术中计算机断层扫描与术后磁共振成像确定脑深部电刺激电极坐标的比较
该回顾性研究纳入2015至2017年单中心83例患者共145根脑深部电刺激电极,比较术中CT与术后MRI确定的电极x、y、z坐标绝对差异及欧几里得向量。结果显示x坐标差异无统计学意义,y和z坐标差异显著,术中CT确认x坐标更准确,确认y和z坐标准确性较低。
要点
- 术中CT与术后MRI确定的x坐标平均绝对差异为0.01±0.09 mm(P=0.89),无显著差异。
- y坐标平均绝对差异为1.67±0.14 mm(P<0.001),z坐标为2.75±0.15 mm(P<0.001),差异显著。
- 平均欧几里得向量差异为3.21±0.15 mm(P<0.001)。
核心结论
- 术中CT与术后MRI确定的脑深部电刺激电极y和z坐标存在显著差异,而x坐标无显著差异。
- 基于本系列,术中CT确认x坐标更准确,确认y和z坐标准确性较低。
研究设计与样本
- 回顾性研究,纳入2015至2017年单中心83例患者共145根脑深部电刺激电极。
- 将术中CT与术前MRI融合,分别在磁共振和CT图像上独立确定电极坐标,计算x、y、z坐标绝对差异及欧几里得向量。
主要发现
- 术中CT与术后MRI坐标平均绝对差异±标准误:x=0.01±0.09 mm(P=0.89),y=1.67±0.14 mm(P<0.001),z=2.75±0.15 mm(P<0.001)。
- 平均欧几里得向量差异为3.21±0.15 mm(P<0.001)。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Deep brain stimulation (DBS) is an effective therapy for a variety of refractory movement disorders. Accurate lead placement in the target nucleus is critical to ensure therapeutic effects and to minimize side effects, and intraoperative computed tomography (iCT) scan has been used to target and confirm lead position. The objective of this study is to compare the accuracy of determining the x, y, and z coordinates of final lead placement using iCT scan relative to postoperative magnetic resonance imaging (MRI). METHODS: We conducted a retrospective study on 83 patients who underwent insertion of 145 DBS leads from 2015 to 2017 at a single institution. iCT scan was merged with the preoperative MRI to determine lead coordinates on both magnetic resonance and computed tomography images independently, and the absolute differences between the x, y, and z coordinates between the 2 scans along with the Euclidean vectors were calculated. RESULTS: The mean absolute differences ± standard error of the mean between iCT scan and postoperative MRI coordinates were as follows: x = 0.01 ± 0.09 mm (P = 0.89), y = 1.67 ± 0.14 mm (P < 0.001), and z = 2.75 ± 0.15 mm (P < 0.001). The average Euclidean vector difference was 3.21 ± 0.15 mm (P < 0.001). CONCLUSIONS: Significant differences exist between iCT scan and postoperative MRI DBS y and z lead coordinates, but not with x coordinates. Based on this series, iCT scan is more accurate when confirming x coordinates, and less accurate for confirming y and z coordinates during DBS operations.
引用
- Girgis F, Zarabi H, Said M, et al. Comparison of Intraoperative Computed Tomography Scan with Postoperative Magnetic Resonance Imaging for Determining Deep Brain Stimulation Electrode Coordinates. World Neurosurg. 2020. PMID: 32112943. doi:10.1016/j.wneu.2020.02.108
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。