文献速递功能神经外科

三维透视(XT)与计算机断层扫描(CT)注册在脑深部电刺激(DBS)手术中的准确性比较

The accuracy of 3D fluoroscopy (XT) vs computed tomography (CT) registration in deep brain stimulation (DBS) surgery.
PubMed 文献 · Acta Neurochir (Wien) · 2020-04-16

本研究比较脑深部电刺激手术中三维透视(XT)与计算机断层扫描(CT)立体定向注册的准确性、效率和辐射暴露。41例患者中CT组25例、XT组16例,两组植入准确性无显著差异,XT组辐射暴露显著低于CT组,注册时间无显著差异。

要点

  • XT组与CT组在x、y、z方向的平均绝对差异均无统计学意义。
  • XT组平均辐射暴露220.0 ± 0.1 mGy*cm,显著低于CT组的1269.3 ± 112.9 mGy*cm(p < 0.001)。
  • CT组注册时间107.8 ± 23.1 min与XT组106.0 ± 18.2 min无显著差异(p = 0.518)。

核心结论

  • 基于XT的框架注册与CT相比可获得相似的植入准确性,且辐射暴露显著更低。
  • 作者指出注册时间无显著差异,但这可能归因于学习曲线效应。

研究设计与样本

  • 本研究为比较XT与CT立体定向注册准确性、效率及辐射暴露的研究,并评估XT与术前MRI融合。
  • 共纳入41例接受DBS手术的患者,其中CT组25例,XT组16例。
  • 准确性通过计算计划(根据术中测试调整)与实际电极轨迹之间的平均绝对差异和欧几里得距离来评估。
  • 每次扫描的辐射剂量以剂量长度乘积(DLP)记录。
  • 效率以患者进入手术室至初始皮肤切口之间的时间衡量。

主要发现

  • CT组与XT组在x方向(p = 0.331)、y方向(p = 0.951)和z方向(p = 0.807)的平均绝对差异均无统计学意义。
  • 两组间的欧几里得距离无显著差异(p = 0.874)。
  • XT组平均辐射暴露为220.0 ± 0.1 mGy*cm,显著低于CT组的1269.3 ± 112.9 mGy*cm(p < 0.001)。
  • CT组注册时间为107.8 ± 23.1 min,XT组为106.0 ± 18.2 min,两组间无显著差异(p = 0.518)。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 作者指出注册时间无显著差异可能归因于学习曲线效应。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Stereotactic registration is the most critical step ensuring accuracy in deep brain stimulation (DBS) surgery. 3D fluoroscopy (XT) is emerging as an alternative to CT. XT has been shown to be safe and effective for intraoperative confirmation of lead position following implantation. However, there is a lack of studies evaluating the suitability of XT to be used for the more crucial step of registration and its capability of being merged to a preoperative MRI. This is the first study comparing accuracy, efficiency, and radiation exposure of XT- vs CT-based stereotactic registration and XT/MRI merging in deep brain stimulation. METHODS: Mean absolute differences and Euclidean distance between planned (adjusted for intraoperative testing) and actual lead trajectories were calculated for accuracy of implantation. The radiation dose from each scan was recorded as the dose length product (DLP). Efficiency was measured as the time between the patient entering the operating room and the initial skin incision. A one-way ANOVA compared these parameters between patients that had either CT- or XT-based registration. RESULTS: Forty-one patients underwent DBS surgery-25 in the CT group and 16 in the XT group. The mean absolute difference between CT and XT was not statistically significant in the x (p = 0.331), y (p = 0.951), or z (p = 0.807) directions. The Euclidean distance between patient groups did not differ significantly (p = 0.874). The average radiation exposure with XT (220.0 ± 0.1 mGy*cm) was significantly lower than CT (1269.3 ± 112.9 mGy*cm) (p < 0.001). There was no significant difference in registration time between CT (107.8 ± 23.1 min) and XT (106.0 ± 18.2 min) (p = 0.518). CONCLUSION: XT-based frame registration was shown to result in similar implantation accuracy and significantly less radiation exposure compared with CT. Our results surprisingly showed no significant difference in registration time, but this may be due to a learning curve effect.

来源:Acta Neurochir (Wien),查看原文。版权归原作者所有。

引用

  1. Cooper MD, Restrepo C, Hill R, et al. The accuracy of 3D fluoroscopy (XT) vs computed tomography (CT) registration in deep brain stimulation (DBS) surgery. Acta Neurochir (Wien). 2020. PMID: 32300988. doi:10.1007/s00701-020-04322-9

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。