文献速递功能神经外科

迷走神经刺激器更换术中高阻抗需修正的案例回顾

High lead impedances requiring revision during vagal nerve stimulator generator replacement.
PubMed 文献 · Acta Neurochir (Wien) · 2020-09-24

该回顾性研究分析了2009至2018年间570例因难治性癫痫行VNS手术的病例,其中338例为修正手术。术中发现意外高阻抗10例(3%),导致手术复杂度显著增加。作者提醒在更换刺激器时应预判可能需要更广泛的手术操作。

要点

  • VNS修正手术中意外高阻抗发生率约3%。
  • 高阻抗原因包括断开、电极断裂、硬件故障或组织瘢痕。
  • 术前应告知患者可能需要超出简单发生器更换的手术。

核心结论

  • VNS术中意外高阻抗需扩大手术范围,应提前预判。

研究设计与样本

  • 回顾性分析2009年1月至2018年9月570例VNS手术记录,含232例初次植入和338例修正手术。

主要发现

  • 全部病例均为难治性癫痫;术中发现意外高阻抗10例(3%),导致手术复杂度显著增加。

关键数值与推荐

  • 高阻抗可能源于断开、电极断裂、硬件故障或组织瘢痕,最终需更广泛手术。

临床怎么用

  • 在VNS发生器更换术前应预判可能需行更广泛手术,以减少术中困扰。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Vagal nerve stimulation (VNS) therapy is among the growing options in the treatment of intractable epilepsy. The phenomenon of surprise lead impedance issues found at the time of surgery resulting in unplanned lead revision is a challenge with this type of device. We reviewed our experience with VNS revisions. MATERIAL AND METHODS: We retrospectively reviewed the records of all adult and pediatric patients between January 2009 and September 2018 who underwent surgery for VNS therapy, including revision surgery. Office and operative notes were reviewed to obtain the indications and operative details for VNS placement. RESULTS: A total of 570 operations were reviewed. The indication was intractable epilepsy in all cases. Primary implantation was performed in 232 patients, while the remaining 338 cases were revision cases of various natures. Surprise high lead impedance was found in 10 (3%) of these cases, resulting in a significantly increased complexity of surgery in those instances. CONCLUSION: Lead impedance issues can be caused by disconnection, electrode fracture, hardware failure, or tissue scarring but ultimately require a more extended surgery than may be initially planned. Anticipating the potential for a more extensive operation than a simple generator replacement may prevent perioperative frustrations on both sides.

来源:Acta Neurochir (Wien),查看原文。版权归原作者所有。