基于显示器的3D4k外视镜神经外科干预:新型混合设备的两阶段前瞻性随机临床评估
该研究评估新型3D4k混合外视镜(结合数字与光学可视化)在神经外科中的手术表现与团队流程。方法包括临床前工作坊(n=12)、颅脑与脊柱手术术中探索(n=9)及幕上脑肿瘤随机临床研究(n=20,比较目镜模式与显示器模式)。结果显示显示器模式坐位手术工效学显著改善(P=.03),无显著额外手术时间;当前3D4k显示器图像质量仍逊于光学可视化。
要点
- 显示器模式较目镜模式显著改善坐位手术工效学(P=.03)
- 额叶、额顶或颞部入路脑肿瘤是外视镜手术的有利用例
- 当前3D4k显示器图像质量仍逊于光学可视化,有待更新
核心结论
- 混合外视镜设备可整合入既有神经外科工作流程。
- 当前外视镜干预最适合非深部颅脑肿瘤手术。
- 显示器模式较传统显微手术改善工效学。
研究设计与样本
- 临床前工作坊通过技能场景评估可用性与可实施性,神经外科参与者n=12。
- 颅脑与脊柱手术术中探索n=9。
- 随机临床研究比较幕上脑肿瘤病例的目镜模式与显示器模式,n=20。
- 上述临床评估在12个月内完成。
主要发现
- 额叶、额顶或颞部入路的脑肿瘤被确定为引入外视镜神经外科的有利用例。
- 平均显示器距离与角度分别为180 cm和10°。
- 显示器模式坐位手术工效学较目镜模式显著改善(P=.03)。
- 显示器模式的手眼协调需要适应。
- 临床前数据显示侧向摄像头倾斜与手眼协调影响呈正相关(rs=0.756,P=.01)。
临床怎么用
- 可将混合外视镜设备整合入既有神经外科工作流程。
- 当前外视镜干预最适合非深部颅脑肿瘤手术。
- 显示器模式较传统显微手术改善工效学,但手眼协调需适应。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Promoting a disruptive innovation in microsurgery, exoscopes promise alleviation of physical strain and improved image quality through digital visualization during microneurosurgical interventions. This study investigates the impact of a novel 3D4k hybrid exoscope (i.e., combining digital and optical visualization) on surgical performance and team workflow in preclinical and clinical neurosurgical settings. METHODS: A pre-clinical workshop setting has been developed to assess usability and implementability through skill-based scenarios (neurosurgical participants n = 12). An intraoperative exploration in head and spine surgery (n = 9) and a randomized clinical study comparing ocular and monitor mode in supratentorial brain tumor cases (n = 20) followed within 12 months. Setup, procedure, case characteristics, surgical performance, and user experience have been analyzed for both ocular group (OG) and monitor group (MG). RESULTS: Brain tumor cases using frontal, frontoparietal, or temporal approaches have been identified as favorable use cases for introducing exoscopic neurosurgery. Mean monitor distance and angle were 180 cm and 10°. Surgical ergonomics when sitting improved significantly in MG compared with OG (P = .03). Hand-eye coordination required familiarization in MG. Preclinical data showed a positive correlation between lateral camera inclination and impact on hand-eye coordination (rs = 0.756, P = .01). There was no significant added surgical time in MG. Image quality in current generation 3D4k monitors has been rated inferior to optic visualization yet awaits updates. CONCLUSIONS: The hybrid exoscopic device can be integrated into established neurosurgical workflows. Currently, exoscopic interventions seem most suited for cranial tumor surgery in lesions that are not deep-seated. Ergonomics improve in monitor mode compared to conventional microsurgery.
引用
- Roethe AL, Landgraf P, Schröder T, et al. Monitor-based exoscopic 3D4k neurosurgical interventions: a two-phase prospective-randomized clinical evaluation of a novel hybrid device. Acta Neurochir (Wien). 2020. PMID: 32424568. doi:10.1007/s00701-020-04361-2
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。