深部脑刺激器患者进行3特斯拉磁共振成像:体模研究与初步患者研究结果
该研究通过人体仿真体模测试激活状态下DBS设备在3T MRI中的温度和电压安全性,并基于此开展6例患者队列研究。体模中所有序列温度升高低于2°C,射频脉冲可调制DBS电压幅度;6例患者激活状态下行MRI未发生严重不良事件,1例在T1-MPRAGE扫描中出现刺激相关副作用。结论认为经仔细体模安全测试后,激活DBS患者可安全进行3T结构和功能MRI。
要点
- 体模测试中DBS激活与未激活状态下所有序列温度升高均低于2°C。
- 射频脉冲可调制DBS电输出,导致电压幅度轻微波动。
- 6例患者激活DBS下行3T MRI无严重不良事件,1例出现刺激相关副作用。
核心结论
- 经仔细体模安全测试后,激活深部脑刺激器患者可安全进行3T结构和功能MRI。
研究设计与样本
- 采用人体仿真3D打印体模对激活DBS设备进行3T MRI下温度和电压安全测试,并基于结果开展6例患者队列研究。
主要发现
- 体模研究中DBS激活和未激活状态下所有序列温度升高均低于2°C。
- MRI射频脉冲可调制DBS电输出,导致电压幅度轻微波动。
- 6例患者激活DBS下行MRI未发生严重不良事件。
- 1例患者在DBS激活状态行T1-MPRAGE扫描时因射频诱导电压幅度调制出现刺激相关副作用。
关键数值与推荐
- 体模测试中所有序列温度升高低于2°C(DBS激活与未激活状态)。
- 患者队列共6例,DBS激活状态下行MRI,无严重不良事件。
临床怎么用
- 在体模安全测试基础上,可考虑对激活DBS患者进行3T结构和功能MRI。
- 扫描中需关注射频诱导的电压幅度调制可能引起刺激相关副作用。
局限性
- 患者队列仅6例,样本量小。
- 基于单一 anthropomorphic 体模设计的安全测试结果。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Deep brain stimulation (DBS) is a standard of care treatment for multiple neurologic disorders. Although 3-tesla (3T) magnetic resonance imaging (MRI) has become the gold-standard modality for structural and functional imaging, most centers refrain from 3T imaging in patients with DBS devices in place because of safety concerns. 3T MRI could be used not only for structural imaging, but also for functional MRI to study the effects of DBS on neurocircuitry and optimize programming. OBJECTIVE: To use an anthropomorphic phantom design to perform temperature and voltage safety testing on an activated DBS device during 3T imaging. METHODS: An anthropomorphic 3D-printed human phantom was constructed and used to perform temperature and voltage testing on a DBS device during 3T MRI. Based on the phantom assessment, a cohort study was conducted in which 6 human patients underwent MRI with their DBS device in an activated (ON) state. RESULTS: During the phantom study, temperature rises were under 2°C during all sequences, with the DBS in both the deactivated and activated states. Radiofrequency pulses from the MRI appeared to modulate the electrical discharge from the DBS, resulting in slight fluctuations of voltage amplitude. Six human subjects underwent MRI with their DBS in an activated state without any serious adverse events. One patient experienced stimulation-related side effects during T1-MPRAGE scanning with the DBS in an ON state because of radiofrequency-induced modulation of voltage amplitude. CONCLUSION: Following careful phantom-based safety testing, 3T structural and functional MRI can be safely performed in subjects with activated deep brain stimulators.
引用
- Davidson B, Tam F, Yang B, et al. Three-Tesla Magnetic Resonance Imaging of Patients With Deep Brain Stimulators: Results From a Phantom Study and a Pilot Study in Patients. Neurosurgery. 2021. PMID: 33045736. doi:10.1093/neuros/nyaa439
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。