosurg · 2020-11-13

研究用Tinkercad设计并以聚乳酸三维打印含动脉瘤缺损的模型血管,在可达生理压力的加压箱中于透视下测试Hyperglide、Transform与Scepter C三种顺应性重塑球囊,记录名义与超名义充盈时球囊最大直径及内部压力,并用多变量线性回归建模球囊顺应性。结果显示Scepter球囊顺应性显著高于Hyperglide与Transform(P<0.001),在较大缺损中顺应性更高,且性能差异与血管尺寸相关。

要点

  • Scepter球囊顺应性显著高于Hyperglide与Transform(P<0.001)。
  • Scepter在较大缺损中顺应性高于其他球囊,性能差异与血管尺寸相关。
  • 在较小血管直径的较大瘤颈中,Scepter变形所需压力更低。

核心结论

  • Scepter球囊通过模型缺损时所需压力低于对照球囊,尤其在较小血管直径的较大动脉瘤颈中。
  • 该结果可帮助术者在脆弱脑动脉瘤中减少球囊压力时选择球囊类型与尺寸。

研究设计与样本

  • 体外研究:用Tinkercad创建含动脉瘤缺损的模型血管,以聚乳酸经Ultimaker2三维打印。
  • 构建可达生理压力的加压箱容纳模型血管,于透视下测试3家厂商的顺应性重塑球囊。
  • 测试球囊为Hyperglide、Transform与Scepter C,每个缺损处分别充盈至名义与超名义点。
  • 每次充盈记录球囊最大直径与内部球囊压力,采用Phillips Intellivue监测内部压力。
  • 采用多变量线性回归分析建模球囊顺应性(球囊直径除以压力)。

主要发现

  • 多变量回归建模显示Scepter球囊顺应性显著高于Hyperglide与Transform(P<0.001)。
  • Scepter球囊在较大缺损中的顺应性高于其他类型球囊。
  • 球囊性能差异与血管尺寸相关。

临床怎么用

  • 在脆弱脑动脉瘤需尽量减少球囊压力时,可参考本研究结果选择球囊类型与尺寸。
  • 对于较小血管直径中的较大瘤颈,Scepter球囊变形所需压力较低。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Balloon remodeling microcatheters are essential in assisting endovascular coiling of brain aneurysms, but the performance and pressure requirements of different balloon types when used in aneurysmal defects are currently unknown. METHODS: We used Tinkercad (Autodesk, Montreal, Quebec) to create model vessels with aneurysmal defects and 3-dimensionally printed these designs with polylactic acid using the Ultimaker2 (Ultimaker, Geldermalsen, Netherlands). We constructed a pressurized box capable of reaching physiologic pressures that housed our vessels and then tested compliant remodeling balloons under fluoroscopy from 3 manufacturers: Hyperglide (Medtronic, Minneapolis, Minnesota, USA), Transform (Stryker Neurovascular, Fremont, California, USA), and Scepter C (Microvention, Tustin, California, USA). Each balloon was inflated to a nominal and supranominal point at each defect, and at each inflation the maximum diameter of the balloon and internal balloon pressure were recorded. The Phillips Intellivue (Phillips, Amsterdam, The Netherlands) monitor was adapted for internal balloon pressure monitoring. A multivariate linear regression analysis was performed to model balloon compliance (balloon diameter divided by pressure). RESULTS: Multivariate regression modeling demonstrated the Scepter balloon had significantly greater compliance compared with both the Hyperglide and Transform balloons (P < 0.001). In addition, we found that Scepter balloons had higher compliance in larger defects compared with the other types of balloons and performance differences based on vessel size. CONCLUSIONS: Scepter balloons require less pressure compared with their counterparts to adequately deform through model defects, specifically in larger aneurysm necks in smaller vessel diameters. This result could inform operators of optimal balloon type and size when trying to minimize balloon pressure in fragile brain aneurysms.