深部脑刺激患者脊柱磁共振成像的安全性评估
该研究通过体模测试评估1.5-T脊柱MRI中DBS装置产热,并将不安全序列修改为升温<2°C的安全方案,随后对9例DBS患者前瞻性完成67个脊柱MRI序列。结果显示无急性或长期不良事件,脑MRI无急性发现,装置阻抗稳定。
要点
- 颈椎扫描升温最高,腰椎扫描升温最低。
- 将3D序列改为2D并减少层数可使升温<2°C,但层数仍多于厂商指南允许值。
- 9例DBS患者完成67个脊柱MRI序列后无急性或长期不良事件,阻抗稳定。
核心结论
- 植入DBS装置的患者可安全接受层数少于常规临床方案的脊柱MRI。
- 安全性数据采集可能允许在厂商指南之外使用最大化层数的方案,减少放射科医师监督需求。
研究设计与样本
- 先用体模评估1.5-T脊柱MRI中DBS装置产热,将升温>2°C的常规临床序列修改为安全方案。
- 随后在前瞻性队列中,对9例需临床 imaging 的DBS参与者运行67个脊柱MRI序列,主要结局为急性不良反应,次要结局包括长期不良临床效应、脑MRI急性发现和装置阻抗稳定性。
主要发现
- 颈椎扫描时温度升高最高,腰椎扫描时最低。
- 将3D序列改为2D并将层数较常规脊柱MRI方案最少程度减少时,实现温度升高<2°C(但层数仍多于厂商指南允许值)。
- 脊柱MRI后未检测到急性或长期不良反应,脑MRI无急性发现,装置阻抗保持稳定。
关键数值与推荐
- 体模测试中,常规临床序列若温度升高>2°C则视为不安全并加以修改。
- 修改后安全方案实现温度升高<2°C。
临床怎么用
- 作者认为,植入DBS装置的患者可安全接受层数少于常规临床方案的脊柱MRI。
- 作者指出,安全性数据采集可能允许在厂商指南之外使用最大化层数的方案,减少放射科医师监督需求。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Many centers are hesitant to perform clinically indicated MRI in patients who have undergone deep brain stimulation (DBS). Highly restrictive guidelines prohibit the use of most routine clinical MRI protocols in these patients. The authors' goals were to assess the safety of spine MRI in patients with implanted DBS devices, first through phantom model testing and subsequently through validation in a DBS patient cohort. METHODS: A phantom was used to assess DBS device heating during 1.5-T spine MRI. To establish a safe spine protocol, routinely used clinical sequences deemed unsafe (a rise in temperature > 2°C) were modified to decrease the rise in temperature. This safe phantom-based protocol was then used to prospectively run 67 spine MRI sequences in 9 DBS participants requiring clinical imaging. The primary outcome was acute adverse effects; secondary outcomes included long-term adverse clinical effects, acute findings on brain MRI, and device impedance stability. RESULTS: The increases in temperature were highest when scanning the cervical spine and lowest when scanning the lumbar spine. A temperature rise < 2°C was achieved when 3D sequences were modified to 2D and when the number of slices was decreased by the minimum amount compared to routine spine MRI protocols (but there were still more slices than allowed by vendor guidelines). Following spine MRI, no acute or long-term adverse effects or acute findings on brain MR images were detected. Device impedances remained stable. CONCLUSIONS: Patients with DBS devices may safely undergo spine MRI with a fewer number of slices compared to those used in routine clinical protocols. Safety data acquisition may allow protocols outside vendor guidelines with a maximized number of slices, reducing the need for radiologist supervision.Clinical trial registration no.: NCT03753945 (ClinicalTrials.gov).
引用
- Boutet A, Elias GJB, Gramer R, et al. Safety assessment of spine MRI in deep brain stimulation patients. J Neurosurg Spine. 2020. PMID: 32059193. doi:10.3171/2019.12.SPINE191241
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。