文献速递脊柱脊髓外科功能神经外科

原型硬膜内脊髓刺激器在猪模型中的体内测试

In Vivo Testing of a Prototype Intradural Spinal Cord Stimulator in a Porcine Model.
PubMed 文献 · World Neurosurg · 2020-02-26

本研究在6只雌性幼猪中测试原型硬膜内脊髓刺激器的安全性与可行性,其中1只仅行椎板切除术作为对照,5只植入装置,植入后3个月行终末手术并取脊柱行影像学与组织病理学检查,结果显示所有动物存活且无神经功能缺损或感染,未见脊髓损伤、脑脊液漏、骨过度生长或硬膜缺损,肌电图记录到诱发的运动单位。

要点

  • 6只雌性幼猪中5只植入硬膜内脊髓刺激器原型,1只仅行椎板切除作为对照。
  • 所有动物存活植入手术,无神经功能缺损或感染证据。
  • 影像学与组织病理学检查未见脊髓损伤、脑脊液漏、异常骨过度生长或硬膜缺损,装置内外板间存在有活力的硬膜。

核心结论

  • 慢性植入的硬膜内装置在猪模型中展示了安全性和向人类转化的可行性。

研究设计与样本

  • 研究在6只雌性幼猪中进行手术探索,其中1只对照动物仅行椎板切除术,另5只动物植入硬膜内脊髓刺激器原型。

主要发现

  • 所有动物均存活植入手术,无神经功能缺损或感染证据。
  • 脊柱死后影像学(微计算机断层扫描)和组织病理学检查未发现脊髓损伤、脑脊液瘘形成、异常骨过度生长或硬膜缺损。
  • 装置的内外板之间存在有活力的硬膜。
  • 肌电图记录显示刺激器诱发了运动单位。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Chronic midline low back pain is the number one reason for disability in the United States despite the prolific use of medical and surgical interventions. Notwithstanding the widespread use of epidural spinal cord stimulators (SCSs), there remains a large portion of the population with inadequate pain control thought to be because of the limited volume of stimulated neural tissue. Intradural SCSs represent an underexplored alternative strategy with the potential to improve selectivity, power efficiency, and efficacy. We studied and carried out development of an intradural form of an SCS. Herein we present the findings of in vivo testing of a prototype intradural SCS in a porcine model. METHODS: Six female juvenile pigs underwent surgical investigation. One control animal underwent a laminectomy only, whereas the 5 other animals had implantation of an intradural SCS prototype. One of the prototypes was fully wired to enable acute stimulation and concurrent electromyographic recordings. All animals underwent terminal surgery 3 months postimplantation, with harvesting of the spinal column. Imaging (microcomputed tomography scan) and histopathologic examinations were subsequently performed. RESULTS: All animals survived implantation without evidence of neurologic deficits or infection. Postmortem imaging and histopathologic examination of the spinal column revealed no evidence of spinal cord damage, cerebrospinal fluid fistula formation, abnormal bony overgrowth, or dural defect. Viable dura was present between the intra- and extradural plates of the device. Electromyographic recordings revealed evoked motor units from the stimulator. CONCLUSIONS: Chronically implanted intradural device in the porcine model demonstrated safety and feasibility for translation into humans.

来源:World Neurosurg,查看原文。版权归原作者所有。

引用

  1. Holland MT, Seaman SC, Woodroffe RW, et al. In Vivo Testing of a Prototype Intradural Spinal Cord Stimulator in a Porcine Model. World Neurosurg. 2020. PMID: 32112934. doi:10.1016/j.wneu.2020.02.100

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。