本文报告一例32岁因早产致分流依赖性脑积水男性患者,其14岁时唯一一次分流翻修时将新远端导管置于旧导管旁,残留原分流管引起明显不适。采用心胸外科内镜下静脉采集系统经小皮肤切口微创完整取出钙化的脑室腹腔分流管,患者症状完全缓解。作者认为该器械适合微创取出长段残留脑室腹腔分流管,据其所知为首次报道。

要点

  • 残留分流管可因钙化而难以取出,部分患者可出现疼痛不适。
  • 采用心胸外科内镜下静脉采集系统经小皮肤切口微创取出钙化分流管。
  • 患者术后症状完全缓解,作者称为首例微创取出残留脑室腹腔导管的报道。

核心结论

  • 心胸外科使用的内镜下静脉采集系统是经小皮肤切口微创取出长段残留脑室腹腔分流管的合适器械。
  • 据作者所知,这是首例微创取出残留脑室腹腔导管的报道。

研究设计与样本

  • 技术说明,报告一例32岁因早产致分流依赖性脑积水男性患者。
  • 患者14岁时接受唯一一次分流翻修,将新远端导管置于旧导管旁。

主要发现

  • 患者因残留原分流管出现明显不适。
  • 使用内镜下静脉采集系统微创取出分流管后,患者症状完全缓解。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Retained old cerebrospinal fluid diversion shunt catheters in the neck, chest, or abdominal walls are frequently encountered in patients with lifelong shunt-dependent hydrocephalus who have undergone multiple shunt revisions. Particularly in cases where years and decades go between shunt revisions, the distal catheter portion can get calcified and nearly impossible to remove. Most patients tolerate a retained shunt catheter without problems. In some patients, however, retained catheters can cause pain and discomfort, particularly over the clavicle with head movements. Albeit trivial, we are unaware of innovative solutions to this problem. Here, we describe the use of an endoscopic vein harvest device used in cardiothoracic surgery to completely remove an old, calcified shunt catheter. METHODS: Removal of a calcified ventriculoperitoneal shunt catheter using an endoscopic vein harvesting system was performed in a 32-year-old man with shunt-dependent hydrocephalus from premature birth. At 14 years of age, the patient had his only shunt revision consisting of a new distal catheter being placed adjacent to the old catheter. The patient presented with significant discomfort from the retained original shunt catheter. RESULTS: Using the endoscopic vein harvesting system, the shunt catheter was removed minimally invasively and the patient had complete resolution of his symptomatology. CONCLUSIONS: The endoscopic vein harvesting system used in cardiothoracic surgery is a suitable instrument to remove long segments of a retained ventriculoperitoneal shunt catheter minimally invasively through a small skin incision. To our knowledge, this is the first report of minimally invasive removal of a retained ventriculoperitoneal catheter.

来源:World Neurosurg,查看原文。版权归原作者所有。