无创热流评估脑室腹腔分流管故障的诊断准确性:前瞻性多中心操作者盲法研究

Diagnostic Accuracy of Non-Invasive Thermal Evaluation of Ventriculoperitoneal Shunt Flow in Shunt Malfunction: A Prospective, Multi-Site, Operator-Blinded Study.
PubMed 文献 · Neurosurgery · 2020-01-01

本研究为前瞻性多中心操作者盲法试验,纳入406例急诊疑似分流管故障患者,比较两种热流评估方案(TFE+MP与TFE-only)与神经影像对分流管梗阻的诊断准确性。主要结局为7天内需神经外科翻修的分流管梗阻。结果显示TFE在排除分流管故障方面不劣于影像,且TFE-only优于TFE+MP。

要点

  • TFE在排除分流管故障方面不劣于神经影像,可帮助避免影像检查及其他步骤。
  • 对于排除分流管故障,TFE-only方案优于TFE+MP方案。
  • TFE-only检测梗阻的敏感性为100%(95%CI:88%-100%)。

核心结论

  • TFE在排除分流管故障方面不劣于影像检查,可能有助于避免影像及其他步骤。
  • 用于此目的时,TFE-only优于TFE+MP。

研究设计与样本

  • 前瞻性多中心操作者盲法试验,连续纳入因疑似分流管故障接受评估的患者。
  • 共纳入406例患者,来自10个中心。
  • 主要结局为7天内需神经外科翻修的分流管梗阻。
  • 采用预先设定的非劣效界值-2.5%检验TFE与神经影像检测分流管梗阻敏感性的非劣效性。

主要发现

  • TFE+MP组68/348(20%)存在分流管梗阻,TFE-only组30/215(14%)存在分流管梗阻。
  • TFE-only检测梗阻的敏感性为100%(95%CI:88%-100%)。
  • TFE+MP检测梗阻的敏感性为90%(95%CI:80%-96%)。
  • TFE+MP队列中影像检测梗阻的敏感性为76%(95%CI:65%-86%)。
  • TFE-only队列中影像检测梗阻的敏感性为77%(95%CI:58%-90%)。

临床怎么用

  • 对急诊疑似分流管故障患者,可优先采用TFE-only方案排除分流管梗阻。
  • TFE结果可帮助判断是否需进一步影像检查及其他步骤。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Thermal flow evaluation (TFE) is a non-invasive method to assess ventriculoperitoneal shunt function. Flow detected by TFE is a negative predictor of the need for revision surgery. Further optimization of testing protocols, evaluation in multiple centers, and integration with clinical and imaging impressions prompted the current study. OBJECTIVE: To compare the diagnostic accuracy of 2 TFE protocols, with micropumper (TFE+MP) or without (TFE-only), to neuro-imaging in patients emergently presenting with symptoms concerning for shunt malfunction. METHODS: We performed a prospective multicenter operator-blinded trial of a consecutive series of patients who underwent evaluation for shunt malfunction. TFE was performed, and preimaging clinician impressions and imaging results were recorded. The primary outcome was shunt obstruction requiring neurosurgical revision within 7 d. Non-inferiority of the sensitivity of TFE vs neuro-imaging for detecting shunt obstruction was tested using a prospectively determined a priori margin of -2.5%. RESULTS: We enrolled 406 patients at 10 centers. Of these, 68/348 (20%) evaluated with TFE+MP and 30/215 (14%) with TFE-only had shunt obstruction. The sensitivity for detecting obstruction was 100% (95% CI: 88%-100%) for TFE-only, 90% (95% CI: 80%-96%) for TFE+MP, 76% (95% CI: 65%-86%) for imaging in TFE+MP cohort, and 77% (95% CI: 58%-90%) for imaging in the TFE-only cohort. Difference in sensitivities between TFE methods and imaging did not exceed the non-inferiority margin. CONCLUSION: TFE is non-inferior to imaging in ruling out shunt malfunction and may help avoid imaging and other steps. For this purpose, TFE only is favored over TFE+MP.

来源:Neurosurgery,查看原文。版权归原作者所有。

引用

  1. Madsen JR, Boyle TP, Neuman MI, et al. Diagnostic Accuracy of Non-Invasive Thermal Evaluation of Ventriculoperitoneal Shunt Flow in Shunt Malfunction: A Prospective, Multi-Site, Operator-Blinded Study. Neurosurgery. 2020. PMID: 32459841. doi:10.1093/neuros/nyaa128