基于组织激活体积软件选择丘脑底核刺激触点
该研究分析26例接受丘脑底核脑深部电刺激的帕金森病患者,用组织激活体积(VTA)软件由两名对临床数据设盲的专家选择触点并确定治疗窗,与患者实际参数比较。结果显示90.4%(95%CI 82%-98%)的患者VTA软件识别触点与临床有效触点一致,虚拟治疗窗涵盖81.3%的临床刺激幅度。
要点
- VTA软件选触点与临床有效触点一致性达90.4%(95%CI 82%-98%)。
- 虚拟治疗窗涵盖81.3%的临床刺激幅度。
- VTA软件可辅助术后随访与程控。
核心结论
- VTA软件在丘脑底核脑深部电刺激触点与刺激参数选择上与临床数据呈现显著一致性,可能有助于术后随访和程控。
研究设计与样本
- 单中心研究,分析26例接受丘脑底核脑深部电刺激的帕金森病患者的脑影像。
- 由两名对临床数据设盲的脑深部电刺激专家用VTA软件选择触点并确定幅度治疗窗。
- 术前采集临床与神经心理数据,术后3个月和1年获取结果并与患者实际参数比较。
主要发现
- 90.4%(95%CI 82%-98%)的患者中,VTA软件识别的触点与临床有效触点或逐触点测试中的有效触点一致。
- 虚拟选择的幅度治疗窗涵盖81.3%的临床刺激幅度。
临床怎么用
- 术后程控时可借助VTA软件可视化触点位置与患者解剖关系,辅助触点与刺激幅度选择。
局限性
- 样本量仅26例,随访至术后1年。
- 专家选触点基于设盲条件,与真实程控流程存在差异。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: High-frequency deep brain stimulation (DBS) of the subthalamic nucleus (STN) is effective in the treatment of motor symptoms of Parkinson's disease. Using a patient-specific lead and volume of tissue activated (VTA) software, it is possible to visualize contact positions in the context of the patient's own anatomy. In this study, the authors' aim was to demonstrate that VTA software can be used in clinical practice to help determine the clinical effectiveness of stimulation in patients with Parkinson's disease undergoing DBS of the STN. METHODS: Brain images of 26 patients undergoing STN DBS were analyzed using VTA software. Preoperative clinical and neuropsychological data were collected. Contacts were chosen by two experts in DBS blinded to the clinical data. A therapeutic window of amplitude was determined. These results were compared with the parameter settings for each patient. Data were obtained at 3 months and 1 year postsurgery. RESULTS: In 90.4% (95% CI 82%-98%) of the patients, the contacts identified by the VTA software were concordant with the clinically effective contacts or with an effective contact in contact-by-contact testing. The therapeutic window of amplitude selected virtually included 81.3% of the clinical amplitudes. CONCLUSIONS: VTA software appears to present significant concordance with clinical data for selecting contacts and stimulation parameters that could help in postoperative follow-up and programming.
引用
- Devaluez M, Tir M, Krystkowiak P, et al. Selection of deep brain stimulation contacts using volume of tissue activated software following subthalamic nucleus stimulation. J Neurosurg. 2020. PMID: 33096524. doi:10.3171/2020.6.JNS192157
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。