硬膜外心电门控振荡颅内植入物调节脑血流的动物研究

Epidural Oscillating Cardiac-Gated Intracranial Implant Modulates Cerebral Blood Flow.
PubMed 文献 · Neurosurgery · 2020-01-01

该研究在13只犬中植入硬膜外球囊导管并连接心电门控振荡泵,比较ICP脉冲增强、减弱和反转协议对脑血流的影响。结果显示ICP脉冲增强使同侧/对侧脉冲幅度分别增加217%/202%,而反转和减弱协议可增加CBF(28%/7.4%和2.4%/1.3%),但增强协议未增加CBF;CBF变化与ICP脉冲幅度和收缩峰变化相关,与平均ICP或全身变量无关。

要点

  • 心电门控硬膜外球囊振荡可差异调节ICP脉冲波形并改变CBF。
  • ICP脉冲反转和减弱协议增加CBF,增强协议未增加CBF。
  • CBF变化与ICP脉冲幅度及收缩峰变化相关,非平均ICP或全身变量。

核心结论

  • 心电门控操纵ICP搏动性可用于研究颅内搏动动力学,并可能成为改变CBF的手段。

研究设计与样本

  • 研究纳入13只犬,单侧硬膜外球囊植入于外侧裂上方,连接心电门控振荡泵,比较ICP脉冲增强、减弱和反转协议,并直接对比系统激活与失活时间段。

主要发现

  • ICP脉冲增强使同侧/对侧脉冲幅度分别增加217%/202%(P < .0005),反转协议使同侧/对侧分别增加139%/120%(P < .0005)。
  • ICP均值在增强协议下同侧/对侧分别增加23%/31%(P < .0001),反转协议下分别降低7%/11%(P = .006/.0003),减弱协议下分别降低4%/5%(P < .0005)。
  • 反转和减弱协议下CBF增加:反转协议同侧/对侧分别增加28%/7.4%(P < .0001/P = .006),减弱协议同侧/对侧分别增加2.4%/1.3%(P < .0001/P = .003);增强协议未观察到CBF增加。
  • CBF变化与ICP脉冲幅度和收缩峰变化相关,与平均ICP变化或全身变量(心率、动脉血压)无关。

临床怎么用

  • 该技术为研究颅内搏动动力学提供了可控工具,并提示通过调节ICP脉冲波形可能影响CBF,但尚未进入临床决策。

局限性

  • 研究为犬类动物实验,样本量13只,结果向人类临床转化的有效性尚未验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: We have previously reported a method and device capable of manipulating ICP pulsatility while minimally effecting mean ICP. OBJECTIVE: To test the hypothesis that different modulations of the intracranial pressure (ICP) pulse waveform will have a differential effect on cerebral blood flow (CBF). METHODS: Using an epidural balloon catheter attached to a cardiac-gated oscillating pump, 13 canine subjects underwent ICP waveform manipulation comparing different sequences of oscillation in successive animals. The epidural balloon was implanted unilaterally superior to the Sylvian sulcus. Subjects underwent ICP pulse augmentation, reduction and inversion protocols, directly comparing time segments of system activation and deactivation. ICP and CBF were measured bilaterally along with systemic pressure and heart rate. CBF was measured using both thermal diffusion, and laser doppler probes. RESULTS: The activation of the cardiac-gate balloon implant resulted in an ipsilateral/contralateral ICP pulse amplitude increase with augmentation (217%/202% respectively, P < .0005) and inversion (139%/120%, P < .0005). The observed changes associated with the ICP mean values were smaller, increasing with augmentation (23%/31%, P < .0001) while decreasing with inversion (7%/11%, P = .006/.0003) and reduction (4%/5%, P < .0005). CBF increase was observed for both inversion and reduction protocols (28%/7.4%, P < .0001/P = .006 and 2.4%/1.3%, P < .0001/P = .003), but not the augmentation protocol. The change in CBF was correlated with ICP pulse amplitude and systolic peak changes and not with change in mean ICP or systemic variables (heart rate, arterial blood pressure). CONCLUSION: Cardiac-gated manipulation of ICP pulsatility allows the study of intracranial pulsatile dynamics and provides a potential means of altering CBF.