颅内压对分流阀压力设置调整的矛盾反应

Paradoxical response of intracranial pressure to shunt valve setting adjustments.
PubMed 文献 · Acta Neurochir (Wien) · 2020-06-24

结果显示57%调整后颅内压反应符合预期,18%出现矛盾反应,24%无效应(夜间颅内压变化<1 mmHg),且夜间颅内压变化与阀门设置变化幅度在Miethke ProGAV(p=0.01)和Medtronic Strata(p=0.02)中均显著相关。

要点

  • 分流阀调整后18%出现颅内压矛盾反应,需警惕。
  • 24%调整无效应,夜间颅内压变化<1 mmHg。
  • 夜间颅内压变化与阀门设置变化幅度显著相关。

核心结论

  • 分流阀设置调整后可出现颅内压矛盾反应,在颅内压引导的阀门调整中需考虑此现象,并对未来智能分流系统开发有重要意义。

研究设计与样本

  • 单中心回顾性观察研究,纳入2014至2019年接受24小时持续颅内压监测下分流阀设置调整的患者,排除疑似分流故障者。

主要发现

  • 基础疾病最常见为脑积水(47%)或特发性颅内高压(43%)。
  • 颅内压反应分类:预期54例(57%)、矛盾17例(18%)、无效应(夜间颅内压变化<1 mmHg)23例(24%)。
  • 夜间颅内压变化与阀门设置变化幅度在Miethke ProGAV(p=0.01)和Medtronic Strata(p=0.02)中均显著相关。

临床怎么用

  • 进行颅内压引导的分流阀调整时,应考虑到矛盾颅内压反应的可能。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The hydrodynamics of cerebrospinal fluid shunts have been described in vitro; however, knowledge on the response of intracranial pressure (ICP) to valve settings adjustments in vivo is limited. This study describes the effect of adjusting the shunt valve setting on ICP in a cohort of patients with complex symptom management. METHOD: Single-centre retrospective observational study. Patients who underwent ICP-guided valve setting adjustments during 24-h continuous ICP monitoring, between 2014 and 2019, were included. Patients with suspected shunt malfunction were excluded. Median night ICP before and after the valve adjustments were compared (Δ night ICP). The responses of ICP to valve adjustment were divided into 3 different groups as follows: expected, paradoxical and no response. The frequency of the paradoxical response and its potential predicting factors were investigated. RESULTS: Fifty-one patients (37 females, 14 males, mean age 38 years) receiving 94 valve setting adjustments met the study inclusion criteria. Patients' underlying conditions were most commonly hydrocephalus (47%) or idiopathic intracranial hypertension (43%). The response of ICP to valve setting adjustments was classified as 'expected' in 54 cases (57%), 'paradoxical' in 17 cases (18%) and 'no effect' (Δ night ICP < 1 mmHg) in 23 cases (24%). There was a significant correlation between the Δ night ICP and the magnitude of valve setting change in both the investigated valves (Miethke ProGAV, p = 0.01 and Medtronic Strata, p = 0.02). CONCLUSIONS: Paradoxical ICP changes can occur after shunt valve setting adjustments. This observation should be taken into account when performing ICP-guided valve adjustments and is highly relevant for the future development of "smart" shunt systems.

来源:Acta Neurochir (Wien),查看原文。版权归原作者所有。

引用

  1. D'Antona L, Craven CL, Merchan MAJ, et al. Paradoxical response of intracranial pressure to shunt valve setting adjustments. Acta Neurochir (Wien). 2020. PMID: 32583083. doi:10.1007/s00701-020-04462-y

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。